Aurélie Maiguy-Foinard1,2, Nicolas Blanchemain3,4, Christine Barthélémy1, Bertrand Décaudin1,2, Pascal Odou1,2.
Abstract
PURPOSE: Plastic materials such as polyurethane (PUR), polyethylene (PE), polypropylene (PP) and polyvinyl chloride (PVC) are widely used in double-lumen extension tubing. The purposes of our study were to 1) compare in vitro drug delivery through the double extension tubes available on the market 2) assess the plastic properties of PUR in infusion devices and their impact on drug delivery.Entities:
Mesh:
Substances:
Year: 2016 PMID: 27153224 PMCID: PMC4859544 DOI: 10.1371/journal.pone.0154917
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Double-lumen extension tubes made from different types of plastics (PUR, PVC and PE) assessed in our study.
1) Octopus 2 ref. 841.264. 2) Octopus 2 ref. 5841.208. 3) Spider ref. PY2101NCM. 4) Smallbore ref. 011-MC33076. 5) Smallbore ref. 011-MC33077. 6) Smallbore ref. 011-MC33165. 7) MaxPlus Clear ref. MFX2502MP. 8) Edelvaiss CW2+.
Characteristics of double-lumen extension tubes used for our study.
| Double-lumen extension tube | Manufacturer | Reference | Type of plastic | Characteristics | Internal surface (mm2) | Type of sterilisation |
|---|---|---|---|---|---|---|
| Octopus 2 | Vygon | 841.264 | PUR | Ø = 1.50 x 2.50 mm, L = 10 cm, Vol. = 0.34 mL | 471 | Gamma irradiation |
| Octopus 2 | Vygon | 5841.208 | PUR | Ø = 1.50 x 2.50 mm, L = 10 cm, Vol. = 0.44 mL | 471 | Ethylene oxide gas |
| Spider double lumen | Cair LGL | PY2101NCM | PUR | Ø = 1 x 2 mm, L = 8 cm, Vol. = 0.40 mL | 251.20 | Ethylene oxide gas |
| Smallbore double lumen | ICU Medical | 011-MC33076 | PUR | Ø = 1.2 x 2.1 mm, L = 12 cm, Vol. = 0.51 mL | 452.16 | Gamma irradiation |
| Smallbore double lumen | ICU Medical | 011-MC33077 | PUR | Ø = 1.2 x 2.1 mm, L = 8 cm, Vol. = 0.47 mL | 301.44 | Gamma irradiation |
| Smallbore double lumen | ICU Medical | 011-MC33165 | PVC | Ø = 1.2 x 2.1 mm, L = 9 cm, Vol. = 0.43 mL | 339.12 | Gamma irradiation |
| MaxPlus Clear 2 way connector | Carefusion | MFX2502MP | PVC | Ø = 1.0 x 0.55 mm, L = 5 cm, Vol. > 0.50 mL | 172.70 | Ethylene oxide gas |
| Edelvaiss-CW2+ | Doran International | Edelvaiss-CW2+ | PE/PVC | Ø = 0.7 x 1.7 mm, L = 12 cm, Vol. = 0.05 mL | 263.76 | Ethylene oxide gas |
PUR: polyurethane, PVC: polyvinyl chloride, PE: polyethylene
Ø: diameter, L: length, Vol.: volume (per tube).
Fig 2Mean concentrations of isosorbide dinitrate (as % of the initial syringe concentrations) delivered at the egress of the double-lumen extension tubes during the 150-minute simulated infusion.
(A) PUR tubes. (B) PVC and PE tubes.
Fig 3Mean concentrations of diazepam (as % of the initial syringe concentrations) delivered at the egress of the double-lumen extension tubes during the 150-minute simulated infusion.
(A) PUR tubes. (B) PVC and PE tubes.
Fig 4Mean concentrations of noradrenaline (as % of the initial syringe concentrations) delivered at the egress of the double-lumen extension tubes during the 150-minute simulated infusion.
