Marion Morena1, Caroline Creput2, Mouloud Bouzernidj3, Annie Rodriguez4,5, Lotfi Chalabi5, Bruno Seigneuric6, Céline Lauret1, Anne-Sophie Bargnoux1, Anne-Marie Dupuy4, Jean-Paul Cristol7,8.
Abstract
This prospective multicenter randomized comparative cross-over trial aimed at evaluating the influence of hemodialysis vs post-dilution hemodiafiltration with high-flux dialyzers in solute clearance and biocompatibility profile. 32 patients were sequentially dialyzed with Leoceed-21HX, Polypure-22S+, Rexsys-27H and VIE-21A. Primary outcome was β2-microglobulin removal. Secondary outcomes were (i) extraction of other uremic solutes (ii) parameters of inflammation and nutrition and (iii) comparative quantification of perdialytic albumin losses (using total 'TDC' vs partial 'PDC' collection of dialysate). Significant increases in removal rates of β2-microglobulin (84.7 ± 0.8 vs 71.6 ± 0.8 mg/L), myoglobin (65.9 ± 1.3 vs 38.6 ± 1.3 µg/L), free immunoglobulin light chains Kappa (74.9 ± 0.8 vs 55.6 ± 0.8 mg/L), β-trace protein (54.8 ± 1.3 vs 26.8 ± 1.4 mg/L) and orosomucoid (11.0 ± 1.1 vs 6.0 ± 1.1 g/L) but not myostatin (14.8 ± 1.5 vs 13.0 ± 1.5 ng/mL) were observed in HDF compared to HD when pooling all dialyzers. Rexsys and VIE-A use in both HD and HDF subgroups was associated to a better removal of middle/large-size molecules compared to Leoceed and Polypure, except β2-microglobulin for Rexsys. Inflammatory parameters were unchanged between dialyzers without any interaction with dialysis modality. Mean dialysate albumin loss was comparable between TDC and PDC (1.855 vs 1.826 g/session for TDC and PDC respectively). In addition, a significant difference in albumin loss was observed between dialyzers with the highest value (4.5 g/session) observed using Rexsys. Use of all dialyzers was associated with good removals of the large spectrum of uremic toxins tested and good biocompatibility profiles, with an additional gain in removal performances with HDF. Larger surface area, thinner wall and resultant very high ultrafiltration coefficient of Rexsys should be taken into account in its clear performance advantages.Entities:
Mesh:
Substances:
Year: 2019 PMID: 31797880 PMCID: PMC6892817 DOI: 10.1038/s41598-019-54404-7
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Baseline characteristics of the studied population.
| Parameter | All | HD group | HDF group |
|---|---|---|---|
| N | 32 | 16 | 16 |
| Gender, Male (%) | 20 (62.5%) | 9 (56.3%) | 11 (68.8%) |
| Age (years) | 72 (10); 73 [42–89] | 74 (10); 76 [55–89] | 69 (11); 70 [42–84] |
| Dry weight (kg) | 73.0 (17.1); 71.8 [39.5–114.0] | 69.9 (13.7); 70.5 [39.5–87.5] | 76.1 (19.8); 76.0 [47.5–114.0] |
