Time-related study on external exposure dose of 2-deoxy-2-[F-18]fluoro-D-glucose PET for workers' safety.

Yasuyuki Takahashi1, Shota Hosokawa2, Takakiyo Tsujiguchi2, Satoru Monzen2, Takao Kanzaki2,3, Koji Shirakawa4, Ayaka Nemoto5, Hayato Ishimura6, Noboru Oriuchi7.   

Abstract

Time-course study of individual dose equivalents of 2-deoxy-2-[F-18]fluoro-D-glucose positron emission tomography (18F-FDG PET) was conducted in different hospital workers, and the daily work duties were analyzed. For the measurements, a semiconductor dosimeter was used. The values at intervals of 1 min and 1 h, the monthly cumulative and daily cumulative doses, and trend graphs were acquired with dedicated software and displayed on the reader. The following radiation workers with duties involving maximum external exposure work were included: doctors making diagnoses (4.8 μSv/procedure), nurses removing injection needles (3.1 μSv/procedure), pharmacists performing quality control tests (2.9 μSv/procedure), nuclear medicine technologists assisting patient positioning (6.5 μSv/procedure), and cyclotron engineers performing daily checks (13.4 μSv/procedure). The results of analysis of daily work duties revealed the influencing factors of external exposure dose. To reduce the external exposure dose, investigators should shorten the patient's contact time with the 18F-FDG source or patient tracer.

Entities:  

Keywords:  18F-FDG PET; Radiation workers; Si solid-state detector

Mesh:

Substances:

Year:  2019        PMID: 31832908     DOI: 10.1007/s12194-019-00548-0

Source DB:  PubMed          Journal:  Radiol Phys Technol        ISSN: 1865-0333


  11 in total

1.  Efficiency of a radiophotoluminescence glass dosemeter for low-earth-orbit space radiation.

Authors:  H Yasuda; K Fujitaka
Journal:  Radiat Prot Dosimetry       Date:  2002       Impact factor: 0.972

2.  The introduction of automated dispensing and injection during PET procedures: a step in the optimisation of extremity doses and whole-body doses of nuclear medicine staff.

Authors:  P Covens; D Berus; F Vanhavere; V Caveliers
Journal:  Radiat Prot Dosimetry       Date:  2010-03-23       Impact factor: 0.972

3.  The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103.

Authors: 
Journal:  Ann ICRP       Date:  2007

4.  INVESTIGATION OF EXTERNAL RADIATION DOSES DURING RESIDENTS' TEMPORAL STAY TO NAMIE TOWN, FUKUSHIMA PREFECTURE.

Authors:  T Tsujiguchi; Y Shiroma; T Suzuki; Y Tamakuma; M Yamaguchi; K Iwaoka; M Hosoda; S Tokonami; I Kashiwakura
Journal:  Radiat Prot Dosimetry       Date:  2019-10-01       Impact factor: 0.972

5.  Radiation dose rates from patients undergoing PET: implications for technologists and waiting areas.

Authors:  N A Benatar; B F Cronin; M J O'Doherty
Journal:  Eur J Nucl Med       Date:  2000-05

6.  Scintillation crystals for PET.

Authors:  C L Melcher
Journal:  J Nucl Med       Date:  2000-06       Impact factor: 10.057

7.  Radiation dose to technicians per nuclear medicine procedure: comparison between technetium-99m, gallium-67, and iodine-131 radiotracers and fluorine-18 fluorodeoxyglucose.

Authors:  C Chiesa; V De Sanctis; F Crippa; M Schiavini; C E Fraigola; A Bogni; C Pascali; D Decise; R Marchesini; E Bombardieri
Journal:  Eur J Nucl Med       Date:  1997-11

8.  [Measurement of radiation exposure to a PET institution driver from patients injected with FDG].

Authors:  Fumitoshi Nakamura; Toshihiko Kanno; Hiroyuki Okada; Etsuji Yoshikawa; Ichirou Andou; Masami Futatsubashi; Tomomi Shinke; Yasuomi Ouchi; Tatsuo Torizuka
Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi       Date:  2006-08-20

9.  Radiation dose to technologists per nuclear medicine examination and estimation of annual dose.

Authors:  Tuncay Bayram; A Hakan Yilmaz; Mustafa Demir; Bircan Sonmez
Journal:  J Nucl Med Technol       Date:  2011-02-14

10.  The radiation dose to accompanying nurses, relatives and other patients in a nuclear medicine department waiting room.

Authors:  L K Harding; N J Harding; H Warren; A Mills; W H Thomson
Journal:  Nucl Med Commun       Date:  1990-01       Impact factor: 1.690

View more
  1 in total

1.  Exposure Doses to Technologists Working in 7 PET/CT Departments.

Authors:  Weiguo Li; Lianying Fang; Jieqing Li
Journal:  Dose Response       Date:  2020-07-09       Impact factor: 2.658

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.