Literature DB >> 32108520

Three-Dimensional Printed Targeting Device for Scaphoid Fracture Fixation.

Matthew C DeWolf1, Alexander Hartov2, Thomas A Fortney1, Lance G Warhold1,3.   

Abstract

Background: Percutaneous guide wire insertion for scaphoid screw fixation can be challenging and often requires multiple attempts with significant radiation exposure to the surgical team. A 3-dimensional (3D) printed targeting device has the potential to reduce procedure time and intraoperative radiation exposure.
Methods: Our targeting device protocol included a preprocedure computed tomography (CT) scan of a casted cadaver wrist, followed by 3D printing of a customized targeting guide. In a comparison trial, seven orthopedic surgery residents performed percutaneous scaphoid guide wire insertion on different cadaver specimens by both freehand technique and using our targeting device. Radiation exposure and procedure times were compared. All specimens underwent postprocedure CT to assess Kirschner wire (K-wire) accuracy, determined by central third placement. Pre- and postprocedure CT scans from the targeting device group were co-registered to compare planned and actual K-wire trajectories.
Results: Using the freehand technique, mean fluoroscopy time was 120 seconds (standard deviation: ±53 seconds) generating 2.45 milligray of radiation. Average procedure time was 21 minutes with a mean of 6.4 (range: 3-9) insertion attempts. A single insertion attempt was made using the targeting device with an average procedure time of 30 seconds and no fluoroscopy exposure. Four K-wires were placed within the central scaphoid in both groups. Using the targeting device, average linear deviation from the planned trajectory was 2.1 mm, while the maximum linear deviation was 3.75 mm.
Conclusion: When compared to freehand scaphoid guide wire insertion, our targeting device provides similar accuracy while significantly reducing intraoperative radiation exposure and procedure time.

Entities:  

Keywords:  diagnosis; fracture/dislocation; research & health outcomes; scaphoid; specialty; surgery; trauma; treatment; wrist

Mesh:

Year:  2020        PMID: 32108520      PMCID: PMC8721798          DOI: 10.1177/1558944720906502

Source DB:  PubMed          Journal:  Hand (N Y)        ISSN: 1558-9447


  27 in total

1.  Radiation exposure of the hands and thyroid gland during percutaneous wiring of wrist and hand procedures.

Authors:  Syah Bahari; Seamus Morris; David Broe; Colm Taylor; Brian Lenehan; John McElwain
Journal:  Acta Orthop Belg       Date:  2006-04       Impact factor: 0.500

2.  A comparison of optical and electromagnetic computer-assisted navigation systems for fluoroscopic targeting.

Authors:  William M Ricci; Thomas A Russell; David M Kahler; Lauralan Terrill-Grisoni; Patrick Culley
Journal:  J Orthop Trauma       Date:  2008-03       Impact factor: 2.512

3.  Computer-assisted navigation of volar percutaneous scaphoid placement.

Authors:  Eric Walsh; Joseph J Crisco; Scott W Wolfe
Journal:  J Hand Surg Am       Date:  2009-11       Impact factor: 2.230

4.  Radiation exposure issues in orthopaedics.

Authors:  Brian D Giordano; Jonathan N Grauer; Christopher P Miller; Thomas L Morgan; Glenn R Rechtine
Journal:  J Bone Joint Surg Am       Date:  2011-06-15       Impact factor: 5.284

5.  Scaphoid fracture epidemiology.

Authors:  Andrew D Duckworth; Paul J Jenkins; Stuart A Aitken; Nicholas D Clement; Charles M Court-Brown; Margaret M McQueen
Journal:  J Trauma Acute Care Surg       Date:  2012-02       Impact factor: 3.313

6.  Current methods of diagnosis and treatment of scaphoid fractures.

Authors:  Steven J Rhemrev; Daan Ootes; Frank Jp Beeres; Sven Ag Meylaerts; Inger B Schipper
Journal:  Int J Emerg Med       Date:  2011-02-04

7.  Fixation of nondisplaced scaphoid fractures: making treatment cost effective. Prospective controlled trial.

Authors:  R Arora; M Gschwentner; D Krappinger; M Lutz; M Blauth; M Gabl
Journal:  Arch Orthop Trauma Surg       Date:  2006-09-27       Impact factor: 3.067

Review 8.  Fractures in brief: scaphoid fractures.

Authors:  Thomas Tysver; Andrew Jawa
Journal:  Clin Orthop Relat Res       Date:  2010-09       Impact factor: 4.176

9.  Percutaneous screw fixation versus conservative treatment for fractures of the waist of the scaphoid: a prospective randomised study.

Authors:  M M McQueen; M K Gelbke; A Wakefield; E M Will; C Gaebler
Journal:  J Bone Joint Surg Br       Date:  2008-01

Review 10.  Scaphoid nonunion: diagnosis and treatment.

Authors:  Victoria S Pao; James Chang
Journal:  Plast Reconstr Surg       Date:  2003-11       Impact factor: 4.730

View more
  1 in total

1.  Overview of In-Hospital 3D Printing and Practical Applications in Hand Surgery.

Authors:  Marco Keller; Alissa Guebeli; Florian Thieringer; Philipp Honigmann
Journal:  Biomed Res Int       Date:  2021-03-26       Impact factor: 3.411

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.