| BMI (kg/m2) | 26.6 (6.9); 25.3 [15.5–49.4] | 26.4 (7.7); 23.6 [16.9–49.4] | 26.9 (6.2); 27.0 [15.5–36.4] |
| Hypertension (%) | 24 (75.0%) | 14 (87.5%) | 10 (62.5%) |
| Diabetes mellitus (%) | 10 (31.3%) | 6 (37.5%) | 4 (25.0%) |
| Dialysis vintage (years) | 4.8 (3.3); 4.4 [0.4–16.6] | 5.5 (4.2); 4.7 [0.4–16.6] | 4.1 (2.0); 3,6 [1.6–8.6] |
| Vascular access, AVF (%) | 28 (87.5%) | 13 (81.3%) | 15 (93.8%) |
| Erythropoiesis stimulating agents (%) | 25 (78.1%) | 12 (75.0%) | 13 (81.3%) |
| Erythropoiesis stimulating agents (IU/kg/week) | 84.8 (60.6); 71.4 [15.7–231.5] | 83.2 (64.1); 70.0 [15.7–231.5] | 86.3 (59.7); 71.4 [23.6–210.5] |
| Iron injection (%) | 20 (62.5%) | 10 (62.5%) | 10 (62.5%) |
| Iron injection (mg/week) | 60.0 (35.4); 50.0 [20.0–100.0] | 35.0 (25.8); 22.5 [20.0–100.0] | 85.0 (24.2); 100.0 [50.0–100.0] |
| Hemoglobin (g/dL) | 11.5 (1.2); 11.5 (8.9–14.2) | 11.6 (1.4); 11.6 (8.9–14.2) | 11.5 (1.0); 11.4 (9.8–13.7) |
| Hematocrit (%) | 35.4 (3.8); 35.2 (27.0–44.9) | 35.7 (4.5); 35.3 (27.0–44.9) | 35.0 (3.1); 35.2 (29.9–42.7) |
| Transthyretin (g/L) | 0.31 (0.08); 0.32 (0.14–0.46) | 0.29 (0.08); 0.30 (0.14–0.46) | 0.32 (0.06); 0.33 (0.22–0.45) |
| Leukocytes (G/L) | 6.5 (2.1); 6.1 (3.8–14.5) | 7.2 (2.6); 6.4 (4.3–14.5) | 5.8 (1.3); 5.4 (3.8–8.3) |
| Inorganic phosphates (mmol/L) | 1.5 (0.4); 1.4 (0.6–2.3) | 1.4 (0.3); 1.4 (0.6–1.9) | 1.6 (0.4); 1.6 (0.9–2.3) |
| C reactive protein (mg/L) | 12.4 (22.3); 4.7 (0.6–100.4) | 21.1 (29.4); 9.0 (0.8–100.4) | 3.7 (2.3); 3.0 (0.6–7.4) |
Values were described using proportions for categorical variables and as mean (sd); median [minimum-maximum] for quantitative variables.
Characteristics of treatment according to the two groups (HD and HDF) during follow-up.
| Parameter | Leoceed 21HX | Polypure 22S+ | Rexsys 27H | VIE-21A | ||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| HD | HDF | HD | HDF | HD | HDF | HD | HDF | |||||||||
| mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | |
| Duration of dialysis session (h) | 3.90 | (0.20) | 4.00 | (0.00) | 4.00 | (0.10) | 3.90 | (0.20) | 4.00 | (0.10) | 4.00 | (0.10) | 3.90 | (0.40) | 4.00 | (0.00) |
| Blood flow (mL/min) | 348.8 | (35.8) | 374.1 | (23.8) | 341.3 | (35.9) | 375.8 | (22.9) | 346.9 | (39.6) | 379.4 | (23.6) | 335.3 | (34.2) | 370.9 | (22.3) |
| Dialysate flow (mL/min) | 561.5 | (50.6) | 650.0 | (51.6) | 546.2 | (66.0) | 650.0 | (51.6) | 546.2 | (66.0) | 650.0 | (51.6) | 546.2 | (51.9) | 646.7 | (51.6) |
| Intradialytic weight change (kg) | 2.5 | (3.7) | 2.3 | (1.0) | 1.5 | (1.1) | 2.2 | (1.3) | 1.6 | (1.2) | 2.5 | (1.2) | 2.3 | (1.5) | 2.3 | (1.2) |
| Convection volume (L/session) | — | — | 23.8 | (1.6) | — | — | 24.3 | (2.0) | — | — | 24.6 | (1.9) | — | — | 22.8 | (1.6) |
Values were described using mean (sd).