Mean concentrations of isosorbide dinitrate (as % of the initial syringe concentrations) delivered at the egress of the double-lumen extension tubes and mean concentrations of isosorbide dinitrate sorbed per internal surface (in % per mm2) at T0+150 min.
| Double-lumen extension tube | Reference | Tubing material | Mean concentrations of isosorbide dinitrate delivered (%) | Mean concentrations of isosorbide dinitrate sorbed per internal surface (% per mm2) |
|---|---|---|---|---|
| Octopus 2 | 841.264 | PUR | 81.37 ± 2.35 | 0.040 ± 0.005 |
| Octopus 2 | 5841.208 | PUR | 80.53 ± 1.66 | 0.041 ± 0.004 |
| Spider double lumen | PY2101NCM | PUR | 83.40 ± 0.61 | 0.066 ± 0.002 |
| Smallbore double lumen | 011-MC33076 | PUR | 89.15 ± 2.09 | 0.024 ± 0.005 |
| Smallbore double lumen | 011-MC33077 | PUR | 89.81 ± 1.40 | 0.034 ± 0.005 |
| Smallbore double lumen | 011-MC33165 | PVC | 89.75 ± 1.16 | 0.030 ± 0.003 |
| MaxPlus Clear 2 way connector | MFX2502MP | PVC | 85.43 ± 1.22 | 0.084 ± 0.007 |
| Edelvaiss-CW2+ | Edelvaiss-CW2+ | PE/PVC | 92.84 ± 2.73 | 0.027 ± 0.010 |
PUR: polyurethane, PVC: polyvinyl chloride, PE: polyethylene
Mean concentrations of isosorbide dinitrate sorbed per internal surface at T0+150 min are significantly different (analysis of variance: ANOVA, p ≤ 0.0001).
1 Significant difference vs. Spider, Smallbore 011-MC33076, MaxPlus Clear and Edelvaiss-CW2+ (Tukey test, p ≤ 0.030)
2 Significant difference vs. Spider, Smallbore 011-MC33076, MaxPlus Clear and Edelvaiss-CW2+ (Tukey test, p ≤ 0.009)
3 Significant difference vs. Octopus 2 ref. 841.264, Octopus 2 ref. 5841.208, Smallbore 011-MC33076, Smallbore 011-MC33077, Smallbore 011-MC33165, MaxPlus Clear and Edelvaiss-CW2+ (Tukey test, p < 0.0001)
4 Significant difference vs. Octopus 2 ref. 841.264, Octopus 2 ref. 5841.208, Spider and MaxPlus Clear (Tukey test, p ≤ 0.001)
5 Significant difference vs. Spider and MaxPlus Clear (Tukey test, p < 0.0001)
6 Significant difference vs. Spider and MaxPlus Clear (Tukey test, p < 0.0001)
7 Significant difference vs. Octopus 2 ref. 841.264, Octopus 2 ref. 5841.208, Spider, Smallbore 011-MC33076, Smallbore 011-MC33077, Smallbore 011-MC33165 and Edelvaiss CW2+ (Tukey test, p < 0.0001)
8 Significant difference vs. Octopus 2 ref. 841.264, Octopus 2 ref. 5841.208, Spider and MaxPlus Clear (Tukey test, p ≤ 0.030).
Mean concentrations of diazepam (as % of the initial syringe concentrations) delivered at the egress of the double-lumen extension tubes and mean concentrations of diazepam sorbed per internal surface (in % per mm2) at T0+150 min.
