Michiharu Sekimoto1, Yoh Katoh, Tsuguhisa Katoh.
Abstract
A wide variety of existing combinations of target and filter materials (target/filter combinations) are used in mammography equipment. The patient dose depends on the X-ray quality that is derived from the target/filter combination, and a calibration of the dosimeter that is used in the measurement that corresponds to the specific target/filter combination is necessary. However, dosimeters in mammography are generally calibrated with reference to the X-ray quality of Mo/Mo or W/Al combinations, and it is unclear whether or not the X-ray quality that is derived from other target/filter combinations will affect the calibration coefficients. In this paper, the calibration coefficients of different dosimeters were evaluated for target/filter combinations. For an ionization chamber-type dosimeter, good energy dependence was found and the effect of the target/filter combination was small. However, for a semiconductor dosimeter, a large energy dependence was found, and different calibration coefficients that depended on the target/filter combination used were required.Entities:
Mesh:
Year: 2015 PMID: 26699570 PMCID: PMC5691010 DOI: 10.1120/jacmp.v16i6.5604
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Value (HVL, Eeff, QI) of Beam quality with a combination of target / filter and each tube voltage. All conditions including 3 mm PMMA
|
|
|
|
|
|
|
| |
|---|---|---|---|---|---|---|---|
| W / 0.5 mm Al | HVL (mmAl) | 0.341 | 0.357 | 0.374 | 0.391 | 0.408 | 0.433 |
| Eeff (keV) | 15.701 | 16.086 | 16.441 | 16.757 | 17.049 | 17.465 | |
| QI | 0.654 | 0.619 | 0.587 | 0.559 | 0.533 | 0.499 | |
| W / 0.025 mm Rh | HVL (mmAl) | 0.341 | 0.357 | 0.374 | 0.391 | 0.408 | 0.433 |
| Eeff (keV) | 15.274 | 15.529 | 15.776 | 16.016 | 16.252 | 16.587 | |
| QI | 0.636 | 0.597 | 0.563 | 0.534 | 0.508 | 0.474 | |
| W / 0.05 mm Ag | HVL (mmAl) | 0.555 | 0.582 | 0.610 | 0.637 | 0.666 | 0.705 |
| Eeff (keV) | 18.062 | 18.366 | 18.662 | 18.949 | 19.244 | 19.627 | |
| QI | 0.753 | 0.706 | 0.666 | 0.632 | 0.601 | 0.561 | |
| W / 0.075 mm Ag | HVL (mmAl) | 0.661 | 0.685 | 0.709 | 0.733 | 0.759 | 0.792 |
| Eeff (keV) | 19.193 | 19.432 | 19.666 | 19.895 | 20.137 | 20.438 | |
| QI | 0.800 | 0.747 | 0.702 | 0.663 | 0.629 | 0.584 | |
| Mo / 0.03 mm Mo | HVL (mmAl) | 0.331 | 0.357 | 0.379 | 0.397 | 0.414 | 0.433 |
| Eeff (keV) | 15.123 | 15.516 | 15.841 | 16.097 | 16.329 | 16.590 | |
| QI | 0.630 | 0.597 | 0.566 | 0.537 | 0.510 | 0.474 | |
| Mo / 0.025 mm Rh | HVL (mmAl) | 0.354 | 0.390 | 0.412 | 0.426 | 0.440 | 0.457 |
| Eeff (keV) | 15.479 | 16.002 | 16.306 | 16.495 | 16.679 | 16.898 | |
| QI | 0.645 | 0.615 | 0.582 | 0.550 | 0.521 | 0.483 | |
List of the calibration coefficients of the ionization chamber dosimeter in the combination of the target / filter. All conditions including 3 mm PMMA
