Characterization of the Mixed Radiation Field Produced by Carbon and Oxygen Ion Beams of Therapeutic Energy: A Monte Carlo Simulation Study.

C K Ying1, David Bolst2, Anatoly Rosenfeld2, Susanna Guatelli2.   

Abstract

PURPOSE: The main advantages of charged particle radiotherapy compared to conventional X-ray external beam radiotherapy are a better tumor conformality coupled with the capability of treating deep-seated radio-resistant tumors. This work investigates the possibility to use oxygen beams for hadron therapy, as an alternative to carbon ions.
MATERIALS AND METHODS: Oxygen ions have the advantage of a higher relative biological effectiveness (RBE) and better conformality to the tumor target. This work describes the mixed radiation field produced by an oxygen beam in water and compares it to the one produced by a therapeutic carbon ion beam. The study has been performed using Geant4 simulations. The dose is calculated for incident carbon ions with energies of 162 MeV/u and 290 MeV/u, and oxygen ions with energies of 192 MeV/u and 245 MeV/u, and hence that the range of the primary oxygen ions projectiles in water was located at the same depth as the carbon ions.
RESULTS: The results show that the benefits of oxygen ions are more pronounced when using lower energies because of a slightly higher peak-to-entrance ratio, which allows either providing higher dose in tumor target or reducing it in the surrounding healthy tissues. It is observed that, per incident particle, oxygen ions deliver higher doses than carbon ions.
CONCLUSIONS: This result coupled with the higher RBE shows that it may be possible to use a lower fluence of oxygen ions to achieve the same therapeutic dose in the patient as that obtained with carbon ion therapy. Copyright:
© 2019 Journal of Medical Physics.

Entities:  

Keywords:  Carbon ions; Geant 4; Monte Carlo simulation; charged particle therapy; oxygen ions

Year:  2019        PMID: 31908385      PMCID: PMC6936202          DOI: 10.4103/jmp.JMP_40_19

Source DB:  PubMed          Journal:  J Med Phys        ISSN: 0971-6203


INTRODUCTION

Proton and carbon ion therapy are two types of hadron therapy which have been increasingly used in recent years for cancer treatment, with many new facilities under construction or planned for the near future.[123456] One of the main advantages of charged particle radiotherapy compared to conventional X-ray external beam radiotherapy is a better dose conformality. Heavy charged particles deposit the major part of their energy at the end of their range, at the Bragg peak. This allows to provide greater dose in the tumor while sparing the surrounding healthy tissues. For this reason, hadron therapy is particularly suitable for deep-seated tumors.[78] Currently, hadron therapy centers in the United States offer only proton beams.[9] Carbon ion therapy is currently available in Japan, for example, at the heavy-ion medical accelerator (HIMAC) at the National Institute of Radiological Sciences (NIRS, Chiba), and in Europe, for example, the Heidelberg Ion-Beam Therapy Center (HIT). Oxygen ions are currently considered as a potential alternative to carbon ions. Because of their mass, oxygen ions have less lateral scattering. This contributes favorably to the tumor conformality. The high linear energy transfer (LET) of oxygen ions when compared to carbon ions is associated with higher radiobiological effectiveness (RBE) which translates to better treatment effectiveness when treating hypoxic tumors.[101112] Kurz et al. published the results of the first experimental-based study of an oxygen ion beam in 2012. In particular, they measured the depth-dose distributions of oxygen ions beams at the HIT. Their work contributed to the development of a preclinical oxygen ion-beam experimental database, already being used in research purpose treatment planning systems to support radiobiological experiments with oxygen ion beams at HIT.[13] With respect to carbon ions, oxygen ions produce more nuclear fragments, which need to be investigated, as they can affect the RBE and the sensitivity to oxygenation.[14] In addition, the dose due to such fragments needs to be investigated not only in-field but also out-of-field, laterally and beyond the Bragg peak, to study the effect of the mixed radiation field in the healthy tissues surrounding the tumor target. The possible use of helium and oxygen ions in hadron therapy has recently gained more attention. Yu-Shen et al. investigated the longitudinal and lateral dose profiles and RBE produced by protons, α particles, and carbon ions.[15] Tessonnier et al. studied the basic dosimetric features of the different ions, including oxygen ions by experiments.[16] Mattei et al. studied the prompt gamma-ray emission by the interaction of helium, carbon, and oxygen ion beams with a polymethyl methacrylate phantom.[17] Burigo et al. compared the dose distributions and cell survival fractions for various incident ions. They concluded that the optimal projectile depends on the location of the tumor and on the radiosensitivity of the irradiated tissues.[18] This work compares the dose produced in a water phantom by an oxygen ion (16O) beam with respect to a carbon (12C) ion beam. This work has been done by using Geant4 Monte Carlo simulations.[19]