| Double-lumen extension tube | Reference | Tubing material | Mean concentrations of diazepam delivered (%) | Mean concentrations of diazepam sorbed per internal surface (% per mm2) |
|---|---|---|---|---|
| Octopus 2 | 841.264 | PUR | 48.58 ± 2.88 | 0.109 ± 0.006 |
| Octopus 2 | 5841.208 | PUR | 50.30 ± 1.46 | 0.106 ± 0.003 |
| Spider double lumen | PY2101NCM | PUR | 59.30 ± 1.21 | 0.162 ± 0.005 |
| Smallbore double lumen | 011-MC33076 | PUR | 78.83 ± 2.26 | 0.047± 0.005 |
| Smallbore double lumen | 011-MC33077 | PUR | 81.88 ± 2.49 | 0.060 ± 0.008 |
| Smallbore double lumen | 011-MC33165 | PVC | 77.89 ± 2.51 | 0.065 ± 0.007 |
| MaxPlus Clear 2 way connector | MFX2502MP | PVC | 59.99 ± 1.24 | 0.232 ± 0.007 |
| Edelvaiss-CW2+ | Edelvaiss-CW2+ | PE/PVC | 85.06 ± 3.94 | 0.057 ± 0.015 |
PUR: polyurethane, PVC: polyvinyl chloride, PE: polyethylene
Mean concentrations of diazepam sorbed per internal surface at T0+150 min are significantly different (analysis of variance: ANOVA, p <0.0001).
1 Significant difference vs. Spider, Smallbore 011-MC33076, Smallbore011-MC33077, Smallbore011-MC33165, MaxPlus Clear and Edelvaiss-CW2+ (Tukey test, p < 0.0001)
2 Significant difference vs. Spider, Smallbore 011-MC33076, Smallbore011-MC33077, Smallbore011-MC33165, MaxPlus Clear and Edelvaiss-CW2+ (Tukey test, p < 0.0001)
3 Significant difference vs. Octopus 2 ref. 841.264, Octopus 2 ref. 5841.208, Smallbore 011-MC33076, Smallbore011-MC33077, Smallbore011-MC33165, MaxPlus Clear and Edelvaiss-CW2+ (Tukey test, p < 0.0001)
4 Significant difference vs. Octopus 2 ref. 841.264, Octopus 2 ref. 5841.208, Spider, Smallbore 011-MC33165 and MaxPlus Clear (Tukey test, p ≤ 0.015)
5 Significant difference vs. Octopus 2 ref. 841.264, Octopus 2 ref. 5841.208, Spider and MaxPlus Clear (Tukey test, p < 0.0001)
6 Significant difference vs. Octopus 2 ref. 841.264, Octopus 2 ref. 5841.208, Spider, Smallbore 011-MC33076 and MaxPlus Clear (Tukey test, p ≤ 0.015)
7 Significant difference vs. Octopus 2 ref. 841.264, Octopus 2 ref. 5841.208, Spider, Smallbore 011-MC33076, Smallbore 011-MC33077, Smallbore011-MC33165 and Edelvaiss CW2+ (Tukey test, p < 0.0001)
8 Significant difference vs. Octopus 2 ref. 841.264, Octopus 2 ref. 5841.208, Spider and MaxPlus Clear (Tukey test, p < 0.0001).
Fig 5Mean concentrations of drug sorbed per internal surface (in % per mm2) at T0+150 min for each double-lumen extension tube investigated in the case of isosorbide dinitrate and diazepam.
Determining the type of plastic (thermoplastic or thermosetting) for PUR double-lumen extension tubes through different tests (heat resistance and reaction in an organic solvent-type methanol).
| Double-lumen extension tube | Manufacturer | Reference | Heat resistance | Reaction in an organic solvent (methanol) | Type of PUR |
|---|---|---|---|---|---|
| Octopus 2 | Vygon | 841.264 | Deformable | Stretching | Thermoplastic |
| Octopus 2 | Vygon | 5841.208 | Deformable | Stretching | Thermoplastic |
| Spider double lumen | Cair LGL | PY2101NCM | Deformable | Breaking | Thermoplastic |
| Smallbore double lumen | ICU Medical | 011-MC33076 | Non-deformable | Stretching | Thermosetting |
| Smallbore double lumen | ICU Medical | 011-MC33077 | Non-deformable | Stretching | Thermosetting |
PUR: polyurethane.