Tianrui Zhao1, Adrien E Desjardins2,3, Sebastien Ourselin1,2, Tom Vercauteren1,2, Wenfeng Xia1,2.
Abstract
Photoacoustic imaging (PAI) is an emerging biomedical imaging modality that is based on optical absorption contrast, capable of revealing distinct spectroscopic signatures of tissue at high spatial resolution and large imaging depths. However, clinical applications of conventional non-invasive PAI systems have been restricted to examinations of tissues at depths less than a few cm due to strong light attenuation. Minimally invasive photoacoustic imaging (miPAI) has greatly extended the landscape of PAI by delivering excitation light within tissue through miniature fibre-optic probes. In the past decade, various miPAI systems have been developed with demonstrated applicability in several clinical fields. In this article, we present an overview of the current status of miPAI and our thoughts on future perspectives.Entities:
Keywords: Interventional photoacoustic imaging; Minimally invasive procedures; Multi-modal imaging; Photoacoustic computed tomography; Photoacoustic endoscopy; Photoacoustic imaging; Photoacoustic microscopy
Year: 2019 PMID: 31871889 PMCID: PMC6909166 DOI: 10.1016/j.pacs.2019.100146
Source DB: PubMed Journal: Photoacoustics ISSN: 2213-5979
Fig. 1Endogenous tissue chromophores in photoacoustic imaging. (a) Optical absorption spectra of the main tissue chromophores, including DNA (data adapted from [20]), RNA (data adapted from [22]), oxyhaemoglobin, deoxyhaemoglobin (150 g L−1), melanin (data from https://omlc.org/spectra/), water and lipid (data adapted from [18]). (b) Ultraviolet localised photoacoustic (PA) image of a fibroblast cell with lipids, protein and nucleic acids contents shown in pseudo-coloured blue, green and red, respectively. This image was adapted from Ref [21] with permission. (c) PA image of a melanoma mouse model in vivo. Melanoma (pseudo-coloured brown) visualised with 784 nm excitation was surrounded by microvasculature (pseudo-coloured red) visualised at 584 nm with six orders of vessel branching. M, Melanoma. This image was adapted from Ref. [15] with permission. (d) PA image of a human lymph node with lipid (grayscale) and haemoglobin (pseudo-coloured red) contrast with optical excitation at 1210 nm and 530 nm, respectively. This image was adapted from Ref. [26] with permission. (e) Multispectral PAI of a mouse tumour model. Oxyhaemoglobin and deoxyhaemoglobin distributions were pseudo-coloured in red and blue, respectively. White arrows indicated the region of a tumour core with a low level of blood oxygen concentration. Inset is a photograph of the corresponding cross-section of the imaged tumour region. This image was adapted from Ref. [6] with permission.
Fig. 2Spatial resolution versus depth of examinations for conventional and minimally invasive photoacoustic imaging (miPAI) techniques. OR-PAM, optical resolution-photoacoustic microscopy; AR-PAM, acoustic resolution-photoacoustic microscopy; PACT, photoacoustic computed tomography. For simplicity purposes, the spatial resolution of OR-RAM only represents the lateral resolution. With excitation light delivered within tissue through miniature fibre-optic probes, miPAI greatly extended the depths of examinations of conventional non-invasive photoacoustic imaging modalities including OR-PAM, AR-PAM and PACT.
Fig. 3Schematic illustrations of the three embodiments of minimally invasive photoacoustic imaging (miPAI) techniques including (a) Interventional photoacoustic imaging (iPAI), (b) Forward-viewing photoacoustic endoscopy (FV-PAE), and (c) Side-viewing photoacoustic endoscopy (SV-PAE).
Representative interventional photoacoustic imaging systems. L, lateral; A, axial.
