Abstract

PURPOSE: To investigate a computer-assisted technique for retrograde insertion of a percutaneous scaphoid screw and compare insertion time, accuracy, and radiation exposure to the traditional technique. We hypothesize that computer-assisted navigation of volar percutaneous scaphoid screw placement would improve accuracy, require less time, and diminish radiation exposure when compared to the traditional technique.
METHODS: Ten matched pairs of cadaveric wrists were randomized to computer-assisted versus traditional volar percutaneous scaphoid screw placement. Time of the overall procedure, set-up time, time for ideal guide wire placement, and radiation time were recorded. Number of K-wire attempts was also recorded. Finally, accuracy of planned screw axis and actual screw axis were compared. Student's t-tests and rank sums were used to determine whether the differences in outcome variables between computer-assisted and traditional techniques was significant, with an alpha level of 0.05.
RESULTS: Although the overall time of the 2 procedures and the set-up time were not different between the 2 groups, the time for placement of the K-wire was halved in the computer-assisted percutaneous scaphoid fixation group, and the number of K-wire attempts needed for accurate screw placement approached clinical significance. Although the radiation exposures for the individual components of set-up time and final check time were not different, the radiation exposures for global time of the procedure, K-wire placement, and screw placement were clinically significant.
CONCLUSIONS: Computer-assisted navigation of volar percutaneous scaphoid screw placement takes no more time that traditional methods and significantly reduces the amount of radiation exposure to the patient. Although not statistically significant, the technique reduced the number of incorrect passes of the K-wire, requiring a single attempt in 4 of the 5 specimens.

Entities:  

Mesh:

Year:  2009        PMID: 19896013     DOI: 10.1016/j.jhsa.2009.08.009

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  8 in total

Review 1.  Computer-assisted Surgery for Scaphoid Fracture.

Authors:  Zi-Run Xiao; Ge Xiong
Journal:  Curr Med Sci       Date:  2018-12-07

2.  Volume slicing of cone-beam computed tomography images for navigation of percutaneous scaphoid fixation.

Authors:  Erin J Smith; Hisham A Al-Sanawi; Braden Gammon; Paul J St John; David R Pichora; Randy E Ellis
Journal:  Int J Comput Assist Radiol Surg       Date:  2011-06-25       Impact factor: 2.924

3.  Investigating the performance of a wrist stabilization device for image-guided percutaneous scaphoid fixation.

Authors:  Erin J Smith; Gregory Allan; Braden Gammon; Richard W Sellens; Randy E Ellis; David R Pichora
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-04-16       Impact factor: 2.924

4.  Computer-assisted percutaneous scaphoid fixation: concepts and evolution.

Authors:  Erin J Smith; Randy E Ellis; David R Pichora
Journal:  J Wrist Surg       Date:  2013-11

5.  Influence of screw design, sex, and approach in scaphoid fracture fixation.

Authors:  Geert Meermans; Frederik Verstreken
Journal:  Clin Orthop Relat Res       Date:  2011-12-17       Impact factor: 4.176

6.  Image guidance can support scaphoid K-wire insertion: an experimental study and initial clinical experience.

Authors:  Hendrik Schöll; Martin Mentzel; Almut Jones; Joachim Gülke; Florian Gebhard; Michael Kraus
Journal:  Int J Comput Assist Radiol Surg       Date:  2012-11-30       Impact factor: 2.924

7.  Three-Dimensional Printed Targeting Device for Scaphoid Fracture Fixation.

Authors:  Matthew C DeWolf; Alexander Hartov; Thomas A Fortney; Lance G Warhold
Journal:  Hand (N Y)       Date:  2020-02-28

8.  Cold ablation robot-guided laser osteotomy in hand, wrist and forearm surgery-A feasibility study.

Authors:  Philipp Honigmann; Maximilian Hofer; Sibylle Hirsch; Marta Morawska; Magdalena Müller-Gerbl; Florian M Thieringer; Enrico Coppo
Journal:  Int J Med Robot       Date:  2022-07-08       Impact factor: 2.483

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.