Primary outcome: b2m performances with the 4 tested dialyzers using HD and HDF modalities.
| Parameter | Leoceed 21HX | Polypure 22 S+ | Rexsys 27 H | VIE-21A | ANOVA | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| HD | HDF | HD | HDF | HD | HDF | HD | HDF | Dialyzer | Mode | Dial* Mode | |||||||||
| mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | ||||
| RR (%) | 70.9 | (1.5) | 84.3 | (1.5) | 67.6 | (1.6) | 83.5 | (1.5) | 74.0 | (1.6) | 85.6 | (1.5) | 73.9 | (1.5) | 85.6 | (1.6) | 0.023 | <0.001 | NS |
| RR-eq (%) | 57.3 | (1.6) | 72.1 | (1.6) | 54.0 | (1.8) | 70.9 | (1.6) | 60.5 | (1.7) | 73.6 | (1.6) | 59.9 | (1.6) | 73.6 | (1.7) | 0.025 | <0.001 | NS |
| Kt/V | 0.87 | (0.05) | 1.29 | (0.05) | 0.79 | (0.05) | 1.25 | (0.05) | 0.95 | (0.05) | 1.34 | (0.04) | 0.93 | (0.05) | 1.34 | (0.05) | 0.026 | <0.001 | NS |
| K skm (mL/min) | 45.7 | (3.4) | 74.3 | (3.4) | 40.8 | (3.8) | 72.9 | (3.4) | 50.7 | (3.5) | 76.3 | (3.3) | 50.7 | (3.4) | 76.0 | (3.5) | NS | <0.001 | NS |
| Mass (mg) | 183.9 | (14.9) | 216.6 | (14.9) | 168.6 | (16.5) | 197.0 | (14.9) | 186.3 | (15.4) | 196.4 | (14.4) | 189.3 | (14.9) | 197.4 | (15.4) | NS | NS | NS |
| RR (%) | NS | NS | NS | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| RR-eq (%) | NS | NS | NS | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| Kt/V | NS | NS | NS | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| K skm (mL/min) | — | — | — | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| Mass (mg) | — | — | — | — | — | — | — | ||||||||||||
Values were described using mean (sd).
In the ANOVA column, Dialyzer means dialyzer effect; Mode means mode effect and Dial * Mode means Interaction between dialyzer and mode.
Bonferroni post-hoc tests correspond to dialyzers effect whatever the mode was. ANOVA Mode corresponds to ANOVA testing dialyzers effect according to mode. ANOVA Dialyzer corresponds to ANOVA testing mode effects according to dialyzers. (skm, simplified kinetic model).
Small solute performances with the 4 tested dialyzers using HD and HDF modalities.
| Parameter | Leoceed 21HX | Polypure 22S+ | Rexsys 27H | VIE-21A | ANOVA | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| HD | HDF | HD | HDF | HD | HDF | HD | HDF | Dialyzer | Mode | Dial * Mode | |||||||||
| mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | ||||
| ( | |||||||||||||||||||
| spKt/V | 1.80 | (0.09) | 1.99 | (0.08) | 1.84 | (0.09) | 1.94 | (0.09) | 1.75 | (0.10) | 1.93 | (0.08) | 1.90 | (0.09) | 2.03 | (0.09) | NS | 0.016 | NS |
| eqKt/V | 1.56 | (0.07) | 1.72 | (0.07) | 1.59 | (0.08) | 1.68 | (0.07) | 1.52 | (0.08) | 1.66 | (0.07) | 1.63 | (0.07) | 1.76 | (0.07) | NS | 0.012 | NS |
| Kt/V Urea skm | 1.33 | (0.11) | 1.48 | (0.11) | 1.47 | (0.11) | 1.59 | (0.11) | 1.53 | (0.12) | 1.42 | (0.11) | 1.40 | (0.11) | 1.50 | (0.11) | NS | NS | NS |
