| Literature DB >> 31854210 |
Lei Xu1, Hao Li2, Sishi Yang1, Wenjuan Zeng1, Shan Gan2, Xin Chen2, Ling Duan2, Hongbing Hu2.
Abstract
BACKGROUND: The indirect antiglobulin test (IAT) and direct antiglobulin test (DAT) have been used as common tests for transfusion. Recently, we have found that in addition to causing false increases, rheumatoid factor (RF) can also cause false decreases in immunoassays for hepatitis B surface antigen and B-type natriuretic peptide. However, it remains unclear whether RF also interferes with the IAT and DAT.Entities:
Keywords: Indirect antiglobulin test; direct antiglobulin test; immunoassay; interference; red blood cell; rheumatoid factor
Mesh:
Substances:
Year: 2019 PMID: 31854210 PMCID: PMC7782952 DOI: 10.1177/0300060519892386
Source DB: PubMed Journal: J Int Med Res ISSN: 0300-0605 Impact factor: 1.671
Ratios of the RF and IgG concentrations in one-step IAT models and two-step IAT models.
| Models | Decrease group | Increase group | Consistent group |
|---|---|---|---|
| One-step IAT# | 106.9 (70.5, 174.2) | 112.5 (96.1, 233.3) | 120.8 (85.1, 178.1) |
| Two-step IAT§ | 166.1 (114.2, 232.0) | 180.2 (106.8, 246.2) | 130.7 (96.0, 188.5) |
Data are presented as medians (interquartile ranges). #H=3.0, p=0.23; there were no significant differences in the decrease group, increase group or consistent group. §H=2.1, p=0.36; there were no significant differences in the decrease group, increase group or consistent group.
IAT, indirect antiglobulin test
RF concentrations of 219 IAT models.
| Groups | N | RF concentrations (IU/mL) |
|---|---|---|
| Decline ≥2 levels of IAT positivity | 20 | 1080 (987, 1550) |
| Decline <2 levels of IAT positivity | 49 | 1710 (1145, 2647) |
| Uniformity | 150 | 1415 (1115, 2325) |
Data are presented as medians (interquartile ranges).
N: The number of IAT models.
RF, rheumatoid factor; IAT, indirect antiglobulin test
Rate of decrease and increase of the DAT models after dilution.
| Number of decreases | Number of increases | Number of models | |
|---|---|---|---|
| IAT models | 69 (69/219, 31.5%) | 16 (16/219, 7.3%) | 219 |
| DAT models after dilution | 46 (46/60, 76.7%)[ | 10 (10/60, 16.7%)§ | 60 |
Data are presented as N (%).
#x2 = 39.6, p<0.0005; §x2 = 4.9, p = 0.027 compared with IAT models.
DAT, direct antiglobulin test; IAT, indirect antiglobulin test
Figure 1.After dilution, the results of control models did not change. As the relative content of RF increased, the degree of decrease in the 17 DAT models became increasingly significant compared with control models. DAT, direct antiglobulin test; RF, rheumatoid factor.
X: DAT-positive sample; R: RF-positive plasma sample.
Figure 2.After dilution, the results of control models did not change, but the results of 8 DAT models changed with an inversely proportional trend. The results for X172+R133, X165+R133 and X133+R109 were higher in HCT samples than control models, and then became lower than the control models after dilution. The other 5 models showed a trend of initially decreasing, then increasing and finally dropping at dilutions of 1:16. DAT, direct antiglobulin test; HCT, haematocrit.
X: DAT-positive sample; R: RF-positive plasma sample.