Álvaro Zubizarreta-Macho1, Ana de Pedro Muñoz1, Elena Riad Deglow1, Rubén Agustín-Panadero2, Jesús Mena Álvarez1.
Abstract
: Purpose: To analyze the accuracy of two computer-aided navigation techniques to guide the performance of endodontic access cavities compared with the conventional access procedure.Entities:
Keywords: computer-assisted treatment; endodontic access cavity; endodontics; image-guided treatment; navigation system; real-time tracking
Year: 2020 PMID: 31906598 PMCID: PMC7019931 DOI: 10.3390/jcm9010129
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.241
Figure 1(A,B) Endodontic access cavities planned with the computer-aided static navigation system incorporating the cone-beam computed tomography (CBCT) scan (pink lines); (C,D) virtual template design according to planned virtual endodontic access cavities; and (E) manufactured stereolithography template fixed over dental surface of teeth placed over epoxy resin model.
Figure 2(A) Planning endodontic access cavities with treatment planning software of computer-aided dynamic navigation system (yellow cylinders), and (B) performing endodontic access cavities (green cylinders) controlled in all planes and depth with the black-and-white jaw tag fixed over the dental surface of the teeth with the thermoplastic template and the black-and-white drill tag (red arrow) attached to the high-speed handpiece (green arrow).
Descriptive deviation values at coronal (mm), apical (mm), and angular (°) levels and statistical significance (p-value).
|
| Mean | SD | Minimum | Maximum | |||
|---|---|---|---|---|---|---|---|
| Coronal | SN | 10 | 7.44 | 1.57 | 5.40 | 10.0 | SN-DN = 0.654 |
| DN | 10 | 3.14 | 0.86 | 2.00 | 5.10 | SN-MN < 0.001 | |
| MN | 10 | 4.03 | 1.93 | 1.10 | 7.10 | DN-MN < 0.001 | |
| Apical | SN | 10 | 7.13 | 1.73 | 4.80 | 9.80 | SN-DN = 0.914 |
| DN | 10 | 2.48 | 0.94 | 1.10 | 3.80 | SN-MN < 0.001 | |
| MN | 10 | 2.43 | 1.23 | 0.80 | 4.50 | DN-MN < 0.001 | |
| Angular | SN | 10 | 10.04 | 5.20 | 4.10 | 19.40 | SN-DN = 0.072 |
| DN | 10 | 5.58 | 3.23 | 1.70 | 10.40 | SN-MN < 0.001 | |
| MN | 10 | 14.95 | 11.15 | 0.80 | 29.70 | DN-MN < 0.001 |
SN, static navigation system; DN, dynamic navigation system; MN, manual navigation.
Figure 3(A,B) Planned endodontic access cavities (yellow cylinders) and access cavities performed manually (red cylinders). Guided endodontic access cavities planned and performed through (C,D) computer-aided static navigation system and (E,F) computer-aided dynamic navigation system.