| W / 0.5 mm Al | Ramtec 1000D and N23344 | 0.999 | 0.997 | 0.996 | 0.993 | 0.995 | 0.989 |
| Model 9015 and 10x5‐6M | 1.013 | 1.012 | 1.011 | 1.011 | 1.011 | 1.010 | |
| AE‐1340C and C‐1340 | 1.010 | 1.010 | 1.010 | 1.008 | 1.008 | 1.008 | |
| W / 0.025 mm Rh | Ramtec 1000D and N23344 | 0.992 | 0.991 | 0.993 | 0.989 | 0.989 | 0.988 |
| Model 9015 and 10x5‐6M | 1.022 | 1.020 | 1.019 | 1.018 | 1.017 | 1.016 | |
| AE‐1340C and C‐1340 | 1.009 | 1.009 | 1.008 | 1.007 | 1.006 | 1.005 | |
| W / 0.05 mm Ag | Ramtec 1000D and N23344 | 0.989 | 0.985 | 0.987 | 0.987 | 0.984 | 0.980 |
| Model 9015 and 10x5‐6M | 1.004 | 1.005 | 1.003 | 1.003 | 1.001 | 1.000 | |
| AE‐1340C and C‐1340 | 0.996 | 0.998 | 0.996 | 0.996 | 0.994 | 0.993 | |
| W / 0.075 mm Ag | Ramtec 1000D and N23344 | 0.986 | 0.986 | 0.982 | 0.982 | 0.980 | 0.978 |
| Model 9015 and 10x5‐6M | 1.002 | 1.002 | 1.002 | 1.001 | 1.001 | 1.000 | |
| AE‐1340C and C‐1340 | 0.996 | 0.998 | 0.995 | 0.996 | 0.995 | 0.994 | |
| Mo / 0.03 mm Mo | Ramtec 1000D and N23344 | 1.015 | 1.012 | 1.011 | 1.010 | 1.008 | 1.004 |
| Model 9015 and 10x5‐6M | 1.018 | 1.016 | 1.014 | 1.014 | 1.013 | 1.012 | |
| AE‐1340C and C‐1340 | 0.995 | 0.995 | 0.995 | 0.994 | 0.994 | 0.994 | |
| Mo / 0.025 mm Rh | Ramtec 1000D and N23344 | 1.006 | 1.004 | 1.000 | 0.999 | 0.996 | 0.994 |
| Model 9015 and 10x5‐6M | 1.032 | 1.028 | 1.024 | 1.023 | 1.028 | 1.027 | |
| AE‐1340C and C‐1340 | 0.991 | 0.986 | 0.985 | 0.983 | 0.982 | 0.981 |
List of the calibration coefficients of the semiconductor detector in a combination of the target / filter. All conditions including 3 mm PMMA
| W / 0.5 mm Al | Eeff (keV) | 15.701 | 16.086 | 16.441 | 16.757 | 17.049 | 17.465 |
| Solidose 308 | 0.993 | 0.951 | 0.915 | 0.884 | 0.859 | 0.829 | |
| Piranha | 1.268 | 1.205 | 1.156 | 1.113 | 1.079 | 1.036 | |
| W / 0.025 mm Rh | Eeff (keV) | 15.274 | 15.529 | 15.776 | 16.016 | 16.252 | 16.587 |
| Solidose 308 | 1.001 | 0.972 | 0.952 | 0.935 | 0.918 | 0.896 | |
| Piranha | 1.360 | 1.350 | 1.340 | 1.330 | 1.326 | 1.105 | |
| W / 0.05 mm Ag | Eeff (keV) | 18.062 | 18.366 | 18.662 | 18.949 | 19.244 | 19.627 |
| Solidose 308 | 0.869 | 0.835 | 0.816 | 0.804 | 0.793 | 0.777 | |
| Piranha | 1.205 | 1.195 | 1.010 | 0.992 | 0.978 | 0.959 | |
| W / 0.075 mm Ag | Eeff (keV) | 19.193 | 19.432 | 19.666 | 19.895 | 20.137 | 20.438 |
| Solidose 308 | 0.808 | 0.779 | 0.765 | 0.754 | 0.746 | 0.734 | |
| Piranha | 1.147 | 1.146 | 0.976 | 0.964 | 0.954 | 0.941 | |
| Mo / 0.03 mm Mo | Eeff (keV) | 15.123 | 15.516 | 15.841 | 16.097 | 16.329 | 16.590 |
| Solidose 308 | 0.981 | 0.940 | 0.910 | 0.886 | 0.871 | 0.847 | |
| Piranha | 1.247 | 1.268 | 1.224 | 1.229 | 1.216 | 1.213 | |
| Mo / 0.025 mm Rh | Eeff (keV) | 15.479 | 16.002 | 16.306 | 16.495 | 16.679 | 16.898 |
| Solidose 308 | 0.948 | 0.913 | 0.889 | 0.872 | 0.859 | 0.843 | |
| Piranha | 1.281 | 1.260 | 1.244 | 1.235 | 1.228 | 1.223 |