Literature DB >> 23537796

False decrease of HBsAg S/CO values in serum with high-concentration rheumatoid factors.

Lei Xu1, Xueping Wang, Ronghong Ma, Xiaoqian Zeng, Bing Wu, Lihua Hu, Yirong Li.   

Abstract

OBJECTIVES: To investigate whether high-concentration RFs cause false decrease of S/CO values of serum HBsAg. DESIGN AND METHODS: Serum HBsAg was determined in 100 RF-positive sera using one-step ELISA. Twenty-three HBsAg-negative sera with high-concentration RFs were selected randomly to perform dilution with mixed normal sera. Serum models consisting of HBsAg and high-concentration RFs were made by blending high-concentration RFs and HBsAg-positive sera at the ratio of 9:1.
RESULTS: In the 23 samples, one-step ELISA showed that HBsAg-positive rate was 69.57% at dilution of 1:2, 60.87% at dilution of 1:4 and 40.00% at dilution of 1:8, whereas two-step ELISA showed that it was 100% without any dilution. Thirty serum models were made and divided into six groups. Median S/CO value of HBsAg was 3.00 in control, whereas it ranged from 0.13 to 1.78 in the six groups.
CONCLUSIONS: High-concentration RFs cause false decrease of S/CO values of HBsAg using one-step ELISA.
Copyright © 2013 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2013        PMID: 23537796     DOI: 10.1016/j.clinbiochem.2013.03.013

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  4 in total

Review 1.  Infection prevention and the medical director: uncharted territory.

Authors:  Toros Kapoian; Klemens B Meyer; Douglas S Johnson
Journal:  Clin J Am Soc Nephrol       Date:  2015-02-20       Impact factor: 8.237

2.  Negative interference by rheumatoid factor of plasma B-type natriuretic peptide in chemiluminescent microparticle immunoassays.

Authors:  Wen Fan; Lei Xu; Liangcai Xie; Decai Yang; Xuezheng Liu; Jiajun Zhang; Yirong Li; Cunjian Yi
Journal:  PLoS One       Date:  2014-08-21       Impact factor: 3.240

3.  Interference in the indirect antiglobulin test and direct antiglobulin test from rheumatoid factor.

Authors:  Lei Xu; Hao Li; Sishi Yang; Wenjuan Zeng; Shan Gan; Xin Chen; Ling Duan; Hongbing Hu
Journal:  J Int Med Res       Date:  2019-12-19       Impact factor: 1.671

4.  Negative interference in serum HBsAg ELISA from rheumatoid factors.

Authors:  Lei Xu; Zhen Yu; Wen Fan; Xueping Wang; Mingshui Xie; Yiting Xu; Lihua Hu; Yirong Li
Journal:  PLoS One       Date:  2013-11-15       Impact factor: 3.240

  4 in total