| Literature DB >> 31916401 |
Richard Ryan Wargo1, Areej Fawzi Aljabal2, Pei-Jan Paul Lin1.
Abstract
PURPOSE: This technical note presents an inexpensive tool and method for determining lead equivalency using digital radiography x-ray equipment.Entities:
Keywords: attenuation measurement; lead and nonlead protection apparel; lead equivalence; lead foil tape
Mesh:
Substances:
Year: 2020 PMID: 31916401 PMCID: PMC7020982 DOI: 10.1002/acm2.12810
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Drawing of test tool using lead 3M Lead Foil Tape 421.
Carestream DRX revolution beam quality.
| X‐ray tube voltage (kV) | HVL (mm Al) |
|---|---|
| 60 | 2.50 |
| 70 | 2.96 |
| 80 | 3.42 |
| 90 | 3.85 |
| 100 | 4.35 |
| 110 | 4.89 |
| 120 | 5.37 |
Figure 2Digital radiographic image of test tool, radiation protection apparel, and lead sheets.
Caliper measurements for lead sheets.
| Commercial lead sheets | Side 1 (mm) | Side 2 (mm) | Side 3 (mm) | Side 4 (mm) |
Average (mm) |
|---|---|---|---|---|---|
| Pb – 1/32”(0.794 mm) | 0.880 | 0.885 | 0.890 | 0.905 | 0.894 |
| 0.900 | 0.890 | 0.905 | 0.895 | ||
| Pb – 1/64”(0.397 mm) | 0.410 | 0.460 | 0.470 | 0.395 | 0.424 |
| 0.420 | 0.420 | 0.380 | 0.440 |
Lead sheets used for validation of the test tool were measured using a caliper.
Detector uniformity.
| ROI position | A (UL) | B (UR) | C (C) | D (LL) | E (LR) | CV | Detector air kerma (μGy) |
|---|---|---|---|---|---|---|---|
| Mean pixel value | 1450 | 1404 | 1455 | 1442 | 1409 | 1.7% | 1.1 |
| 13031 | 12617 | 13118 | 13035 | 12725 | 1.7% | 10.3 | |
| 14328 | 13876 | 14429 | 14340 | 13997 | 1.7% | 11.4 | |
| 29013 | 28099 | 29210 | 29037 | 28354 | 1.7% | 23.2 |
C, Center; CV, Coefficient of Variation; LL, Lower Left; LR, Lower Right; UL, Upper Left; UR, Upper Right.
Uniformity was evaluated objectively and visually at varying detector air kerma levels.
Archer’s equation fitting parameters for transmission curves corresponding to varying tube voltages.
| kV | 60 | 70 | 80 | 90 | 100 | 110 | 120 |
|---|---|---|---|---|---|---|---|
| α | 4.528 | 4.678 | 3.723 | 3.170 | 3.032 | 2.983 | 2.937 |
| β | 14.553 | 11.717 | 9.919 | 8.662 | 7.509 | 6.597 | 5.930 |
| γ | 0.474 | 0.755 | 0.875 | 0.937 | 0.969 | 0.987 | 0.996 |
Calculated versus measured lead thickness for lead sheets using test tool
| Output | kV | DAK (μGy) | kV | DAK (μGy) | kV | DAK (μGy) | kV | DAK (μGy) | kV | DAK (μGy) | kV | DAK (μGy) | kV | DAK (μGy) |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 60 | 21.6 | 70 | 19.7 | 80 | 19.3 | 90 | 21.5 | 100 | 18.5 | 110 | 19.5 | 120 | 22.8 | |
| Calculation method | AF (mm) | LI (mm) | AF (mm) | LI (mm) | AF (mm) | LI (mm) | AF (mm) | LI (mm) | AF (mm) | LI (mm) | AF (mm) | LI (mm) | AF (mm) | LI (mm) |
| Pb ‐ 1/32" (0.894 mm) | 0.807 | 0.957 | 0.875 | 0.930 | 0.874 | 0.915 | 0.878 | 0.913 | 0.886 | 0.920 | 0.892 | 0.927 | 0.898 | 0.933 |
| Pb ‐ 1/64" (0.424 mm) | 0.445 | 0.445 | 0.439 | 0.443 | 0.438 | 0.440 | 0.438 | 0.440 | 0.437 | 0.438 | 0.436 | 0.437 | 0.436 | 0.437 |
| Percent Error | −9.7% | 7.1% | −2.1% | 4.1% | −2.2% | 2.4% | −1.8% | 2.2% | −0.9% | 2.9% | −0.2% | 3.7% | 0.5% | 4.4% |
| 4.9% | 4.9% | 3.4% | 4.4% | 3.2% | 3.7% | 3.2% | 3.7% | 3.0% | 3.2% | 2.7% | 3.0% | 2.7% | 3.0% |
AF, Archer’s Fit; DAK, Detector Air Kerma; LI, Linear Interpolation.
Calculated lead equivalencies for radiation protection apparel using test tool.
| Output | kV | DAK (μGy) | kV | DAK (μGy) | kV | DAK (μGy) | kV | DAK (μGy) | kV | DAK (μGy) | kV | DAK (μGy) | kV | DAK (μGy) |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 60 | 21.6 | 70 | 19.7 | 80 | 19.3 | 90 | 21.5 | 100 | 18.5 | 110 | 19.5 | 120 | 22.8 | |
| Calculation method | AF (mm) | LI (mm) | AF (mm) | LI (mm) | AF (mm) | LI (mm) | AF (mm) | LI (mm) | AF (mm) | LI (mm) | AF (mm) | LI (mm) | AF (mm) | LI (mm) |
| Glasses — Lenses (L) [0.75 mm PbEq] | 0.65 | 0.72 | 0.77 | 0.80 | 0.82 | 0.85 | 0.83 | 0.86 | 0.84 | 0.87 | 0.85 | 0.88 | 0.86 | 0.89 |
| − | − |
|
|
|
|
|
|
|
|
|
|
|
| |
| Glasses — Lenses (R) [0.75 mm PbEq] | 0.68 | 0.78 | 0.77 | 0.81 | 0.81 | 0.84 | 0.82 | 0.85 | 0.82 | 0.85 | 0.83 | 0.85 | 0.83 | 0.86 |
| − |
|
|
|
|
|
|
|
|
|
|
|
|
| |
| Glasses — Side Shield (L) [0.35 mm PbEq] | 0.37 | 0.38 | 0.37 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.39 | 0.39 | 0.39 | 0.39 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| Glasses — Side Shield (R) [0.35 mm PbEq] | 0.37 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 | 0.38 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
| Thyroid Shield [0.5 mm PbEq] | 0.46 | 0.46 | 0.46 | 0.47 | 0.47 | 0.47 | 0.47 | 0.47 | 0.46 | 0.46 | 0.46 | 0.46 | 0.45 | 0.45 |
| − |
|
|
| − | − | − | − | − | − | − | − | − |
| |
| Apron [0.35 mm PbEq] | 0.39 | 0.37 | 0.39 | 0.39 | 0.39 | 0.39 | 0.39 | 0.39 | 0.38 | 0.38 | 0.37 | 0.38 | 0.37 | 0.37 |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
AF, Archer’s Fit; DAK, Detector Air Kerma; LI, Linear Interpolation; PbEq, Lead Equivalency.
Per cent error between vendor/manufacturer reported lead equivalence and calculated lead equivalence using test tool is provided underneath each calculated lead equivalence.