Abstract

STATEMENT OF PROBLEM: Digital scans have increasingly become an alternative to conventional impressions. Although previous studies have analyzed the accuracy of the available intraoral scanners (IOSs), the effect of the light scanning conditions on the accuracy of those IOS systems remains unclear.
PURPOSE: The purpose of this in vitro study was to measure the impact of lighting conditions on the accuracy (trueness and precision) of different IOSs.
MATERIAL AND METHODS: A typodont was digitized by using an extraoral scanner (L2i; Imetric) to obtain a reference standard tessellation language (STL) file. Three IOSs were evaluated-iTero Element, CEREC Omnicam, and TRIOS 3-with 4 lighting conditions-chair light 10 000 lux, room light 1003 lux, natural light 500 lux, and no light 0 lux. Ten digital scans per group were recorded. The STL file was used as a reference to measure the discrepancy between the digitized typodont and digital scans by using the MeshLab software program. The Kruskal-Wallis, 1-way ANOVA, and pairwise comparison were used to analyze the data.
RESULTS: Significant differences for trueness and precision mean values were observed across different IOSs tested with the same lighting conditions and across different lighting conditions for a given IOS. In all groups, precision mean values were higher than their trueness values, indicating low relative precision.
CONCLUSIONS: Ambient lighting conditions influenced the accuracy (trueness and precision) of the IOSs tested. The recommended lighting conditions depend on the IOS selected. For iTero Element, chair and room light conditions resulted in better accuracy mean values. For CEREC Omnicam, zero light resulted in better accuracy, and for TRIOS 3, room light resulted in better accuracy.
Copyright © 2019 Editorial Council for the Journal of Prosthetic Dentistry. Published by Elsevier Inc. All rights reserved.

Mesh:

Year:  2019        PMID: 31864638     DOI: 10.1016/j.prosdent.2019.06.003

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  14 in total

1.  Evaluation of accuracy and characteristics of tooth-color matching by intraoral scanners based on Munsell color system: an in vivo study.

Authors:  Mingming Huang; Hongqiang Ye; Hu Chen; Yongsheng Zhou; Yunsong Liu; Yong Wang; Yuchun Sun
Journal:  Odontology       Date:  2022-02-25       Impact factor: 2.885

Review 2.  The direct digital workflow in fixed implant prosthodontics: a narrative review.

Authors:  George Michelinakis; Dimitrios Apostolakis; Phophi Kamposiora; George Papavasiliou; Mutlu Özcan
Journal:  BMC Oral Health       Date:  2021-01-21       Impact factor: 2.757

3.  Effect of the volumetric dimensions of a complete arch on the accuracy of scanners.

Authors:  Min-Kyu Kim; KeunBaDa Son; Beom-Young Yu; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2020-12-28       Impact factor: 1.904

4.  Full-arch accuracy of five intraoral scanners: In vivo analysis of trueness and precision.

Authors:  Miran Kwon; Youngmok Cho; Dong-Wook Kim; MyungSu Kim; Yoon-Ji Kim; Minho Chang
Journal:  Korean J Orthod       Date:  2021-03-25       Impact factor: 1.372

5.  Analysis of the relationship between the surface topography of prepared tooth surfaces and data quality of digital impressions from an intraoral scanner.

Authors:  Neset Volkan Asar; Sarah Yun; Shelby Schwartz; Ilser Turkyilmaz
Journal:  J Dent Sci       Date:  2021-07-31       Impact factor: 2.080

6.  In-vitro evaluation of marginal and internal fit of 3-unit monolithic zirconia restorations fabricated using digital scanning technologies.

Authors:  Çise Özal; Mutahhar Ulusoy
Journal:  J Adv Prosthodont       Date:  2021-12-22       Impact factor: 1.904

7.  The Effects of Orthodontic Brackets on the Time and Accuracy of Digital Impression Taking.

Authors:  Hyojin Heo; Minji Kim
Journal:  Int J Environ Res Public Health       Date:  2021-05-16       Impact factor: 3.390

8.  Can Dental Office Lighting Intensity Conditions Influence the Accuracy of Intraoral Scanning?

Authors:  Anca Jivanescu; Andrei-Bogdan Faur; Raul Nicolae Rotar
Journal:  Scanning       Date:  2021-05-27       Impact factor: 1.932

9.  Marginal and internal fit and intaglio surface trueness of interim crowns fabricated from tooth preparation of four finish line locations.

Authors:  Keunbada Son; Young-Tak Son; Ji-Min Lee; Kyu-Bok Lee
Journal:  Sci Rep       Date:  2021-07-06       Impact factor: 4.379

10.  Effect of Tooth Types on the Accuracy of Dental 3D Scanners: An In Vitro Study.

Authors:  Keunbada Son; Kyu-Bok Lee
Journal:  Materials (Basel)       Date:  2020-04-09       Impact factor: 3.623

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.