| RR-eq Urea (%) | 73.0 | (1.8) | 76.5 | (1.7) | 75.0 | (1.7) | 76.9 | (1.7) | 74.7 | (1.9) | 75.4 | (1.7) | 74.9 | (1.8) | 77.1 | (1.8) | NS | NS | NS |
| K skm Urea (mL/min) | 209.8 | (13.8) | 249.0 | (13.4) | 234.1 | (13.4) | 265.3 | (13.4) | 237.6 | (14.9) | 239.4 | (13.4) | 224.9 | (13.8) | 251.1 | (13.8) | NS | 0.013 | NS |
| Urea mass (mmol) | 496.8 | (78.3) | 583.9 | (75.8) | 561.8 | (78.3) | 668.1 | (75.8) | 513.9 | (84.1) | 568.3 | (75.8) | 549.7 | (78.3) | 564.7 | (78.3) | NS | NS | NS |
| ( | |||||||||||||||||||
| Kt/V Creatinine skm | 1.00 | (0.06) | 1.26 | (0.06) | 1.00 | (0.06) | 1.13 | (0.06) | 1.11 | (0.07) | 1.12 | (0.06) | 1.04 | (0.07) | 1.16 | (0.07) | NS | NS | NS |
| RR-eq Creatinine (%) | 62.6 | (1.7) | 69.6 | (1.7) | 62.7 | (1.7) | 66.8 | (1.7) | 65.9 | (1.9) | 67.0 | (1.7) | 64.1 | (1.8) | 68.1 | (1.8) | NS | 0.001 | NS |
| K skm Creatinine (mL/min) | 158.1 | (10.1) | 214.2 | (10.1) | 157.0 | (10.5) | 193.3 | (10.1) | 176.8 | (11.2) | 189.0 | (10.1) | 166.9 | (10.5) | 190.1 | (10.8) | NS | <0.001 | NS |
| Creatinine mass (µmol × 10−3) | 15.0 | (1.3) | 18.3 | (1.3) | 15.7 | (1.3) | 18.2 | (1.3) | 16.2 | (1.4) | 18.9 | (1.3) | 15.8 | (1.3) | 18.0 | (1.4) | NS | 0.005 | NS |
Values were described using mean (sd).
In the ANOVA column, Dialyzer means dialyzer effect; Mode means mode effect and Dial * Mode means Interaction between dialyzer and mode.
(skm, simplified kinetic model).
Other middle and large molecule performances with the 4 tested dialyzers using HD and HDF modalities.
| Parameter | Leoceed 21HX | Polypure 22S+ | Rexsys 27H | VIE-21A | ANOVA | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| HD | HDF | HD | HDF | HD | HDF | HD | HDF | Dialyzer | Mode | Dial* Mode | |||||||||
| mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | ||||
| RR (%) | 31.3 | (2.5) | 60.5 | (2.5) | 25.7 | (2.6) | 59.5 | (2.5) | 53.3 | (2.7) | 75.0 | (2.5) | 44.2 | (2.5) | 68.5 | (2.6) | <0.001 | <0.001 | NS |
| RR-eq (%) | 22.9 | (2.1) | 47.5 | (2.1) | 18.4 | (2.2) | 46.2 | (2.1) | 41.1 | (2.3) | 61.4 | (2.1) | 32.6 | (2.1) | 55.1 | (2.2) | <0.001 | <0.001 | NS |
| Kt/V | 0.27 | (0.04) | 0.66 | (0.04) | 0.21 | (0.04) | 0.63 | (0.04) | 0.55 | (0.04) | 0.96 | (0.04) | 0.40 | (0.04) | 0.82 | (0.04) | <0.001 | <0.001 | NS |
| RR (%) | 53.3 | (1.7) | 73.8 | (1.7) | 50.4 | (1.7) | 72.2 | (1.7) | 58.8 | (1.7) | 76.2 | (1.6) | 59.7 | (1.7) | 77.4 | (1.7) | <0.001 | <0.001 | NS |
| RR-eq (%) | 40.8 | (1.6) | 60.3 | (1.6) | 38.2 | (1.6) | 58.3 | (1.6) | 45.7 | (1.6) | 62.7 | (1.5) | 46.1 | (1.6) | 64.1 | (1.6) | <0.001 | <0.001 | NS |
| Kt/V | 0.53 | (0.03) | 0.93 | (0.03) | 0.48 | (0.03) | 0.88 | (0.03) | 0.62 | (0.03) | 1.00 | (0.03) | 0.63 | (0.03) | 1.03 | (0.03) | <0.001 | <0.001 | NS |
| RR (%) | 22.4 | (2.7) | 47.3 | (2.7) | 19.3 | (2.8) | 45.0 | (2.7) | 35.2 | (2.8) | 67.7 | (2.6) | 30.2 | (2.7) | 59.1 | (2.8) | <0.001 | <0.001 | NS |
