Elisabeth Salomon1,2, Peter Homolka2, Istvan Csete1, Paula Toroi1.
Abstract
PURPOSE: To investigate the radiation quality dependence of the response of commercial semiconductor-based dosimeters, and to estimate potential errors and uncertainties related to different measurement and calibration scenarios.Entities:
Keywords: calibration; dosimetry; mammography; radiation quality; semiconductor dosimeter
Year: 2020 PMID: 31889315 PMCID: PMC7079095 DOI: 10.1002/mp.14005
Source DB: PubMed Journal: Med Phys ISSN: 0094-2405 Impact factor: 4.071
Figure 1Calibration setup. S: Shutter, F: Filtration, M: Monitor chamber, Red dashed line: point of test and reference plane of the dosimeters. [Color figure can be viewed at http://wileyonlinelibrary.com]
Characteristics of the radiation qualities chosen for this study.
| Radiation quality | Filtration | 1st HVL in mm Al | Source of traceability | |||
|---|---|---|---|---|---|---|
| 25 kV | 28 kV | 30 kV | 35 kV | |||
| Mo‐Mo | 0.033 mm Mo | 0.29 | 0.32 | 0.34 | 0.38 | BIPM |
| Mo‐Rh | 0.029 mm Rh | – | 0.39 | 0.41 | 0.45 | PTB |
| W‐Al | 0.5 mm Al | 0.31 | 0.35 | 0.38 | 0.44 | PTB |
| W‐Rh | 0.048 mm Rh | 0.47 | 0.50 | 0.51 | 0.55 | PTB |
| W‐Ag | 0.049 mm Ag | 0.48 | 0.53 | 0.56 | 0.61 | PTB |
Dosimeters calibrated in this study.
| Manufacturer | Detector assembly | Measuring assembly | Software |
|---|---|---|---|
| RTI | Internal detector | Piranha 657 | Ocean, Version: 2010.12.15.38 |
| RTI | R100B | Barracuda | Ocean, Version: 2010.12.15.38 |
| Unfors | Internal detector | Mult‐O‐Meter | Display read directly |
| RTI | Internal detector | Black Piranha | Ocean, Version: 2016.06.14.214 |
| PTW | Internal detector | Nomex Multimeter | Nomex Software S030008, 3.0 |
| Unfors | MAM | Xi | RaySafe Xi View, Version: 3.0 (built 47) |
| Radcal | AGMS–DM+ | Accu Gold | Accu‐Gold by Radcal, Version 1.5.3.0 Driver Version 02.10.00 |
| Unfors | MAM | X2 | RaySafe X2 View, Version 1.7.8.0 |
Mölndal, Sweden.
Billdal, Sweden.
Freiburg, Germany.
Monrovia, CA, USA.
Not recommended by the manufacturer for mammography since no automatic compensation is available.
Quantities measured and available selections of anode‐filter combinations in the software menu.
| Measuring assembly | Indications | Calibration in the software | ||||||
|---|---|---|---|---|---|---|---|---|
| Dose rate | HVL | Tube voltage | W‐Al | W‐Rh | W‐Ag | Mo‐Rh | Mo‐Mo | |
| Piranha 657 | × | – | × | × | × | × | × | × |
| Barracuda | × | – | – | – | – | – | – | × |
| Mult‐O‐Meter | × | – | – | – | – | – | – | × |
| Black Piranha | × | × | × | × | × | × | × | × |
| Nomex | × | × | × | – | × | × | × | × |
| Xi | × | × | × | × | × | × | × | × |
| Accu Gold | × | × | × | – | × | × | × | × |
| X2 | × | × | × | × | × | × | × | × |
×, selection available; –, selection not available; HVL, half value layer.
Estimated relative uncertainty in percent for of the calibration of the semiconductor dosimeters against the working standard.
| Type A | Type B | ||
|---|---|---|---|
| Step 1: Working standard | |||
| Calibration against the reference standard | 0.04 | 0.63 | |
| Chamber positioning | 0.03 | ||
| Current measurements including range and time‐based corrections of IAEA electrometer | 0.02 | 0.05 | |
| Uncertainty due to temperature measurements | 0.01 | 0.03 | |
| Uncertainty due to pressure measurements | 0.01 | 0.01 | |
| Monitor chamber contribution | 0.02 | ||
| Relative combined standard uncertainty of | 0.05 | 0.63 | |
| Step 2: Semiconductor dosimeter | |||
| Positioning | 0.03 | ||
| Reading and resolution | 0.12 | ||
| Relative combined standard uncertainty of | 0.12 | ||
| Relative combined standard uncertainty of | 0.05 | 0.65 | |
| Expanded uncertainty (k = 2) | 1.3 | ||
Includes uncorrelated part of uncertainties related to reading, leakage and short‐ term stability of the monitor chamber in both measurements with the reference and user dosimeter.
Figure 2Air kerma calibration coefficients N for all semiconductor dosimeters tested as a function of radiation quality. The horizontal red line indicates the ideal calibration coefficient of 1.0, the dashed black lines represent ±5% tolerance level. Piranha 657, Barracuda, Mult‐O‐Meter, Black Piranha, Nomex, Xi, Accu Gold, X2. [Color figure can be viewed at http://wileyonlinelibrary.com]
Maximum deviation in percent of the half value layer measurements.
| Measuring assembly | Manufacturer specification | Max deviation |
|---|---|---|
| Black Piranha | ±10% | 11% (0.04 mm Al) |
| Nomex | ±0.01 mm Al | 6.7% (0.04 mm Al) |
| Xi | ±5% | 11% (0.05 mm Al) |
| Accu Gold | ±10% (0.05 mm Al) | −1.6% (−0.01 mmAl) |
| X2 | ±5% | −3.7% (0.01 mmAl) |
Maximum deviation in percent of tube voltage measurements.
| Measuring assembly | Manufacturer specification | Max deviation |
|---|---|---|
| Piranha 657 | ±2% or ±1 kV | 10% (2.8 kV) |
| Black Piranha | ±2% or ±1 kV | 10% (3.5 kV) |
| Nomex | ±0.5 kV | 8.2% (2.4 kV) |
| Xi | ±2% or ±0.5 kV | 7.1% (2.5 kV) |
| Accu Gold | ±2% or ±0.7 kV | 8.2% (2.3 kV) |
| X2 | ±2% or ±0.5 kV | −5.9% (2.2 kV) |
Deviation in air kerma rate (error) in case no calibration or one single point calibration coefficient applied.
| Measuring assembly | Maximum deviation | |
|---|---|---|
| No calibration | Single point calibration with Mo‐Mo 28 | |
| Piranha 657 | −16% | −12% |
| Barracuda | 16% | 8.1% |
| Mult‐O‐Meter | −6.1% | −5.1% |
| Black Piranha | −4.2% | −3.2% |
| Nomex | −2.6% | −1.1% |
| Xi | −2.8% | −3.1% |
| Accu Gold | 1.1% | 1.7% |
| X2 | −2.5% | 1.9% |