Technical Note: Effect of transducer position and ground plane configuration on image quality in MR-guided focused ultrasound therapies.

Allison Payne1, Emilee Minalga1, Robb Merrill1, Dennis L Parker1, J Rock Hadley1.   

Abstract

PURPOSE: To evaluate the effect of a focused ultrasound transducer position and ground plane configuration on magnetic resonance image quality.
METHODS: The effect of transducer position with respect to the MRI B0 field and the radiofrequency receive coils was evaluated in a breast-specific MRgFUS system with an integrated RF phased-array coil. Image signal-to-noise ratio was evaluated at different transducer locations. The effect of ultrasound transducer ground plane configuration was evaluated using a replica transducer with 12 ground plane configurations. All evaluations were performed at 3 Tesla.
RESULTS: Both transducer position and ground plane configuration were found to have a considerable effect on overall image SNR. A 67% increase in SNR was achieved by positioning the transducer face perpendicular to the B0 field. A 25% increase in SNR was achieved by segmenting the replica transducer ground plane from one continuous plane to nine individual segments.
CONCLUSIONS: Advances in focused ultrasound hardware allow for integrated radiofrequency MRI coils as well as adjustable transducer positioning. The placement of the ultrasound transducer with respect to both the magnetic field and RF coils can have a considerable effect on image SNR and the resulting MR images that are used for MR-guided focused ultrasound treatment planning, monitoring and assessment.
© 2020 American Association of Physicists in Medicine.

Entities:  

Keywords:  focused ultrasound; magnetic resonance imaging; radiofrequency coils; transducer design

Mesh:

Year:  2020        PMID: 32170866      PMCID: PMC7802121          DOI: 10.1002/mp.14138

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  15 in total

1.  The NMR phased array.

Authors:  P B Roemer; W A Edelstein; C E Hayes; S P Souza; O M Mueller
Journal:  Magn Reson Med       Date:  1990-11       Impact factor: 4.668

2.  Analysis of eddy currents in nuclear magnetic resonance imaging.

Authors:  C B Ahn; Z H Cho
Journal:  Magn Reson Med       Date:  1991-01       Impact factor: 4.668

3.  Design and development of a prototype endocavitary probe for high-intensity focused ultrasound delivery with integrated magnetic resonance imaging.

Authors:  Iain P Wharton; Ian H Rivens; Gail R Ter Haar; David J Gilderdale; David J Collins; Jeff W Hand; Paul D Abel; Nandita M deSouza
Journal:  J Magn Reson Imaging       Date:  2007-03       Impact factor: 4.813

Review 4.  Magnetic resonance thermometry and its biological applications - Physical principles and practical considerations.

Authors:  Henrik Odéen; Dennis L Parker
Journal:  Prog Nucl Magn Reson Spectrosc       Date:  2019-01-31       Impact factor: 9.795

5.  The utility of pelvic coil SNR testing in the quality assurance of a clinical MRgFUS system.

Authors:  Krzysztof R Gorny; Nicholas J Hangiandreou; Heidi A Ward; Gina K Hesley; Douglas L Brown; Joel P Felmlee
Journal:  Phys Med Biol       Date:  2009-03-05       Impact factor: 3.609

6.  An 11-channel radio frequency phased array coil for magnetic resonance guided high-intensity focused ultrasound of the breast.

Authors:  E Minalga; A Payne; R Merrill; N Todd; S Vijayakumar; E Kholmovski; D L Parker; J R Hadley
Journal:  Magn Reson Med       Date:  2012-03-16       Impact factor: 4.668

7.  Clinical improvement associated with targeted interruption of the cerebellothalamic tract following MR-guided focused ultrasound for essential tremor.

Authors:  J Levi Chazen; Harini Sarva; Philip E Stieg; Robert J Min; Douglas J Ballon; Kane O Pryor; Paul M Riegelhaupt; Michael G Kaplitt
Journal:  J Neurosurg       Date:  2017-10-20       Impact factor: 5.115

8.  Development of custom RF coils for use in a small animal platform for magnetic resonance-guided focused ultrasound hyperthermia compatible with a clinical MRI scanner.

Authors:  Christopher Brian Abraham; Jak Loree-Spacek; Robert Andrew Drainville; Samuel Pichardo; Laura Curiel
Journal:  Int J Hyperthermia       Date:  2018-10-08       Impact factor: 3.914

9.  Open-source, small-animal magnetic resonance-guided focused ultrasound system.

Authors:  Megan E Poorman; Vandiver L Chaplin; Ken Wilkens; Mary D Dockery; Todd D Giorgio; William A Grissom; Charles F Caskey
Journal:  J Ther Ultrasound       Date:  2016-09-05

10.  Printed Receive Coils with High Acoustic Transparency for Magnetic Resonance Guided Focused Ultrasound.

Authors:  Joseph Corea; Patrick Ye; Dongjin Seo; Kim Butts-Pauly; Ana Claudia Arias; Michael Lustig
Journal:  Sci Rep       Date:  2018-02-21       Impact factor: 4.379

View more
  1 in total

1.  Transcranial MR-Guided Histotripsy System.

Authors:  Ning Lu; Timothy L Hall; Dave Choi; Dinank Gupta; Badih Junior Daou; Jonathan R Sukovich; Adam Fox; Tyler I Gerhardson; Aditya S Pandey; Douglas C Noll; Zhen Xu
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2021-08-27       Impact factor: 3.267

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.