Steam induced by the activation of energy devices under a wet condition may cause thermal injury.

Masaru Hayami1, Masayuki Watanabe2, Shinji Mine1, Yu Imamura1, Akihiko Okamura1, Masami Yuda1, Kotaro Yamashita1, Yoshiaki Shoji1, Tasuku Toihata1, Ryotaro Kozuki1, Naoki Ishizuka3.   

Abstract

BACKGROUND: During esophagectomy for esophageal cancer, meticulous attention is needed to prevent thermal injury to the vital organs, such as the recurrent laryngeal nerve (RLN) and tracheobronchus. In order to clarify the novel mechanism behind thermal injury induced by energy devices, we investigated the temperature of steam with the use of two different devices under wet and dry conditions.
METHODS: An ultrasonic device (Sonicision™) and a vessel sealing device (Ligasure™) were studied. We evaluated the temperature at the tip of the devices and the steam when the devices were activated under different grasping ranges, under four different combinations of device and muscle, and under four different wet/dry conditions (dry-dry, dry-wet, wet-dry, and wet-wet).
RESULTS: Although the maximum temperature of the devices was significantly higher with Sonicision™ than with Ligasure™, the maximum temperature of the steam was significantly higher with Ligasure™ than with Sonicision™ in almost all situations. At 1 mm away from Sonicision™, the critical temperature more than 60 °C was observed only when used with one-third grasping range under the wet-dry or the wet-wet conditions. In case of Ligasure™, high-temperature steam was observed when used with one-third grasping under the wet-dry or the wet-wet condition and two-third grasping under the dry-wet, the wet-dry, or the wet-wet condition. Under the wet condition, the emission of steam from the non-grasping part of Ligasure™ caused a spike in temperature that exceeded the critical temperature.
CONCLUSION: We demonstrated that the use of energy devices under a wet condition generates steam from the non-grasping part of the devices. The temperatures of steam from Ligasure™ were significantly higher than that from Sonicision™. To prevent thermal injury to the vital organs, a very attentive and meticulous surgical technique is imperative considering the characteristics of each device.

Entities:  

Keywords:  Energy device; Grasping range; Steam; Thermal injury

Mesh:

Substances:

Year:  2019        PMID: 31811453     DOI: 10.1007/s00464-019-07312-w

Source DB:  PubMed          Journal:  Surg Endosc        ISSN: 0930-2794            Impact factor:   4.584


  11 in total

1.  Randomized clinical trial of Ligasure versus conventional surgery for extended gastric cancer resection.

Authors:  W-J Lee; T-C Chen; I-R Lai; W Wang; M-T Huang
Journal:  Br J Surg       Date:  2003-12       Impact factor: 6.939

2.  Determination of temperature elevation in tissue during the application of the harmonic scalpel.

Authors:  Christian Koch; Thomas Friedrich; Frank Metternich; Andrea Tannapfel; Hans-Peter Reimann; Uwe Eichfeld
Journal:  Ultrasound Med Biol       Date:  2003-02       Impact factor: 2.998

3.  Tissue penetration of bipolar electrosurgical currents: Joule overheating beyond the surface layer.

Authors:  Walter Habermann; Wolfram Müller
Journal:  Head Neck       Date:  2012-04-23       Impact factor: 3.147

4.  Evaluation of the safety, efficacy, and versatility of a new surgical energy device (THUNDERBEAT) in comparison with Harmonic ACE, LigaSure V, and EnSeal devices in a porcine model.