| Reference | Light delivery probe | Laser source | Ultrasound system | Spatial Resolution (μm) | Targeted application | Validation |
|---|---|---|---|---|---|---|
| Piras et al. (2013) [ | 600-μm-diameter optical fibre in a biopsy needle | 1064 nm, 250 Hz (Diny pQ, IB laser) | 7.5 MHz, 128 elements (L10-5, Picus, Esaote Europe BV) | – | Breast biopsy guidance | Breast phantom with fish heart as inclusions |
| Kruizinga et al. (2014) [ | 400-μm-core-diameter, side-firing fibre in a 1.25 mm rigid steel tube | 1130 - 1250 nm (OPOTEK Vibrant B/355-II) | 8 MHz, 256 elements (ATL 12L5 + Verasonics V1) | – | Carotid artery atherosclerosis | Human carotid artery |
| Xia et al. (2015) [ | 910-μm-cores-diameter optical fibre in a needle | 750 - 900 nm, 1150 - 1300 nm, 10 Hz (VersaScan L-532, GWU-Lasertechnik) | 5 - 14 MHz, 128 elements (L14-5/38, SonixMDP, Analogic Ultrasound) | L: 600 - 1000 | Fetal surgery and nerve blocks guidance | Human placenta ex vivo + nerve and vessel phantom (porcine fat and human blood in tubes) |
| Bell et al. (2015) [ | 1-mm-core diameter optical fibre attached to a surgical tool | 1064 nm | 5-14 MHz (L14-5W/60, Ultrasonix) | – | Endonasal Surgeries | Bone and blood vessel phantom |
| Bell et al. (2015) [ | 1-mm-core diameter, side-firing optical fibre in a 2 mm quartz tube | 1064 nm, 10 Hz (Phocus InLine, Opotek) | 4 - 8 MHz linear array + 5 - 9 MHz curvilinear array (BPL9-5 + BPC8-4, Ultrasonix + SonixTouch, Ultrasonix) | – | Prostate cancer detection and treatment | Brachytherapy seeds |
| Singh et al. (2016) [ | 600-μm-diameter, side-firing optical fibre in a biopsy needle | 720 - 860 nm, 10 Hz (Quanta-Ray Pro 250, Spectra Physics, VersaScan-L532, GWU) | 7.5 MHz, 128 elements (SL3323, MyLab_One, Esaote Europe BV) | – | Prostate cancer detection and treatment | Brachytherapy seeds |
| Allard et al. (2018) [ | Fibres attached to a da Vinci® scissor tool | 1064 nm | Alpinion L3-8, Alpinion ECUBE 12R | – | Teleoperated hysterectomies | Blood vessel phantom |
| Li et al. (2018) [ | 1.5-mm-core-diameter optical fibre fused with a 3-cm-long active diffuser made of silica | 1064 nm (Q-smart 850, Quantel Laser) | 5 MHz, 128 elements (L7-4, Vantage 128, Verasonics) | L: 931 | Cancer screening and intraoperative guidance | Mouse |
Fig. 4Embodiments of interventional photoacoustic imaging (iPAI) systems and their targeted clinical applications. (a) Schematic diagram of an iPAI system for guiding breast biopsy, in which the excitation light (1064 nm) was delivered through an optical fibre embedded in a breast biopsy needle and photoacoustic (PA) signals were detected by a clinical ultrasound imaging probe. To demonstrate the concept, (b) Ultrasound (US) and (c) PA images were obtained during needle insertions towards a tumour-mimicking target (fish heart). (b)-(c) were adapted from Ref. [62] with permission. (d) Schematic diagram of an iPAI system for the diagnosis of carotid artery atherosclerosis. Excitation light was proposed to be delivered transnasally via a side-firing optical fibre to illuminate the carotid artery in the pharynx cavity, while US detection used an external linear array transducer placed at the neck side. The system was validated with an ex vivo diseased human carotid artery embedded in a neck phantom. Co-registered grayscale US (e) and colour-coded PA (f) images demonstrated complementary information, with US visualising anatomical structure and PA revealing the lipid composition of the plaque. (d)-(f) were adapted from Ref. [81] with permission. (g) Schematic diagram of a multispectral iPAI system for guiding the treatment of twin-to-twin transfusion syndrome, in which the light was delivered through the working channel of a fetoscope via an optical fibre to visualise the placenta vasculature. US detection was performed by an external linear array US probe at the abdomen. This concept was demonstrated with a freshly excised human twin placenta. An agar block was placed between the US probe and the placenta to mimic the ammonitic fluid. US images (h) revealed the anatomical structure of the placenta; a few surface blood vessels were barely visible. In contrast, PA images (i) clearly visualised two blood vessels (v1 and v2) under illumination. (h)-(i) were adapted from Ref. [63] with permission.
