Literature DB >> 32142620

Reliability, responsiveness, and validity of handheld dynamometry for assessing quadriceps strength in total knee arthroplasty.

Andrew J Kittelson1, Jesse C Christensen2, Brian J Loyd2, Kristine L Burrows1, Johna Iannitto1, Jennifer E Stevens-Lapsley1.   

Abstract

PURPOSE: To determine the reliability, responsiveness, validity, and agreement of handheld dynamometry, relative to electromechanical dynamometry, for assessing quadriceps strength surrounding total knee arthroplasty.
METHODS: Fifty-six patients (48% female) undergoing total knee arthroplasty were assessed prior to surgery and at 2 and 6 weeks following surgery. Maximum isometric quadriceps force was assessed at each time point by handheld dynamometry and electromechanical dynamometry. Within-session test-retest reliability was determined by intraclass correlation coefficient (ICC). Standard error of measurement and relative standard error (RSE) values were calculated. Standardized response means were used to describe responsiveness. Pearson's correlations examined construct validity. Agreement was assessed by the Bland-Altman method.
RESULTS: Both handheld dynamometry and electromechanical dynamometry demonstrated excellent test-retest reliability (ICC >0.90) and RSE (<15%). Both methods were responsive, with large postoperative standardized response means of 1.57 (handheld dynamometry) and 1.37 (electromechanical dynamometry). Pearson's correlations were moderate to strong. The Bland-Altman analysis revealed underestimation of force by handheld dynamometry, although this effect was diminished in the early postoperative period.
CONCLUSIONS: Our results suggest handheld dynamometry is a promising tool for monitoring quadriceps strength in patients pre- and post-total knee arthroplasty.IMPLICATIONS FOR REHABILITATIONThis study provides evidence of high reliability and responsiveness of handheld dynamometry for assessing quadriceps strength throughout a clinically relevant time frame for patients with total knee arthroplasty.There is construct validity of handheld dynamometry for measuring quadriceps strength, with moderate to high correlations between handheld dynamometry and electromechanical dynamometry, both pre- and post-total knee arthroplasty.Rehabilitation professionals should be aware that handheld dynamometry and electromechanical dynamometry do not agree in terms of the force measurement itself; particularly at higher force values, handheld dynamometry underestimates force output relative to electromechanical dynamometry.Overall, this study supports the use of handheld dynamometry for monitoring quadriceps strength in clinical settings for patient with total knee arthroplasty.

Entities:  

Keywords:  Total knee arthroplasty; handheld dynamometry; quadriceps strength; reliability; responsiveness; validity

Mesh:

Year:  2020        PMID: 32142620      PMCID: PMC7483560          DOI: 10.1080/09638288.2020.1730454

Source DB:  PubMed          Journal:  Disabil Rehabil        ISSN: 0963-8288            Impact factor:   2.439


  33 in total

1.  Clinical evaluator reliability for quantitative and manual muscle testing measures of strength in children.

Authors:  D M Escolar; E K Henricson; J Mayhew; J Florence; R Leshner; K M Patel; P R Clemens
Journal:  Muscle Nerve       Date:  2001-06       Impact factor: 3.217

2.  Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030.

Authors:  Steven Kurtz; Kevin Ong; Edmund Lau; Fionna Mowat; Michael Halpern
Journal:  J Bone Joint Surg Am       Date:  2007-04       Impact factor: 5.284

3.  Does Tourniquet Use in TKA Affect Recovery of Lower Extremity Strength and Function? A Randomized Trial.

Authors:  Douglas A Dennis; Andrew J Kittelson; Charlie C Yang; Todd M Miner; Raymond H Kim; Jennifer E Stevens-Lapsley
Journal:  Clin Orthop Relat Res       Date:  2016-01       Impact factor: 4.176

4.  A comparison of make and break tests using a hand-held dynamometer and the Kin-Com.

Authors:  P W Stratford; B E Balsor
Journal:  J Orthop Sports Phys Ther       Date:  1994-01       Impact factor: 4.751

5.  The reliability and validity of handheld dynamometry for the measurement of lower-extremity muscle strength in older adults.

Authors:  Cathy M Arnold; Kathryn D Warkentin; Philip D Chilibeck; Charlene R A Magnus
Journal:  J Strength Cond Res       Date:  2010-03       Impact factor: 3.775

6.  Early neuromuscular electrical stimulation to improve quadriceps muscle strength after total knee arthroplasty: a randomized controlled trial.

Authors:  Jennifer E Stevens-Lapsley; Jaclyn E Balter; Pamela Wolfe; Donald G Eckhoff; Wendy M Kohrt
Journal:  Phys Ther       Date:  2011-11-17

7.  Association of muscle strength with maximum walking speed in disabled older women.

Authors:  T Rantanen; J M Guralnik; G Izmirlian; J D Williamson; E M Simonsick; L Ferrucci; L P Fried
Journal:  Am J Phys Med Rehabil       Date:  1998 Jul-Aug       Impact factor: 2.159

8.  Younger age increases the risk of early prosthesis failure following primary total knee replacement for osteoarthritis. A follow-up study of 32,019 total knee replacements in the Finnish Arthroplasty Register.

Authors:  Jaakko Julin; Esa Jämsen; Timo Puolakka; Yrjö T Konttinen; Teemu Moilanen
Journal:  Acta Orthop       Date:  2010-08       Impact factor: 3.717

9.  Examining outcomes from total knee arthroplasty and the relationship between quadriceps strength and knee function over time.

Authors:  Yuri Yoshida; Ryan L Mizner; Dan K Ramsey; Lynn Snyder-Mackler
Journal:  Clin Biomech (Bristol, Avon)       Date:  2007-12-03       Impact factor: 2.063

10.  Assessing stability and change of four performance measures: a longitudinal study evaluating outcome following total hip and knee arthroplasty.

Authors:  Deborah M Kennedy; Paul W Stratford; Jean Wessel; Jeffrey D Gollish; Dianne Penney
Journal:  BMC Musculoskelet Disord       Date:  2005-01-28       Impact factor: 2.362

View more
  1 in total

1.  Development of reference charts for monitoring quadriceps strength with handheld dynamometry after total knee arthroplasty.

Authors:  Jeremy Graber; Elizabeth Juarez-Colunga; Charles Thigpen; Dawn Waugh; Michael Bade; Jennifer Stevens-Lapsley; Andrew Kittelson
Journal:  Disabil Rehabil       Date:  2021-11-09       Impact factor: 2.439

  1 in total