MATERIALS AND METHODS

Geant4 Monte Carlo simulation toolkit, from the European Organization for Nuclear Research, CERN (Geneva, Switzerland) version 4.10.1.p01 was used to model and compare the radiation fields produced by carbon and oxygen ion beams in a 30 cm × 30 cm × 30 cm water phantom. A 5 cm × 5 cm square beam is simulated normally incident on the water phantom with an air gap of 30 cm between the particle source and the surface of the phantom. Carbon beams were generated with energies of 162 MeV/u and 290 MeV/u, which are typical energies used in carbon ion therapy at HIMAC.[20] The simulations were repeated substituting the carbon with an oxygen beam with energies 192 MeV/u and 345 MeV/u (corresponding to ranges of approximately 60 mm and 161 mm in water, similar to the carbon ion beam). The G4EmStandardPhysics_Option3 and QGSP_BIC_EMY Geant4 Physics List were adopted to model the electromagnetic and hadronic physics interactions.[21] The results obtained with oxygen and carbon ion beams were compared in terms of dose distribution in the phantom and peak to entrance dose ratio. The contribution to dose deriving from secondary fragments was studied as well. 106 histories were simulated for each configuration to obtain a statistical uncertainty within 1%, with the corresponding confidence interval of the uncertainty being 95%.

RESULTS AND DISCUSSION

Figure 1 shows the energy deposited in the phantom per incident particle in the case of incident 162 MeV/u 12C and 192 MeV/u 16O ions. The 16O beam produces approximately 50% higher energy deposition than the 12C beam at the entrance, at 2 mm depth from the surface of the phantom. At the Bragg peak, approximately at 60-mm depth from the surface of the phantom, the 16O beam produces approximately 60% higher energy deposition than the 12C beam. In addition, the 16O beam produces 150% higher energy deposition tail just after the Bragg peak (at 2-mm depth after the Bragg peak). The peak-to-entrance ratio is 9.0 for the 16O, just slightly higher than that of 12C beams, which is 8.5.
Figure 1

Depth-energy deposition profile in water for 162 MeV/u 12C (red) and 192 MeV/u 16O beams (blue) per incident particle, with 30 cm × 30 cm × 0.01 cm voxels (0.01 cm along the direction of the beam)

Depth-energy deposition profile in water for 162 MeV/u 12C (red) and 192 MeV/u 16O beams (blue) per incident particle, with 30 cm × 30 cm × 0.01 cm voxels (0.01 cm along the direction of the beam) Figure 2 shows the energy deposition per incident particle, for 12C and 16O beams with energies 290 MeV/u and 345 MeV/u, respectively. In this case, it was found that the dose at the entrance (at 2-mm depth from the surface of the phantom) and at the Bragg peak (approximately at 60-mm depth), produced by the 16O ions was about 60% higher than in the case of the 12C beam. Still, the dose at the distal edge of the Bragg peak (2 mm after the Bragg peak) is 140% higher in the case of irradiation with oxygen beams. This is a shortcoming of oxygen ions (organs at risk may be located at the distal edge) and should be taken into consideration if switching from carbon ions to oxygen ions in hadron therapy. The peak-to-entrance ratio for 16O and 12C beams is found to be 5.4 and 5.8, respectively
Figure 2

Depth-energy deposition profile in water for 290 MeV/u 12C (red) and 345 MeV/u 16O beams (blue) per incident particle, with 30 cm × 30 cm × 0.01 cm voxels (0.01 cm along the direction of the beam)

