The effect of tubing dwell time on insulin adsorption during intravenous insulin infusions.

Cecilia D Thompson1, Jessica Vital-Carona, E Vincent S Faustino.   

Abstract

BACKGROUND: Insulin adsorbs to plastic tubing, which decreases the concentration of an insulin solution delivered from an intravenous infusion set. Dwelling insulin within tubing before starting the infusion decreases adsorption but delays treatment initiation and wastes time in infusion preparation. The lack of data on dwell time effects results in wide variability in practice. We aim to determine the effect of dwell time on insulin concentration from intravenous infusion tubing.
MATERIALS AND METHODS: In this in vitro study, we used insulin solutions with concentrations of 0.1 unit/mL, 1 unit/mL, and 10 units/mL. Each solution dwelled in intravenous infusion sets for 0, 15, 30, or 60 min. After the dwell, we measured insulin concentrations from the solution bags and tubing. We repeated each insulin concentration-dwell time combination five times. Comparisons were performed using analyses of variance.
RESULTS: For each of the three insulin concentrations, the mean insulin concentrations from the tubing were not significantly different between dwell times. Duration of dwell time did not affect insulin adsorption in polypropylene intravenous infusion sets.
CONCLUSIONS: We recommend that following a 20-mL flush, insulin infusions can be started without any dwell time. Removal of dwell times may improve clinical practice by minimizing preparation time and will allow faster initiation of insulin infusion therapy.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22746979      PMCID: PMC3458999          DOI: 10.1089/dia.2012.0098

Source DB:  PubMed          Journal:  Diabetes Technol Ther        ISSN: 1520-9156            Impact factor:   6.118


  14 in total

1.  The impact of insulin adsorption onto the infusion sets in the adult intensive care unit.

Authors:  Therese Jakobsson; Robert Shulman; Hardyal Gill; Kevin Taylor
Journal:  J Diabetes Sci Technol       Date:  2009-01

2.  Insulin adsorption by glass infusion bottles, polyvinylchloride infusion containers, and intravenous tubing.

Authors:  C Petty; N L Cunningham
Journal:  Anesthesiology       Date:  1974-04       Impact factor: 7.892

3.  Adsorption of insulin to infusion bottles and tubing.

Authors:  S Weisenfeld; S Podolsky; L Goldsmith; L Ziff
Journal:  Diabetes       Date:  1968-12       Impact factor: 9.461

4.  Lispro insulin: adsorption and stability in selected intravenous devices.

Authors:  J Ling; M Hu; T Hagerup; R K Campbell
Journal:  Diabetes Educ       Date:  1999 Mar-Apr       Impact factor: 2.140

5.  Continuous insulin infusions in neonates: pharmacologic availability of insulin in intravenous solutions.

Authors:  P S Simeon; M E Geffner; S R Levin; A M Lindsey
Journal:  J Pediatr       Date:  1994-05       Impact factor: 4.406

6.  Clinical signigicance of insulin adsorption by polyvinyl chloride infusion systems.

Authors:  J I Hirsch; M J Fratkin; J H Wood; R B Thomas
Journal:  Am J Hosp Pharm       Date:  1977-06

7.  Insulin infusions in the neonatal unit: delivery variation due to adsorption.

Authors:  M Hewson; V Nawadra; J Oliver; C Odgers; J Plummer; K Simmer
Journal:  J Paediatr Child Health       Date:  2000-06       Impact factor: 1.954

8.  Carrier solutions for low-level intravenous insulin infusion.

Authors:  E W Kraegen; L Lazarus; H Meler; L Campbell; Y O Chia
Journal:  Br Med J       Date:  1975-08-23

9.  Insulin adsorbance to polyvinylchloride (PVC) surfaces of fluid container and infusion-set.

Authors:  A Seifi; A Mowla; Moien M T Vaziri; A R Talei; M R Namazy
Journal:  Middle East J Anaesthesiol       Date:  2004-06

10.  Effect of whole blood on insulin adsorption onto intravenous infusion systems.

Authors:  J Kerchner; D M Colaluca; R P Juhl
Journal:  Am J Hosp Pharm       Date:  1980-10
View more
  7 in total

1.  Clinical Recommendations for Managing the Impact of Insulin Adsorptive Loss in Hospital and Diabetes Care.

Authors:  Jennifer L Knopp; J Geoffrey Chase
Journal:  J Diabetes Sci Technol       Date:  2020-04-24

2.  Do bevacizumab solutions interact with silicone or polyurethane catheters during an infusion through implantable venous access ports?

Authors:  Nicolas Tokhadzé; Philip Chennell; Régis Cueff; Valérie Sautou
Journal:  J R Soc Interface       Date:  2019-09-25       Impact factor: 4.118

3.  Capacity of Infusion Lines for Insulin Adsorption: Effect of Flow Rate on Total Adsorption.

Authors:  Jennifer L Knopp; Kaia Bishop; Theodore Lerios; J Geoffrey Chase
Journal:  J Diabetes Sci Technol       Date:  2019-09-27

4.  Effect of Tubing Flush or Preconditioning on Available Insulin Concentration for IV Infusion: A Pilot Project.

Authors:  Brandi Newby; Daniel T Holmes
Journal:  Can J Hosp Pharm       Date:  2017-08-31

5.  Quantitative assessment of required separator fluid volume in multi-infusion settings.

Authors:  Frank Doesburg; Daniek Middendorp; Willem Dieperink; Wouter Bult; Maarten W Nijsten; Daan J Touw
Journal:  J Vasc Access       Date:  2020-05-04       Impact factor: 2.283

6.  Impact of alternative materials to plasticized PVC infusion tubings on drug sorption and plasticizer release.

Authors:  N Tokhadze; P Chennell; L Bernard; C Lambert; B Pereira; B Mailhot-Jensen; V Sautou
Journal:  Sci Rep       Date:  2019-12-12       Impact factor: 4.379

7.  Addition of Regular Insulin to Ternary Parenteral Nutrition: A Stability Study.

Authors:  Heloise Henry; Damien Lannoy; Patrice Maboudou; David Seguy; Thierry Dine; Pascal Pigny; Pascal Odou
Journal:  Pharmaceutics       Date:  2021-03-27       Impact factor: 6.321

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.