Arman Sharifi1, Rahman Hallaj2,3, Soleiman Bahar1, Bahareh Babamiri1.
Abstract
This work describes an aptamer based method for highly sensitive determination of Hg(II). A Hg(II)-binding ssDNA aptamer was linked to silica-coated magnetic nanoparticles (magNPs). Then, a conjugate composed of graphene and CdS quantum dots (Gr-CdS) was linked to the complementary ssDNA. On mixing the two components, a duplex of type magNP-dsNNA-Gr/CdS is generated. If Hg(II) is added, it wills capturing the aptamer, and this leads to the release of Gr/CdS because of the formation of a stable thymine-Hg2+-thymine link. External magnetic force is used to remove the remaining complex. The released graphene-CdS is decomposed by HNO3 and injected into a graphite furnace AAS. The detectable amount of Cd is proportional to the concentration of Hg(II) in the sample. Under the optimal conditions, the method has a linear response in the 2.50 aM to 0.25 nM Hg(II) concentration range, and the detection limit is as low as 7.6 aM (at S/N = 3). It has high selectivity for Hg(II) over other metal ions. Graphical abstract.Entities:
Keywords: Aptamer; Cold vapor atomic absorption spectroscopy; Determination of Hg2 + ; Graphite furnace atomic absorption; Nanoprobe
Mesh:
Substances:
Year: 2020 PMID: 31897821 DOI: 10.1007/s00604-019-4029-x
Source DB: PubMed Journal: Mikrochim Acta ISSN: 0026-3672 Impact factor: 5.833