Tanja Grobecker-Karl1, Victor Palarie2, Sonja Schneider1, Matthias Karl1.
Abstract
Objectives: Bone density, surgical protocol, and implant design are the major determinants of primary stability. The goal of this animal trial was to investigate potential correlations of intraoperative bone density testing with clinical and histologic parameters of primary implant stability. Material and methods: Following extractions of all mandibular premolars and subsequent healing, four implants each were placed in a total of four minipigs. Bone density was determined by applying intraoperative compressive tests using a device named BoneProbe whereas measurements of implant insertion torque and resonance frequency analysis were used for evaluating implant stability. Bone mineral density (BMD) and bone to implant contact were quantified after harvesting mandibular block sections. Spearman rank correlation tests were performed for evaluating correlations (α = .05).Entities:
Keywords: bone density; dental implantation; osteotomy; torque
Mesh:
Substances:
Year: 2019 PMID: 31890296 PMCID: PMC6934340 DOI: 10.1002/cre2.224
Source DB: PubMed Journal: Clin Exp Dent Res ISSN: 2057-4347
Figure 1Creation of a 2.8‐mm osteotomy in a healed alveolar ridge (a) was followed by BoneProbe measurements depicted here with the sensing element of the BoneProbe inserted into the drill hole and gradually expanded recording the required torque as a measure of bone density (b). As a final step, a bone level implant was inserted while actively measuring the torque required (c)
Figure 2Typical microradiograph (a) and histologic section following toluidine blue staining (b) used for determining bone mineral density and bone to implant contact
Results of clinical, microradiographic, and histomorphometric measurements performed
| Mean |
| ||
|---|---|---|---|
| Clinical measurements | BoneProbe cortical | 0.68 | 0.41 |
| BoneProbe trabecular | 1.25 | 0.61 | |
| BoneProbe overall | 0.96 | 0.48 | |
| Implant insertion torque (Ncm) | 39.72 | 10.73 | |
| Implant stability (ISQ) | 69.04 | 6.24 | |
| Microradiograph | BMD (%) | 75.36 | 5.59 |
| Histomorphometry | BIC (%) | 91.03 | 2.13 |
Abbreviations: BIC, bone to implant contact; BMD, bone mineral density.
Results of Spearman rank correlation tests (Spearman's rho) for all parameters recorded in this study
| BIC | BMD | BoneProbe cortical | BoneProbe trabecular | BoneProbe overall | Implant insertion torque | Implant stability | |
|---|---|---|---|---|---|---|---|
| BIC | .53 | .23 | .24 | .17 | .41 | .15 | |
| BMD | −.09 | −.05 | −.04 | .17 | −.61 | ||
| BoneProbe cortical | .68 | .88 | .60 | .44 | |||
| BoneProbe trabecular | .91 | .60 | .21 | ||||
| BoneProbe overall | .66 | .29 | |||||
| Implant insertion torque | .31 | ||||||
| Implant stability |
Abbreviations: BIC, bone to implant contact; BMD, bone mineral density.
Results of Spearman rank correlation tests (p values) for all parameters recorded in this study
| BIC | BMD | BoneProbe cortical | BoneProbe trabecular | BoneProbe overall | Implant insertion torque | Implant stability | |
|---|---|---|---|---|---|---|---|
| BIC |
| .44 | .41 | .56 | .14 | .63 | |
| BMD | .77 | .88 | .89 | .55 |
| ||
| BoneProbe cortical |
|
|
| .13 | |||
| BoneProbe trabecular |
|
| .49 | ||||
| BoneProbe overall |
| .33 | |||||
| Implant insertion torque | .30 | ||||||
| Implant stability |
Note. Significant correlations are written in bold.
Abbreviations: BIC, bone to implant contact; BMD, bone mineral density.