| Literature DB >> 31831771 |
N Tokhadze1, P Chennell2, L Bernard1, C Lambert3, B Pereira3, B Mailhot-Jensen4, V Sautou1.
Abstract
Medical tubings in plasticized polyvinylchloride (PVC) are widely used for the infusion of medications but are known in some cases to cause content-container interactions (drug sorption and plasticizer release). The aim of this study was to assess interactions between drugs and five alternative materials to a reference plasticized PVC intravenous (IV) infusion tubing: three were PVC coextruded with polyethylene (PE), polyurethane (PU) or a thermoplastic elastomer (Styrene-EthyleneButadiene-Styrene (SEBS)) and two were SEBS or thermoplastic olefin (TPO) monolayer tubings. Diazepam and insulin were chosen as respective reference of absorption and adsorption while paracetamol acted as a negative control. The concentration of each drug was quantified with liquid chromatography to evaluate a potential loss after a static contact condition and simulated infusion at 1 mL/h and 10 mL/h dynamic condition by an electric syringe pump. A characterization of each material's surface was performed by Fourier transform infrared spectroscopy in attenuated total reflection mode (ATR-FTIR) and by measurement of surface zeta potential. Plasticizer release was quantified by gas chromatography coupled with mass spectrometry (GC-MS). For all tubings except PVC/PU, no loss of paracetamol was observed in any condition. Diazepam sorption appeared to be less important with PVC/PE, PVC/SEBS, SEBS and TPO tubings than with PVC, but was more important when using PVC/PU tubings. PVC tubings induced the least loss of insulin amongst all the studied materials. Surface analysis by ATR-FTIR highlighted the presence of a plasticizer (that could be attributed to Tris (2-Ethylhexyl) Trimellitate (TOTM)) in the coextruded SEBS layer of PVC/SEBS, which could have influenced drug sorption, probably as a consequence of a migration from the PVC layer. Coextruded PVC/SEBS and PVC/PE presented the lowest zeta potential of all studied materials with respective values of -39 mV and -36 mV and were related to the highest sorption of insulin while PVC/PU with the highest zeta potential (about -9 mV) presented the highest absorption of diazepam. Coextruded layered materials appeared to have a lower plasticizer release than PVC alone. As a conclusion, PVC/PE and thermoplastic elastomers alone or coextruded with PVC could be interesting alternatives to PVC tubings with regards to sorption phenomena and plasticizer release.Entities:
Year: 2019 PMID: 31831771 PMCID: PMC6908714 DOI: 10.1038/s41598-019-55113-x
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Description of IV-tubings (PVC: Polyvinylchloride; PE: Polyethylene; PU: Polyurethane; SEBS: Styrene-Ethylenebutadiene-Styrene; TPO: Thermoplastic olefin)
| Inner surface material | PVC | PE | PU | SEBS | SEBS* | TPO* |
|---|---|---|---|---|---|---|
| Coextrusion with | / | PVC | PVC | PVC | / | / |
Reference Batch/Grade Manufacturer | PN10318-1 118D06-TE Cair LGL, France | RPB5320 16F27 Cair LGL, France | PY2301NCM 17F06-T Cair LGL, France | PT050117 17B03-TE Cair LGL, France | Cawiton tubing grade PR13816 Wittenburg, The Netherlands | Cawiton tubing grade PR13997 Wittenburg, The Netherlands |
| Length (cm) | 200 | 200 | 10 | 150 | 200 | 200 |
| Internal/external Diameter (mm) | 2.5/4.1 | 2.5/4.1 | 2.5/4.1 | 2.5/4.1 | 2.5/4.1 | 2.5/4.1 |
| Inner Surface Area (cm²) | 157.08 | 157.08 | 7.85 | 117.81 | 157.08 | 157.08 |
*SEBS and TPO tubings were of medical grade but not obtained from commercialized extension sets.
Figure 1Picture of the experimental setup in dynamic condition (A) electric syringe pump; (B) 50 mL syringe; (C) infusion tubing; (D) withdrawing site at the tip of the tube).
Chromatography methods used for quantification of paracetamol, diazepam and insulin. TFA: trifluoroacetic acid.
| API | Chromatography column | Mobile Phase | Mode | Flow rate | Oven Temperature (°C) | Detection wavelength (nm) | Injection volume (µL) |
|---|---|---|---|---|---|---|---|
| Paracetamol | Gemini C18 3 µm, 150 × 4,6 mm Precolumn Gemini C18 5 µm, 0,04 × 0,03 cm (Phenomenex, France) | Water adjusted to pH 2.75 with formic acid 98% Acetonitrile | Gradient (% Phase A): 0–3 min: 90% 3–7 min: 90% to 40% 7–10 min: 40% | 1.3 | 40 | 243 | 10 |
| Diazepam | Nucleodur C18 HTEC 5 µm, 125 × 4,6 mm (Macherey-Nagel, France) | v/v/v 22% Acetonitrile 34% Methanol 44% phosphate buffer (3.4 g/L, pH = 5.0) | isocratic | 1.0 | 30 | 254 | 20 |
| Insulin | Nucleodur C18 ec 5 µm, 250 × 4,6 mm (Macherey Nagel, France) | TFA/water 0,1% (v/v) TFA/acetonitrile 0,08% (v/v) | Gradient (% Phase A): 0–15 min: 80% to 35% 15–17 min: 35% 17–20 min: 35% to 80% | 1.3 | 25 | 280 | 20 |
Analytical method validation data. CV: coefficient of variation.