| RR-eq (%) | 16.2 | (2.1) | 35.7 | (2.1) | 13.8 | (2.2) | 33.7 | (2.1) | 26.0 | (2.2) | 54.0 | (2.1) | 21.7 | (2.1) | 46.2 | (2.2) | <0.001 | <0.001 | NS |
| Kt/V | 0.19 | (0.03) | 0.45 | (0.03) | 0.15 | (0.03) | 0.42 | (0.03) | 0.31 | (0.03) | 0.78 | (0.03) | 0.25 | (0.03) | 0.63 | (0.03) | <0.001 | <0.001 | NS |
| RR (%) | 14.8 | (3.0) | 15.0 | (3.0) | 11.9 | (3.1) | 17.4 | (3.0) | 7.6 | (3.1) | 11.9 | (2.9) | 17.6 | (3.0) | 14.8 | (3.1) | NS | NS | NS |
| RR-eq (%) | 10.7 | (2.1) | 10.7 | (2.1) | 8.5 | (2.2) | 12.5 | (2.1) | 5.3 | (2.2) | 8.5 | (2.1) | 12.4 | (2.1) | 10.5 | (2.2) | NS | NS | NS |
| Kt/V | 0.12 | (0.03) | 0.12 | (0.03) | 0.09 | (0.03) | 0.14 | (0.03) | 0.06 | (0.03) | 0.10 | (0.02) | 0.14 | (0.03) | 0.11 | (0.03) | NS | NS | NS |
| RR (%) | 5.4 | (2.1) | 10.6 | (2.1) | 3.8 | (2.3) | 9.2 | (2.1) | 6.8 | (2.2) | 11.4 | (2.1) | 7.9 | (2.1) | 12.6 | (2.2) | NS | 0.002 | NS |
| RR-eq (%) | 3.9 | (1.5) | 7.5 | (1.5) | 2.7 | (1.6) | 6.5 | (1.5) | 4.7 | (1.6) | 8.1 | (1.5) | 5.5 | (1.5) | 8.9 | (1.6) | NS | 0.002 | NS |
| Kt/V | 0.05 | (0.02) | 0.08 | (0.02) | 0.03 | (0.02) | 0.07 | (0.02) | 0.05 | (0.02) | 0.09 | (0.02) | 0.06 | (0.02) | 0.10 | (0.02) | NS | 0.003 | NS |
| RR (%) | Rexsys, VIE-A > Leoceed, Polypure | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| RR-eq (%) | Rexsys > VIE-A > Leoceed, Polypure | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| Kt/V | Rexsys > VIE-A > Leoceed, Polypure | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| RR (%) | Rexsys, VIE-A > Leoceed, Polypure | 0.008 | 0.006 | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| RR-eq (%) | Rexsys, VIE-A > Leoceed, Polypure | 0.011 | 0.005 | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| Kt/V | Rexsys, VIE-A > Leoceed, Polypure | 0.011 | 0.004 | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| RR (%) | Rexsys > VIE-A > Leoceed, Polypure | 0.003 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| RR-eq (%) | Rexsys > VIE-A > Leoceed, Polypure | 0.003 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| Kt/V | Rexsys > VIE-A > Leoceed, Polypure | 0.004 | <0.001 | <0.001 | <0.001 | <0.001 | <0.001 | ||||||||||||
| RR (%) | — | — | — | — | — | — | — | ||||||||||||
| RR-eq (%) | — | — | — | — | — | — | — | ||||||||||||
| Kt/V | — | — | — | — | — | — | — | ||||||||||||
| RR (%) | — | — | — | NS | NS | NS | NS | ||||||||||||
| RR-eq (%) | — | — | — | NS | NS | NS | NS | ||||||||||||
| Kt/V | — | — | — | NS | NS | NS | NS | ||||||||||||
Values were described using mean (sd).
In the ANOVA column, Dialyzer means dialyzer effect; Mode means mode effect and Dial * Mode means Interaction between dialyzer and mode.
Bonferroni post-hoc tests correspond to dialyzers effect whatever the mode was. ANOVA Mode corresponds to ANOVA testing dialyzers effect according to mode. ANOVA Dialyzer corresponds to ANOVA testing mode effects according to dialyzers. (BTP, β Trace Protein; FLCκ, Free Light Chain kappa).
