Sue Hey Chae1, Hyeck Soo Son2, Ramin Khoramnia2, Kyung Heon Lee1, Chul Young Choi3.
Abstract
BACKGROUND: To experimentally compare the optical performance of two different Extended-Depth-of-Focus (EDOF) intraocular lenses (IOLs) using a standardized optical bench set-up.Entities:
Keywords: Extended-depth-of-focus; Intraocular lenses; Optical quality; Pupil size
Mesh:
Year: 2020 PMID: 32059666 PMCID: PMC7023787 DOI: 10.1186/s12886-020-1332-6
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Overview of the optical properties of the studied intraocular lenses
| IOL | Symfony | AT LARA |
|---|---|---|
| Optic design | Aspheric, diffractive-refractive echelette feature | Aspheric, diffractive-refractive SMP (smooth microphase) design |
| Material | UV blocking hydrophobic acrylic | UV blocking, Hydrophilic acrylate (25% water content) with hydrophobic surface properties |
| Body design | Single-piece/C-loop, square edge | Single-piece, plate-haptic, square edge |
| Optical/total diameter (mm) | 6/13 | 6/11 |
| Dioptric power (D) | 20.00 | 20.00 |
| Spherical aberration (μm) | −0.27 μm | 0.00 μm |
| Diffractive Steps (n) | 9 | 15 |
| Refractive Index | 1.47 | 1.46 |
| Abbe Number | 55 | 56.5 |
Fig. 1Through-focus curves at 50 lp/mm and at pupil sizes of 2, 3, and 4.5 mm, as measured with ISO1 and 2 model corneas
Modulation transfer function values measured with the ISO1 and ISO2 model corneas at primary and secondary peak foci for different pupil sizes and the according Peak 2/1 Ratio percentage. IOL = Intraocular Lens; MTF = Modulation Transfer function
| IOL | Cornea model | Pupil size | Peak 1 | Peak 2 | Peak 2/1 Ratio (%) | ||
|---|---|---|---|---|---|---|---|
| Position in Diopter | Mean MTF value | Position in Diopter | Mean MTF value | ||||
| Symfony | ISO 1 | 2 mm | 20 | 0.361 | + 1.41 | 0.431 | 119.4% |
| 3 mm | 20 | 0.387 | + 1.59 | 0.416 | 107.5% | ||
| 4.5 mm | 20 | 0.225 | + 1.85 | 0.252 | 112.0% | ||
| ISO 2 | 2 mm | 20 | 0.410 | + 1.57 | 0.424 | 103.4% | |
| 3 mm | 20 | 0.418 | + 1.65 | 0.395 | 94.5% | ||
| 4.5 mm | 20 | 0.437 | + 1.69 | 0.344 | 78.7% | ||
| LARA | ISO 1 | 2 mm | 20 | 0.311 | + 1.85 | 0.442 | 142.1% |
| 3 mm | 20 | 0.365 | + 1.87 | 0.323 | 88.5% | ||
| 4.5 mm | 20 | 0.487 | + 1.96 | 0.257 | 52.8% | ||
| ISO 2 | 2 mm | 20 | 0.339 | + 1.85 | 0.380 | 112.1% | |
| 3 mm | 20 | 0.351 | + 1.65 | 0.209 | 59.5% | ||
| 4.5 mm | 20 | 0.169 | + 1.19 | 0.187 | 110.7% | ||
Fig. 2United States Air Force target images of the Symfony IOL at varying pupil sizes and defocus values with ISO1 and 2 model corneas
Fig. 3United States Air Force target images of the AT LARA IOL at varying pupil sizes and defocus values with ISO1 and 2 model corneas