| API | Calibration curve | Slope of the regression line | Intercept of the regression line | Coefficient of determination (r²) | Mean repeatability CV (%) | Mean intermediate precision (%) | Relative mean trueness bias (%) |
|---|---|---|---|---|---|---|---|
| Paracetamol | 60–140 µg/mL | 28742342.0 | −40415.1 | 0.996 | 0.50 | 4.70 | 0.58 |
| Diazepam | 1–40 µg/mL | 64074.1 | −13506.0 | 0.986 | 2.80 | 3.90 | 0.27 |
| Insulin | 0.06–0.14 U/mL | 27308.3 | −1991.1 | 0.984 | 6.40 | 6.40 | 1.04 |
Figure 2Evolution of paracetamol concentrations compared to initial concentration in static condition (A); 1 mL/h dynamic (B) and 10 mL/h dynamic condition (C) for every studied tubing (n = 3, mean ± standard error of mean). (PVC: Polyvinylchloride; PE: Polyethylene; PU: Polyurethane; SEBS: Styrene-Ethylenebutadiene-Styrene; TPO: Thermoplastic olefin).
Figure 3Evolution of diazepam concentrations compared to initial concentration in static condition (A); 1 mL/h dynamic (B) and 10 mL/h dynamic condition (C) for every studied tubing (n = 3, mean ± standard error of mean). (PVC: Polyvinylchloride; PE: Polyethylene; PU: Polyurethane; SEBS: Styrene-Ethylenebutadiene-Styrene; TPO: Thermoplastic olefin).
Figure 4Evolution of insulin concentration compared to initial concentration in static condition (A); 1 mL/h dynamic (B) and 10 mL/h dynamic condition (C) for every studied tubings (n = 3, mean ± standard error of mean). (PVC: Polyvinylchloride; PE: Polyethylene; PU: Polyurethane; SEBS: Styrene-Ethylenebutadiene-Styrene; TPO: Thermoplastic olefin).
Figure 5Effect size of the comparison of each material to PVC after an 8 h infusion at 1 ml/h and 10 ml/h. (A) Paracetamol; (B) Diazepam; (C) Insulin (mean ± confidence interval of 95%). (PVC: Polyvinylchloride; PE: Polyethylene; PU: Polyurethane; SEBS: Styrene-Ethylenebutadiene-Styrene; TPO: Thermoplastic olefin).
Figure 6FTIR spectra of the inner surface before infusion of SEBS (A) and PVC/SEBS (B) tubings compared to TOTM FTIR spectrum (C) (PVC: Polyvinylchloride; SEBS: Styrene-Ethylenebutadiene-Styrene).
zeta potential (mV) of the inner surface of every tubings at pH = 5.0 (PVC: Polyvinylchloride; PE: Polyethylene; PU: Polyurethane; SEBS: Styrene-Ethylenebutadiene-Styrene; TPO: Thermoplastic olefin).
| class | PVC | PVC/PE | PVC/PU | PVC/SEBS | SEBS | TPO |
|---|---|---|---|---|---|---|
| thermoplastic | thermoplastic | thermoplastic | thermoplastic elastomer | thermoplastic elastomer | thermoplastic elastomer | |
| co-extruded | no | PVC | PVC | PVC | no | no |
| pH | 5.06 | 4.98 | 5.06 | 5.04 | 5.00 | 5.04 |
| zeta potential (mv) | −27.4 | −37.0 | −9.5 | −39.6 | −33.1 | −30.1 |
Quantification of plasticizer (Tris(2-ethylheyl) trimellitate: TOTM) in Polyvinyl Chloride (PVC) and PVC coextruded tubings and plasticizer migration assay (n = 3, mean ± standard error of mean).
| Infusion tubings | Percentage of TOTM in PVC (% w/w) | Quantity of TOTM released per cm² of tubing after 24 h (µg/cm²) |
|---|---|---|
| PVC | 48.04 ± 7.39 | 11.45 ± 0.09 |
| PVC/PE | 50.35 ± 1.92 | 3.17 ± 0.01 |
| PVC/PU | 46.14 ± 1.90 | 7.57 ± 0.43 |
| PVC/SEBS | 42.41 ± 2.69 | 5.39 ± 0.15 |