Abstract
INTRODUCTION: Unnecessary and excessive activation of alarms ("false alarm") in neonatal intensive care unit (NICU) often results in alarm fatigue among health-care professionals, which can potentially result in deleterious effects in sick neonates. AIMS ANDEntities:
Keywords: False alarms; multiparameter monitors; quality improvement initiative
Year: 2019 PMID: 31807593 PMCID: PMC6852305 DOI: 10.4103/jehp.jehp_226_19
Source DB: PubMed Journal: J Educ Health Promot ISSN: 2277-9531
Figure 1Fishbone analysis of causes of activation of false alarms from multiparameter monitor
Details of plan-do-study-cycle
| PDSA cycle (time line) | Plan | Do | Study | Act |
|---|---|---|---|---|
| PDSA 1 (11.11.16-24.11.16) | Training of HCP regarding use of MPM | 1.Group teaching, demonstration, counter demonstration on Mindray MPM, | 1.Training on less frequently used para meters like ECG, IBP were not welcomed by new nurses who found it difficult to understand , how ever they were prompt in learning use of parameters like HR and SPO2 | Training was focused on HR, SOP2 first, other parameters like IBP, ECG were dropped |
| PDSA 2 (25.11.16-01.12.16) | Intensification of training | 1.1:1 teaching, | 1.Median no of activation of false alarms decreased to 19 (15-30)/hr/bed | 1.Intensified teaching continued and no further knowledge level scoring was done |
| PDSA 3 (02.12.16-15.12.16) | Team capacity build up by expansion of team members | Same as PDSA 2, validation of new team members | Median no of activation of false alarms decreased to 16 (14-30)/hr/bed, 4new team members were identified | Plan adopted |
| PDSA 4 (16.12.16-22.12.16) | Dissemination of knowledge by use of social media, sustenance of gained results | Same as PDSA 3, Use of social media (whatsapp reminder), making alarm limit check as routine policy during shift change hours | Median no of activation of false alarms decreased to 14 (8-17)/hr/bed | Unit policy was build up and plan was adopted |
| PDSA 5 (23.12.16-29.12.16) | Inclusion of various other types of MPM | As above , increasing frequency and number of MPMs and including on variety of MPMs apart from Mindray, | Median no of activation of false alarms decreased to 9 (6-12)/h/bed | Ongoing training of new staff nurses continues, refreshment training of existing staff continued |
Data is represented as median (range). PDSA:Plan-do-Study-Act, HCP=Health-care provider, MPMs=Multiparameter monitors, ECG=Electrocardiography, BP= Blood pressure, HR=Heart rate, IBP=Intra arterial blood pressure monitoring
Figure 2Process measure run chart showing assessment of knowledge of use of multiparameter monitors
Figure 3Run chart showing activation of false alarm/hour/1MPM
Figure 4Run chart on activation of false alarm/hour/1MPM in sustenance phase