Depth-energy deposition profile in water for 290 MeV/u 12C (red) and 345 MeV/u 16O beams (blue) per incident particle, with 30 cm × 30 cm × 0.01 cm voxels (0.01 cm along the direction of the beam) These results are in agreement with Francesco et al., who compared absolute dose per unit fluence produced by four different types of ion species (protons and helium, carbon and oxygen ions).[12] Higher peak-to-entrance ratio means that the dose may be increased in the tumor target or reduced in the surrounding healthy tissues of the patient, in front of the tumor. Results of the current study show that increasing the beam energy decreases the peak-to-entrance ratio for both oxygen and carbon ions. In terms of peak-to-entrance ratio, the benefits of oxygen ions are more pronounced at lower energy. Figure 3 shows the energy deposited by the secondary radiation field produced by the incident carbon ions (162 MeV/u) and oxygen ions (192 MeV/u) along the depth in the phantom. The results are normalized per incident primary. The results clearly illustrate that fragmentation has a more prominent role in the case of the oxygen beam, as expected, in all regions (i.e., the plateau region, the Bragg peak position, and beyond the Bragg peak). The dose contribution of individual secondary fragment species with higher production yield is presented in Figure 4. Both hydrogen and helium fragments have a longer range than incident ions and heavier fragments, and they contribute most to the dose beyond the Bragg peak.
Figure 3

Depth-energy deposition curve produced by the secondary fragments only, with incident 16O (blue) and 12C beams (red) with energy 162 MeV/u and 192 MeV/u. The voxel sizes are 30 cm × 30 cm × 0.01 cm (0.01 cm along the direction of the beam). The black line at a depth of 60 mm indicates the position of the Bragg peak

Figure 4

Comparison of depth-energy deposition curves produced by some fragment species, for incident oxygen (red) and carbon (blue) ion beams with energy 162 MeV/u and 192 MeV/u, respectively. The black line at a depth of 60 mm indicates the position of the Bragg peak

Depth-energy deposition curve produced by the secondary fragments only, with incident 16O (blue) and 12C beams (red) with energy 162 MeV/u and 192 MeV/u. The voxel sizes are 30 cm × 30 cm × 0.01 cm (0.01 cm along the direction of the beam). The black line at a depth of 60 mm indicates the position of the Bragg peak Comparison of depth-energy deposition curves produced by some fragment species, for incident oxygen (red) and carbon (blue) ion beams with energy 162 MeV/u and 192 MeV/u, respectively. The black line at a depth of 60 mm indicates the position of the Bragg peak It can be observed that secondary protons, produced by incident oxygen ion beams deposit more energy than in the case of carbon ions beams in all regions (i.e., the plateau region, the Bragg peak position, and beyond the Bragg peak position). The contribution to the energy deposition deriving from secondary alpha particles is similar along the Bragg peak when comparing oxygen and carbon ion beams. After the Bragg peak, in the case of oxygen ions, alpha particles deposit more energy. In the case of carbon ion beams, secondary lithium (Li), beryllium (Be), and boron (B) fragments produce more dose in the plateau region and at the Bragg peak position compared to oxygen ions, but, after the Bragg peak, the dose contribution of such fragments is higher with incident oxygen ions. The energy deposition contribution of secondary nitrogen (N) fragments produced by primary oxygen ions is dominant everywhere along the Bragg peak. Lateral dose distributions for several selected secondary fragments produced by incident oxygen and carbon ions have been compared. Figure 5 shows the results for protons (P) and helium (He), lithium (Li) and boron (B) ion fragments produced by incident 162 MeV/u 12C and 192 MeV/u 16O ions. The two-dimensional histogram (normalized per incident primary particle) shows the effects of the scattering outside the radiation field. As expected, the secondary protons scatter more with respect to the other fragments because of their smaller mass. Therefore, lighter fragments contribute more to energy deposition out-of-field. With an increase in the mass of the fragment, the energy is deposited more in a forward direction.
Figure 5

Lateral dose distribution in MeV produced by different fragment species for incident 162 MeV/u carbon ions (left) and 192 MeV/u oxygen ions (right), respectively. The results are normalized per incident primary particle