Representative photoacoustic endoscopy systems. FV, forward-viewing; SV, side-viewing; fps, frames per second; λ, wavelength; f-3 dB, -3 dB frequency bandwidth; OR, optical-resolution; AR, acoustic-resolution; L, lateral; A, axial; PA, photoacoustic; US, ultrasound; OCT, optical coherence tomography.
| Reference | View | Modality | Resolution mode | Light source | Ultrasound sensor | Probe diameter (mm) | Resolution (μm) | Imaging speed (fps) | Validation |
|---|---|---|---|---|---|---|---|---|---|
| Shao et al. (2012) [ | FV | PA + Fluorescence | OR | 532 nm, 160 kHz (GLP-10, IPG Photonics Corporation) | Piezoelectric, Focused, 3.5 MHz | 0.85 | L: ∼7 | 2 | Mouse ear |
| Papadopoulos et al. (2013) [ | FV | PA | OR | 532 nm, (NL-201, EKSPLA) | Piezoelectric, | 0.22 | L: 1.5 | – | Nylon wire phantom |
| Stasio et al. (2015) [ | FV | PA + fluorescence | OR | 532 nm, 200 Hz (NL-201, EKSPLA, Lithuania) | Piezoelectric, | 0.33 | L: 8 | – | Nylon thread phantom |
| Ansari et al. (2018) [ | FV | PA | AR | 410 - 2100 nm, 30 Hz (Innolas Spitlight 600) | Fabry-Perot resonator, | 3.2 | L: 31, | – | Duck embryo + Mouse skin |
| Caravaca-Aguirre et al. (2018) [ | FV | PA + fluorescence | OR | 532 nm, 7 kHz (Cobolt TorTM series) | Fabry-Perot resonator, 250 kHz - 50 MHz | 0.25 | – | – | Red blood cells + Absorbing micro-structure |
| Jansen et al. (2011) [ | SV | PA + US | AR | 715 - 1800 nm, 10 Hz (OPOTEK Vibrant B/ 355-II) | Piezoelectric, | 1.25 | – | – | Human coronary |
| Karpiouk et al. (2012) [ | SV | PA + US | AR | 1064 nm, 20 Hz (Polaris II, New Wave, Inc.) | Piezoelectric, | 2.2 | – | – | Rabit artery + Stent |
| Yang et al. (2012) [ | SV | PA + US | AR | 562 nm + 584 nm/523 nm + 640 nm, (Cobra HRR, Sirah, INNOSLAB IS811-E, EdgeWave) | Piezoelectric, | 3.8 | – | 4 | Rabbit esophagus + Rat colon |
| Dong et al. (2014) [ | SV | PA | OR | 532 nm (TLB-6712, New Focus) | Micro-ring resonator, | 4.5 | L: 4.5 | – | Plastic tube phantom |
| Abran et al. (2014) [ | SV | PA + US + Fluorescence | AR | 710 nm, 20 Hz (Quanta-Ray INDI series, Newport Corporation) | Piezoelectric, | 1.4 | – | – | Blood-mimicking phantom |
| Li et al. (2015) [ | SV | PA + US | AR | 1185-1235 nm, 1 kHz (EKSPLA, NT242) | Piezoelectric (Lead zirconate titanate composite), | 0.9 | L: ∼ 180 | 5 | Stent + Porcine aorta |
| VanderLaan et al. (2016) [ | SV | PA + US | AR | 1064 nm, 10 kHz (SOL40W-1064, RPMC Lasers, Inc.) | Piezoelectric, | 1 | – | 33 | Stent phantom |
| He et al. (2016) [ | SV | PA | OR + AR | 532 nm, 2 kHz (Wedge HB532, BrightSolutions SRL) | Piezoelectric, | 3.6 | L: 13 (OR) | – | Mouse ear |
| Dai et al. (2017) [ | SV | PA + US + OCT | OR | 710 - 1210 nm, 20 Hz | Piezoelectric, | 1 | – | – | Mouse ear + Human hand |
| Wu et al. (2017) [ | SV | PA + US | AR | 1700 - 1750 nm, 5 kHz (FQ-OPO, Elforlight Ltd) | Piezoelectric, | 1.3 | – | 20 | Swine coronary arteries |
| Hui et al. (2017) [ | SV | PA + US | AR | 1700 nm, 2 kHz | Piezoelectric, | 1 | L: 305 | 16 | Human coronary artery |
| Mathews et al. (2018) [ | SV | PA + OCT | OR | 560 - 610 nm, 2.8 kHz (Elforlight) | Fabry-Perot resonator, | 1.25 | L: 18 - 40 | – | Stent + Vascular phantom |
| Liu et al. (2018) [ | SV | PA + Hyperspectral | OR | 532 nm, 10 kHz (Model DTL-314QT) | Piezoelectric, | 12 | L: 40 | – | Rabbit rectum |
| Li et al. (2019) [ | SV | PA + US | AR | 532 nm, 300 kHz (DCH-532-10, Photonics Industries International Inc.) | Piezoelectric, | 1 | L :250 A: 50 | 50 | Rat rectum |
