Torque property of titanium alloy cerebral aneurysm clips in a magnetic resonance scanner.

Minghui Tang1, Shingo Kawahira2, Naoyuki Nomura3, Toru Yamamoto4.   

Abstract

Titanium (Ti) alloys have been introduced in magnetic resonance (MR) safe implantable medical devices because the susceptibility of Ti is approximately 1/10 that of the Co-Cr-Ni alloy (Elgiloy), which was the previously preferred MR-safe material. The torque applied to metallic materials in an MR imaging (MRI) scanner is commonly believed to increase with the susceptibility of the material. However, a visual inspection showed that the torque applied to Ti alloy cerebral aneurysm clips is comparable with that in the case of those of Elgiloy. In this study, we measured the torque applied to the small test pieces of rods and aneurysm clips quantitatively in a 3-T MRI using an accurate self-developed torque measurement apparatus. The maximum torques of Ti alloy and Elgiloy rod test pieces were comparable as 1.1 and 1.2 µN·m, respectively. The values for Ti alloy aneurysm clips were distinctly higher than the values for those of Elgiloy. These contradictory results of a larger torque for smaller-susceptibility products could be explained by our new theory, which takes into account the crystal susceptibility anisotropy in addition to the conventional torque due to the shape anisotropy.

Entities:  

Mesh:

Substances:

Year:  2019        PMID: 31838592     DOI: 10.1007/s10856-019-6329-4

Source DB:  PubMed          Journal:  J Mater Sci Mater Med        ISSN: 0957-4530            Impact factor:   3.896


  17 in total

1.  Magnetic properties of Ni-Co-Cr-base Elgiloy.

Authors:  J C Ho; F G Shellock
Journal:  J Mater Sci Mater Med       Date:  1999-09       Impact factor: 3.896

Review 2.  Testing MR safety and compatibility: an overview of the methods and current standards.

Authors:  Gregory Schaefers
Journal:  IEEE Eng Med Biol Mag       Date:  2008 May-Jun

3.  Effects of phase constitution on magnetic susceptibility and mechanical properties of Zr-rich Zr-Mo alloys.

Authors:  Ryota Kondo; Yusuke Tsutsumi; Hisashi Doi; Naoyuki Nomura; Takao Hanawa
Journal:  Acta Biomater       Date:  2011-07-19       Impact factor: 8.947

4.  3-T MRI safety assessments of magnetic dental attachments and castable magnetic alloys.

Authors:  M Hasegawa; K Miyata; Y Abe; T Ishii; T Ishigami; K Ohtani; E Nagai; T Ohyama; Y Umekawa; S Nakabayashi
Journal:  Dentomaxillofac Radiol       Date:  2015-03-18       Impact factor: 2.419

5.  Cerebral aneurysm clips in the 3-tesla magnetic field. Laboratory investigation.

Authors:  Yukinari Kakizawa; Tatsuya Seguchi; Tetsuyoshi Horiuchi; Kazuhiro Hongo
Journal:  J Neurosurg       Date:  2010-10       Impact factor: 5.115

6.  Magnetic properties of materials for MR engineering, micro-MR and beyond.

Authors:  Matthias C Wapler; Jochen Leupold; Iulius Dragonu; Dominik von Elverfeld; Maxim Zaitsev; Ulrike Wallrabe
Journal:  J Magn Reson       Date:  2014-03-15       Impact factor: 2.229

7.  A Web-based searchable system to confirm magnetic resonance compatibility of implantable medical devices in Japan: a preliminary study.

Authors:  Yasuhiro Fujiwara; Hitoshi Fujioka; Tomoko Watanabe; Maiko Sekiguchi; Ryuji Murakami
Journal:  Radiol Phys Technol       Date:  2017-07-31

Review 8.  Metal-induced artifacts in MRI.

Authors:  Brian A Hargreaves; Pauline W Worters; Kim Butts Pauly; John M Pauly; Kevin M Koch; Garry E Gold
Journal:  AJR Am J Roentgenol       Date:  2011-09       Impact factor: 3.959

9.  Metal artefact reduction in MRI at both 1.5 and 3.0 T using slice encoding for metal artefact correction and view angle tilting.

Authors:  M Reichert; T Ai; J N Morelli; M Nittka; U Attenberger; V M Runge
Journal:  Br J Radiol       Date:  2015-01-23       Impact factor: 3.039

10.  Titanium hip implants for improved magnetic resonance and computed tomography examinations.

Authors:  N A Ebraheim; E R Savolaine; J Zeiss; W T Jackson
Journal:  Clin Orthop Relat Res       Date:  1992-02       Impact factor: 4.176

View more
  1 in total

1.  A Novel Magnetic Resonance Imaging-Compatible Titanium Alloy Wire-Reinforced Endotracheal Tube.

Authors:  Bikei Ryu; Yoshikazu Okada; Nobuko Fujita; Yasuko Nagasaka
Journal:  Materials (Basel)       Date:  2022-08-16       Impact factor: 3.748

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.