Broad beam and narrow beam attenuation in Lipowitz's metal.

E E el-Khatib, E B Podgorsak, C Pla.   

Abstract

Attenuation properties of Lipowitz's metal have been studied for narrow and broad beams of cobalt-60 gamma rays and 4-10 MV x-rays. The measured transmitted fraction for geometries used in radiotherapy depends on the field size and depth of measurement. Therefore a calculation of dose for partially attenuated beams based on narrow beam attenuation coefficients can cause large errors in dosimetry. Our simple calculation of transmitted fractions based on primary attenuation and scattered radiation agrees quite well with the measured data for therapeutic geometries. Also given is a table for linear, mass attenuation, and mass energy absorption coefficients of Lipowitz's metal in the photon energy range from 10 keV to 10 MeV.

Entities:  

Mesh:

Substances:

Year:  1987        PMID: 3104738     DOI: 10.1118/1.596100

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  3 in total

1.  Absorbed dose and transmission factor in blocked fields for radiotherapy treatments with 6 MV X-ray.

Authors:  Ana Elena Buenfil; Marvin Francisco Rodríguez-González; Gabriel Reséndiz González; María Ester Brandan
Journal:  Clin Transl Oncol       Date:  2006-04       Impact factor: 3.405

2.  Effect of electron contamination on in vivo dosimetry for lung block shielding during TBI.

Authors:  Ganesh Narayanasamy; Wilbert Cruz; Daniel L Saenz; Sotirios Stathakis; Niko Papanikolaou; Neil Kirby
Journal:  J Appl Clin Med Phys       Date:  2016-05-08       Impact factor: 2.102

3.  Study of Dosimetric Properties of Flattening Filter Free Photon Beam Passing through Cadmium Free Compensator Alloy.

Authors:  S Kaushik; R Punia; A Tyagi
Journal:  J Biomed Phys Eng       Date:  2019-12-01
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.