| Literature DB >> 31832739 |
Jonathan Nicholls1, Abdulrahman M Alfuraih2,3,4, Elizabeth M A Hensor2,3, Philip Robinson5,6.
Abstract
OBJECTIVE: To determine inter- and intra-reader reproducibility of shear wave elastography measurements for musculoskeletal soft tissue masses.Entities:
Keywords: Elastography; Medical imaging; Muscles; Reliability; Ultrasound
Mesh:
Year: 2019 PMID: 31832739 PMCID: PMC7083807 DOI: 10.1007/s00256-019-03300-2
Source DB: PubMed Journal: Skeletal Radiol ISSN: 0364-2348 Impact factor: 2.199
Fig. 1Example of reader defined region of interest modifiable square box located in most superficial aspect of lesion that appeared solid – myxoid liposarcoma
Shear wave measurements summarised by reader
| Shear wave measure | Reader 1 | Reader 2 |
|---|---|---|
| Transverse velocity, m/s | 2.11 (2.01, 2.21) | 2.11 (2.01, 2.21) |
| Transverse stiffness, kPa | 15.59 (14.24, 17.07) | 15.46 (14.17, 16.88) |
| Cranio-caudal velocity, m/s | 2.14 (2.04, 2.23) | 2.02 (1.93, 2.12) |
| Cranio-caudal stiffness, kPa | 15.97 (14.67, 17.39) | 14.34 (13.14, 15.65) |
All values presented as geometric mean (95% confidence interval)
Repeatability coefficients for each reader and 95% limits of agreement (LOA) between them, adjusted for repeated measures
| Shear wave measure | Coefficient of repeatability | Mean differencea (95% LOA) | Mean ratiob (95% LOA) | |
|---|---|---|---|---|
| Reader 1 | Reader 2 | |||
| Transverse velocity, m/s | 0.34 | 0.40 | −0.01 (−0.45, 0.44) | 0.99 (0.64, 1.55) |
| Transverse stiffness, kPa | 0.68 | 0.82 | 0.00 (−0.89, 0.89) | 1.00 (0.41, 2.44) |
| Cranio-caudal velocity, m/s | 0.35 | 0.38 | 0.05 (−0.48, 0.57) | 1.05 (0.62, 1.78) |
| Cranio-caudal stiffness, kPa | 0.73 | 0.76 | 0.10 (−0.99, 1.19) | 1.10 (0.37, 3.28) |
LOA limits of agreement
aNatural log scale
bBack-transformed from log scale therefore reported values are ratios between reader 1 and reader 2
Fig. 2Bland-Altman plots of shear wave measurements on log scale: a transverse velocity (m/s); b transverse stiffness (kPa); c cranio-caudal velocity (m/s); d cranio-caudal stiffness (kPa)
Limits of agreement (LOA) between the two readers calculated using the means of the five repeated measurements
| Shear wave measure | Mean differencea (95% LOA) | Mean ratiob (95% LOA) |
|---|---|---|
| Transverse velocity, m/s | 0.00 (−0.30, 0.30) | 1.00 (0.74, 1.35) |
| Transverse stiffness, kPa | 0.01 (−0.58, 0.60) | 1.01 (0.56, 1.82) |
| Cranio-caudal velocity, m/s | 0.05 (−0.36, 0.47) | 1.06 (0.69, 1.60) |
| Cranio-caudal stiffness, kPa | 0.11 (−0.76, 0.98) | 1.11 (0.47, 2.65) |
aNatural log scale
bBack-transformed from log scale therefore reported values are ratios between reader 1 and reader 2
Fixed inter- and intra-reader intraclass correlation coefficients for each shear wave measure
| Shear wave measure | Intraclass correlation coefficient (95% CIa) | ||
|---|---|---|---|
| Inter-reader | Intra-reader 1 | Intra-reader 2 | |
| Transverse velocity, m/s | 0.857 (0.811, 1) | 0.887 (0.846, 1) | 0.917 (0.886, 1) |
| Transverse stiffness, kPa | 0.843 (0.795, 1) | 0.872 (0.825, 1) | 0.909 (0.875, 1) |
| Cranio-caudal velocity, m/s | 0.786 (0.716, 1) | 0.895 (0.851, 1) | 0.905 (0.866, 1) |
| Cranio-caudal stiffness, kPa | 0.752 (0.673, 1) | 0.882 (0.831, 1) | 0.890 (0.843, 1) |
aOne-sided lower limit confidence interval (CI)
Fig. 3Association between measured lesion depth (cm; log transformed) and transverse shear wave velocity (m/s; log transformed)