S Singh1, P Raina2, O P Gurjar3.
Abstract
BACKGROUND: Nowadays, advanced radiotherapy equipment includes algorithms to calculate dose. The verification of the calculated doses is important to achieve accurate results. Mostly homogeneous dosimetric phantoms are available commercially which do not mimic the actual patient anatomy; therefore, an indigenous heterogeneous pelvic phantom mimicking actual human pelvic region has been used to verify the doses calculated by different algorithms.Entities:
Keywords: Algorithms ; Dosimetric Phantoms ; Particle Accelerators, Pelvis; Radiotherapy, Conformal
Year: 2019 PMID: 32039091 PMCID: PMC6943848 DOI: 10.31661/jbpe.v0i0.1185
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
The density of different organs in indigenous pelvic phantom and the actual patient’s pelvis
| Organs | Relative Electron Density ± Standard Deviation | |
|---|---|---|
| Pelvic phantom | Actual human | |
| Rectum | 1.068 ± 0.005 | 1.067 ± 0.015 |
| Bladder | 1.044 ± 0.004 | 1.043 ± 0.008 |
| Air Cavity | 0.192 ± 0.196 | 0.204 ± 0.221 |
| Pelvic bone | 1.455 ± 0.112 | 1.497 ± 0.134 |
| Fat | 0.956 ± 0.011 | 0.954 ± 0.007 |
In 3DCRT plans Percentage of variation between planned dose on treatment planning system and measured dose on linear accelerator using pelvic phantom.
| Clarkson | Convolution | Superposition | Fast Superposition | ||
|---|---|---|---|---|---|
|
| Measured Mean Dose (cGy) | 100.66 | 100.72 | 103.4 | 102.98 |
| TPS Dose (cGy) | 100.10 | 100.9 | 100.3 | 100.9 | |
| % Variation | 0.56 | - 0.18 | 3.09 | 2.06 | |
|
| Measured Mean Dose (cGy) | 100.14 | 99.8 | 103.15 | 102.8 |
| TPS Dose (cGy) | 100.7 | 100.4 | 100.1 | 100.9 | |
| % Variation | - 0.56 | - 0.59 | 3.04 | 1.88 | |
|
| Measured Mean Dose (cGy) | 100.56 | 100.82 | 103.01 | 102.98 |
| TPS Dose (cGy) | 100.10 | 100.90 | 100.01 | 100.91 | |
| %Variation | 0.46 | - 0.08 | 2.99 | 2.05 | |
|
| Measured Mean Dose (cGy) | 100.82 | 100.02 | 103.08 | 102.17 |
| TPS Dose (cGy) | 100.05 | 99.9 | 100.45 | 100.30 | |
| % Variation | 0.77 | 0.12 | 2.62 | 1.86 | |
Figure1Dose variation between planned dose and measured dose with 10 x 10cm2 field size for (A) Single Field B. single field with 45oWedge, (C) Opposite field and (D) Four field.
Figure2Doses wash with Superposition Algorithm planed with (A) Single Field (B) Wedge Field(C), Opposite Field and (D) Four Field.