Lateral dose distribution in MeV produced by different fragment species for incident 162 MeV/u carbon ions (left) and 192 MeV/u oxygen ions (right), respectively. The results are normalized per incident primary particle Figures 6-9 show the lateral distribution of the energy deposition per incident oxygen and carbon ion with energy 192 and 162 MeV/u, respectively, including the contributions deriving from secondary protons (P) and lithium ions (Li). Li-ions have been chosen as example of heavier fragment. The comparisons have been done at different depths in the phantom as detailed in the caption of the figures. The plots are normalized per incident primary. The bottom plots indicate the ratio of the dose calculated with incident 16O and 12C beams.
Figure 6

Lateral energy deposition distribution calculated at 15-mm depth in the water phantom (plateau region). Left: total energy deposition. Middle: energy deposition deriving from secondary protons only. Right: energy deposition deriving from secondary lithium ions. The dashed lines at the – 25 mm and 25 mm axis X indicates the beam size of 5 cm

Figure 9

Lateral energy deposition distribution calculated at 62-mm depth in the water phantom (distal edge position). Left: total energy deposition. Middle: energy deposition deriving from secondary protons only. Right: energy deposition deriving from secondary lithium ions

Lateral energy deposition distribution calculated at 15-mm depth in the water phantom (plateau region). Left: total energy deposition. Middle: energy deposition deriving from secondary protons only. Right: energy deposition deriving from secondary lithium ions. The dashed lines at the – 25 mm and 25 mm axis X indicates the beam size of 5 cm Lateral energy deposition distribution calculated at 30-mm depth in the water phantom (plateau region). Left: total energy deposition. Middle: energy deposition deriving from secondary protons only. Right: energy deposition deriving from secondary lithium ions Lateral energy deposition distribution calculated at 60-mm depth in the water phantom (Bragg peak). Left: total energy deposition. Middle: energy deposition deriving from secondary protons only. Right: energy deposition deriving from secondary lithium ions Lateral energy deposition distribution calculated at 62-mm depth in the water phantom (distal edge position). Left: total energy deposition. Middle: energy deposition deriving from secondary protons only. Right: energy deposition deriving from secondary lithium ions The energy deposition of secondary protons is broader out-of-field with respect to other fragments because protons scatter more due to their smaller mass. Furthermore, the secondary proton beam generated by 16O is slightly broader than 12C at the position 60 mm and 62 mm. The relative energy deposition for 16O is higher than 12C. With an increase of mass, the heavier fragments scatter less and mostly in the forward direction of the incident beam. In the case of secondary Li-ions, the lateral dose distribution for both oxygen and carbon ion are also similar at plateau, at Bragg peak and beyond the Bragg peak. The relative energy deposition for 16O is also higher than 12C.

CONCLUSIONS

This work compared the energy deposition in a water phantom, produced by oxygen ions, proposed as alternative to carbon ions for hadron therapy. The results show that oxygen beams are characterized by a higher dose per incident ion and a slightly bigger peak-to-entrance ratio with respect to clinical carbon-ion beams. With respect to carbon ions, oxygen beams are characterized by higher production of nuclear fragments, which deposit energy beyond the Bragg peak and out-of-field. In addition, such nuclear fragments have an impact on the RBE. These factors have to be carefully taken into account when considering oxygen ions as a possible alternative to carbon therapy. This work confirms the results of Kumar et al., who investigated various types of ion species using simulation methods and concluded that oxygen ions are the most suitable heavy ions for hadron therapy in addition to carbon ions due to a higher peak-to-entrance ratio and higher energy deposition at the Bragg peak, even though it has a slightly higher entrance dose compared to carbon ion beams.[22] In the study of the dependence of the RBE on LET by Francesco et al. using research treatment planning system, the results show the RBE of oxygen was 1.68 and 2.83 at the entrance and at the Bragg peak, respectively and the RBE of carbon was 1.44 and 2.61 at the entrance and at the Bragg peak, respectively. The RBE of oxygen ions is slightly higher than carbon ions at the entrance and at the Bragg peak.[12] This may be due to the fact that oxygen ions, carrying higher charge, have comparatively higher LET than carbon ions and hence deposit higher dose from fragmentations, translating into higher RBE for oxygen ions than that for carbon ions. Therefore, oxygen beams may be more efficient than carbon ions with hypoxic tumors.[101112] Usually, LET is a key parameter to explain the biological effects induced by charged particles. However, the LET depth-dose profile is also influenced by the types of ions, and hence LET alone cannot explain the differences in their clinical properties. The RBE depends on many parameters, such as LET, fractionation, oxygenation, cell cycle, and endpoint. This study focused on the physical properties of carbon and oxygen ion beams, but biological aspects should be considered as well so that the choice of the optimal ion depends on the specific tumor configuration.[2324] Future experimental works will be carried out at the HIMAC, NIRS. Other ions lighter than carbon ions will be studied, especially helium, as helium nuclei are expected to have favorable dose distributions and a high potential for application in hadron therapy.[2526]

Financial support and sponsorship

This work was partially supported by Universiti Sains Malaysia short term grant. No. 304/CIPPT/6315065.