Authors:  Jeffrey Milsom; Koiana Trencheva; Sebastien Monette; Raghava Pavoor; Parul Shukla; Junjun Ma; Toyooki Sonoda
Journal:  J Laparoendosc Adv Surg Tech A       Date:  2012-02-24       Impact factor: 1.878

5.  Comparison of lateral thermal spread using monopolar and bipolar diathermy, the Harmonic Scalpel and the Ligasure.

Authors:  P A Sutton; S Awad; A C Perkins; D N Lobo
Journal:  Br J Surg       Date:  2010-03       Impact factor: 6.939

6.  The Effects of Inhaled Heat on the Air Passages and Lungs: An Experimental Investigation.

Authors:  A R Moritz; F C Henriques; R McLean
Journal:  Am J Pathol       Date:  1945-03       Impact factor: 4.307

7.  Electrophysiologic monitoring correlates of recurrent laryngeal nerve heat thermal injury in a porcine model.

Authors:  Yi-Chu Lin; Gianlorenzo Dionigi; Gregory W Randolph; I-Cheng Lu; Pi-Ying Chang; Shan-Yin Tsai; Hoon Yub Kim; Hye Yoon Lee; Ralph P Tufano; Hui Sun; Xiaoli Liu; Feng-Yu Chiang; Che-Wei Wu
Journal:  Laryngoscope       Date:  2015-05-22       Impact factor: 3.325

8.  A comparative study of the tissue-destroying effect of the laser and electrocoagulation.

Authors:  W Matek; H D Reidenbach; A Wittmann; L Beierlein; P Hermanek
Journal:  Endoscopy       Date:  1989-01       Impact factor: 10.093

9.  Use of the harmonic scalpel versus conventional haemostatic techniques in patients with Grave disease undergoing total thyroidectomy: a prospective randomised controlled trial.

Authors:  P Hallgrimsson; L Lovén; J Westerdahl; Anders Bergenfelz
Journal:  Langenbecks Arch Surg       Date:  2008-08-02       Impact factor: 3.445

10.  How to prevent lateral thermal damage to tissue using the harmonic scalpel: experimental study on pig small intestine and abdominal wall.

Authors:  Z Pogorelić; Z Perko; N Druzijanić; S Tomić; I Mrklić
Journal:  Eur Surg Res       Date:  2009-06-26       Impact factor: 1.745

View more
  5 in total

Review 1.  Improving Voice Outcomes after Thyroid Surgery and Ultrasound-Guided Ablation Procedures.

Authors:  Pia Pace-Asciak; Jon O Russell; Ralph P Tufano
Journal:  Front Surg       Date:  2022-05-04

Review 2.  Improving Voice Outcomes After Thyroid Surgery - Review of Safety Parameters for Using Energy-Based Devices Near the Recurrent Laryngeal Nerve.

Authors:  Jia Joanna Wang; Tzu-Yen Huang; Che-Wei Wu; Yi-Chu Lin; Hsin-Yi Tseng; Cheng-Hsin Liu; I-Cheng Lu; Pi-Ying Chang; Hui-Chun Chen; Hsiu-Ya Chen; Gianlorenzo Dionigi; Feng-Yu Chiang; Ling-Feng Wang
Journal:  Front Endocrinol (Lausanne)       Date:  2021-11-24       Impact factor: 5.555

3.  Effects of Intraoperative Neural Tunnel in Protecting Recurrent Laryngeal Nerve: Experiences in Open, Trans Breast, and Transoral Endoscopic Thyroidectomy.

Authors:  Xing Yu; Yujun Li; Chang Liu; Yuancong Jiang; Zhaodi Liu; Qionghua He; Yong Wang; Ping Wang
Journal:  Front Oncol       Date:  2022-02-23       Impact factor: 6.244

Review 4.  Features and applications of energy devices for prone robot-assisted minimally invasive esophagectomy: a narrative review.

Authors:  Noriyuki Hirahara; Takeshi Matsubara; Hikota Hayashi; Yoshitsugu Tajima
Journal:  J Thorac Dis       Date:  2022-09       Impact factor: 3.005

5.  Intraoperative Measures to Reduce the Risk of COVID-19 Transmission During Minimally Invasive Procedures: A Systematic Review and Critical Appraisal of Societies' Recommendations.

Authors:  Giovanni A Tommaselli; Philippe Grange; Crystal D Ricketts; Jeffrey W Clymer; Raymond S Fryrear
Journal:  Surg Laparosc Endosc Percutan Tech       Date:  2021-07-28       Impact factor: 1.719

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.