| Literature DB >> 32039095 |
S Kaushik1, R Punia2,3, A Tyagi4.
Abstract
BACKGROUND: This study aims to investigate radiation beam geometry of Cyberknife beam and change in dosimetric characteristics of six megavoltage (6MV) flattening filter free (FFF) beam after passing through high density cadmium free compensator alloy.Entities:
Keywords: Alloys; Photons ; X-rays
Year: 2019 PMID: 32039095 PMCID: PMC6943851 DOI: 10.31661/jbpe.v0i0.745
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Figure1Experimental setup for in air transmission measurement.
Figure2a) Variation of transmitted intensity (I) with thickness of compensator alloy at different fixed collimator aperture (A) for 800 mm SDD. b) Variation of transmitted intensity (I) with thickness of compensator alloy at different fixed collimator aperture (A) for 1270 mm SDD.
Variation of experimentally measured maximum intensity (I0), maximum intensity obtained from log I vs. t graph (I’0) and best fit parameter (R2) with collimator aperture (A) for 800 mm SDD.
| A (mm) | I0 | I'0 | R2 |
|---|---|---|---|
|
| 2.29E-09 | 2.49E-09 | 0.99676 |
|
| 2.97E-09 | 3.16E-09 | 0.99739 |
|
| 3.02E-09 | 3.20E-09 | 0.99766 |
|
| 3.02E-09 | 3.20E-09 | 0.99789 |
|
| 3.04E-09 | 3.21E-09 | 0.99786 |
|
| 3.05E-09 | 3.21E-09 | 0.99813 |
Variation of experimentally measured maximum intensity (I0), maximum intensity obtained from log I vs. t graph (I’0) and best fit parameter (R2) with field sizes (D) for 1270 mm SDD.
| A (mm) | I0 | I'0 | R2 |
|---|---|---|---|
| 10 | 1.11E-09 | 1.17E-09 | 0.99387 |
| 20 | 1.19E-09 | 1.28E-09 | 0.9957 |
| 30 | 1.21E-09 | 1.28E-09 | 0.99621 |
| 40 | 1.21E-09 | 1.29E-09 | 0.99621 |
| 50 | 1.22E-09 | 1.29E-09 | 0.99613 |
| 60 | 1.22E-09 | 1.29E-09 | 0.99613 |
Figure3Variation in Surface dose (Ds) with thickness (t) of compensator alloy for different fixed collimator aperture (A).
Figure4a) PDDs showing beam hardening with different compensator thickness for A= 60 mm. b) Effect of scattering due to increasing collimator opening on PDDs with 7.1 mm compensator.
Variation of linear attenuation coefficient (µeff) obtained from slope of log I vs. t graph with collimator aperture (A) for 800 mm SDD and 1270 mm SDD.
| A (mm) | µeff (±0.00099 mm-1) measured at 800 mm SDD | µeff (±0.00014 mm-1) measured at 1270 mm SDD |
|---|---|---|
|
| 0.04273 | 0.04295 |
|
| 0.04272 | 0.04214 |
|
| 0.04307 | 0.04236 |
|
| 0.04291 | 0.04236 |
|
| 0.04215 | 0.04246 |
|
| 0.04212 | 0.04246 |