Conflicts of interest

There are no conflicts of interest.
  20 in total

Review 1.  Carbon ion radiotherapy: clinical experiences at National Institute of Radiological Science (NIRS).

Authors:  Tohru Okada; Tadashi Kamada; Hiroshi Tsuji; Jun-etsu Mizoe; Masayuki Baba; Shingo Kato; Shigeru Yamada; Shinji Sugahara; Shigeo Yasuda; Naoyoshi Yamamoto; Reiko Imai; Azusa Hasegawa; Hiroshi Imada; Hiroki Kiyohara; Kenichi Jingu; Makoto Shinoto; Hirohiko Tsujii
Journal:  J Radiat Res       Date:  2010-05-28       Impact factor: 2.724

2.  RBE for carbon track-segment irradiation in cell lines of differing repair capacity.

Authors:  W K Weyrather; S Ritter; M Scholz; G Kraft
Journal:  Int J Radiat Biol       Date:  1999-11       Impact factor: 2.694

3.  Irradiation System for HIMAC.

Authors:  Masami Torikoshi; Shinichi Minohara; Nobuyuki Kanematsu; Masataka Komori; Mitsukata Kanazawa; Koji Noda; Nobuyuki Miyahara; Hiroko Itoh; Masahiro Endo; Tatsuaki Kanai
Journal:  J Radiat Res       Date:  2007       Impact factor: 2.724

4.  First experimental-based characterization of oxygen ion beam depth dose distributions at the Heidelberg Ion-Beam Therapy Center.

Authors:  C Kurz; A Mairani; K Parodi
Journal:  Phys Med Biol       Date:  2012-08-07       Impact factor: 3.609

5.  Optimizing normal tissue sparing in ion therapy using calculated isoeffective dose for ion selection.

Authors:  Nicholas B Remmes; Michael G Herman; Jon J Kruse
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-03-19       Impact factor: 7.038

6.  Particle therapy and treatment of cancer.

Authors:  Edward C Halperin
Journal:  Lancet Oncol       Date:  2006-08       Impact factor: 41.316

7.  Efficacy and safety of carbon ion radiotherapy in bone and soft tissue sarcomas.

Authors:  Tadashi Kamada; Hirohiko Tsujii; Hiroshi Tsuji; Tsuyoshi Yanagi; Jun-etsu Mizoe; Tadaaki Miyamoto; Hirotoshi Kato; Shigeru Yamada; Shinroku Morita; Kyousan Yoshikawa; Susumu Kandatsu; Akio Tateishi
Journal:  J Clin Oncol       Date:  2002-11-15       Impact factor: 44.544

8.  Fragmentation of 120 and 200 MeV u-14He ions in water and PMMA targets.

Authors:  M Rovituso; C Schuy; U Weber; S Brons; M A Cortés-Giraldo; C La Tessa; E Piasetzky; D Izraeli; D Schardt; M Toppi; E Scifoni; M Krämer; M Durante
Journal:  Phys Med Biol       Date:  2017-01-23       Impact factor: 3.609

9.  Assessment of potential advantages of relevant ions for particle therapy: a model based study.

Authors:  Rebecca Grün; Thomas Friedrich; Michael Krämer; Klemens Zink; Marco Durante; Rita Engenhart-Cabillic; Michael Scholz
Journal:  Med Phys       Date:  2015-02       Impact factor: 4.071

10.  Including oxygen enhancement ratio in ion beam treatment planning: model implementation and experimental verification.

Authors:  E Scifoni; W Tinganelli; W K Weyrather; M Durante; A Maier; M Krämer
Journal:  Phys Med Biol       Date:  2013-05-16       Impact factor: 3.609

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.