Fig. 5Embodiments of forward-viewing photoacoustic endoscopy (PAE) systems. (a) Schematic diagram of an optical-resolution PAE system based on a multi-core coherent fibre bundle and 2D galvanometer mirrors. Excitation light was focused at the proximal end of the fibre bundle and raster-scanned by the galvanometer mirrors for photoacoustic (PA) excitation. M, mirror; GS, glass; PD, photodiode; C1, C2, controllers; DX, DY, X and Y axis mirror drivers; OL, objective lens. (b) PA image of carbon fibre network. (c) PA image of the microvascular of a mouse ear. (a)-(c) were adapted from Ref. [65] with permission. (d) Schematic diagram of an optical-resolution PAE system based on a multimode optical fibre and a spatial light modulator (SLM). After calibration, the SLM modulated the incident light field to focus light through the multimode fibre for PA excitation. CMOS, complementary metal oxide semiconductor camera; BS1, beam splitter; L1, tube lens; OBJ, objective. (e) White light optical image and (f) PA image of a wire knot. (d)-(f) were adapted from Ref. [67] with permission. (g) Schematic diagram of an optical-resolution PAE system based on a multimode optical fibre and speckle illuminations. Pre-recorded speckle patterns were generated by a digital micro-mirror device at the distal end of the fibre for PA excitation. A model-based algorithm was used for image reconstruction. f1, tube lens; SMF, single-mode fibre; MMF, multimode fibre; CMOS, complementary metal oxide semiconductor camera. (h) Bright-field microscopy and (i) PA images of an absorbing micro-structure. Scale bar, 30 μm. (g)-(i) were adapted from Ref. [92] with permission. (j) Schematic diagram of an acoustic-resolution PAE system that is based on a multi-core coherent fibre bundle with a Fabry-Pérot (FP) cavity at its distal end to serve as a 2D array of ultrasound detectors. (k-l) PA images of the mouse abdominal skin microvasculature. (j)-(l) were adapted from Ref. [66] with permission.
Fig. 6Schematic diagrams of embodiments of side-viewing photoacoustic endoscopy (PAE) probes. (a) An acoustic-resolution PAE probe with a colinear design. A ring-shaped ultrasound (US) transducer was integrated with an optical fibre, and a mirror was placed in front of the fibre to deflect both light and US. This image was adapted from Ref. [57] with permission. (b) An acoustic-resolution PAE probe with a non-colinear design. The distal end of a multimode fibre was angle-polished at 34° for side-way illuminations, and an US transducer was placed in front of the fibre tip facing the illuminated region. AWG, Arbitrary wave generator; DAQ, data acquisition, exp, expander; lim, limiter; bpf, bandpass filter, and amp, amplifier. This image was adapted from Ref. [56] with permission. (c) An acoustic-resolution PAE probe with a compact colinear design. The distal end of a multimode fibre was polished to 47° to deflect generated photoacoustic signals to an US transducer. This image was adapted from Ref. [104] with permission. (d) An optical-resolution PAE probe with a compact colinear design and a micro-ring resonator US detector. This image was adapted from Ref. [109] with permission.
Fig. 7Multimodal endoscopy. (a-d) Tri-modal photoacoustic (PA), ultrasound (US), optical coherence tomography (OCT) imaging of a human artery. (a) PA; (b) US; (c) OCT, and (d) overlay. (a)-(d) was adapted from Ref. [116] with permission. (e) PA image of a mouse ear. (f) Co-registered PA and fluorescence image of a mouse ear at the same location with (e). (e)-(f) was adapted from Ref. [87] with permission. (g) PA image of a rabbit rectum. The blood oxygen saturation in two regions (dash boxes) are shown in corresponding hyperspectral images in (h) and (i). (g)-(i) was adapted from Ref. [118] with permission.
Fig. 8Embodiments of potential clinical applications of photoacoustic endoscopy (PAE) systems. (a-c) Cross-sectional ultrasound (US), photoacoustic (PA), and PA and US overlay images of a swine coronary artery in vivo. PAI clearly visualised a lipid core (in yellow circle). (a)-(c) was adapted from Ref. [126] with permission. (d-f) Cross-sectional US, PA, and PA and US overlay images of a vascular stent of rabbit thoracic artery in vivo. (d)-(f) was adapted from Ref. [127] with permission. (g) co-registered 3D PA (red) and US (green) image of a rat colon. The dotted arrows marked mesenteric tissue entangled around the tract. SP, sphincter. Horizontal scale bar, 1 cm; vertical scale bar, 5 mm. (h) functional image of the blood oxygen saturation level from the inside of the colon shown in (g). (g)-(h) was adapted from Ref. [57] with permission.