Figure 1FLCκ performances ((a) reduction rates RR and (b) Kt/V) with the 4 tested dialyzers using HD and HDF modalities. (FLCκ, Free Light Chain kappa; RR reduction ratio; skm, simplified kinetic model). (□ HD) (■ on line HDF) (*p < 0.001 vs HD mode).
Figure 2Correlations of the extent of medium molecule removal (reduction rates, %) to their molecular weight with HD or on line HDF technique using the four tested dialyzers. (β2m, beta2 microglobulin; MyoG, myoglobin; FLCκ, Free Light Chain kappa; BTP, Beta trace Protein; MyoST, myostatin; Oroso, orosomucoid). (□ HD) (■ on line HDF).
Biocompatibility parameters with the 4 tested dialyzers using HD and HDF modalities.
| Parameter | Leoceed 21HX | Polypure 22S+ | Rexsys 27H | VIE-21A | ANOVA | ||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| HD | HDF | HD | HDF | HD | HDF | HD | HDF | Dialyzer | Mode | Dial* Mode | |||||||||
| mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | mean | (sd) | ||||
| Albumin (g/L) | 36.1 | (0.8) | 38.5 | (0.8) | 38.0 | (0.8) | 37.6 | (0.8) | 37.1 | (0.8) | 37.8 | (0.8) | 36.2 | (0.8) | 37.9 | (0.8) | NS | NS | NS |
| Transthyretin (g/L) | 0.298 | (0.020) | 0.321 | (0.020) | 0.310 | (0.020) | 0.314 | (0.020) | 0.333 | (0.021) | 0.315 | (0.020) | 0.307 | (0.020) | 0.320 | (0.021) | NS | NS | NS |
| CRP (mg/L) | 23.7 | (8.6) | 6.8 | (8.6) | 11.2 | (8.6) | 4.2 | (8.6) | 11.4 | (8.9) | 3.6 | (8.3) | 29.9 | (8.6) | 11.5 | (8.9) | NS | 0.042 | NS |
| Orosomucoid (g/L) | 1.37 | (0.09) | 1.05 | (0.09) | 1.20 | (0.09) | 0.98 | (0.09) | 1.24 | (0.09) | 0.96 | (0.09) | 1.36 | (0.09) | 0.98 | (0.09) | NS | <0.001 | NS |
| IL-6 (Pre) (pg/mL) | 9.8 | (2.0) | 5.5 | (2.0) | 4.3 | (2.0) | 3.9 | (2.0) | 5.1 | (2.0) | 3.6 | (1.9) | 9.4 | (2.0) | 5.4 | (2.0) | NS | NS | NS |
| IL-6 (Post) (pg/mL) | 9.9 | (1.8) | 6.2 | (1.8) | 4.6 | (1.8) | 3.7 | (1.8) | 3.6 | (1.9) | 3.1 | (1.7) | 7.4 | (1.8) | 4.3 | (1.8) | NS | NS | NS |
| TNF-α (Pre) (pg/mL) | 29.3 | (1.9) | 29.5 | (1.9) | 28.6 | (1.9) | 27.6 | (1.9) | 28.5 | (2.0) | 25.3 | (1.9) | 28.9 | (1.9) | 27.9 | (2.0) | NS | NS | NS |
| TNF-α (Post) (pg/mL) | 13.0 | (1.3) | 11.9 | (1.3) | 12.9 | (1.4) | 11.8 | (1.3) | 13.0 | (1.4) | 11.0 | (1.3) | 14.0 | (1.4) | 10.6 | (1.4) | NS | NS | NS |
Values were described using mean (sd).
In the ANOVA column, Dialyzer means dialyzer effect; Mode means mode effect and Dial * Mode means Interaction between dialyzer and mode.
Figure 3Bland & Altman bias plot for albumin loss measured from direct dialysis quantification using total (TDC) and partial (PDC) spent dialysate collection. For technical reasons, this comparative quantification was evaluated in only half of the patients from center 1 (n = 4). Dotted line represents the mean difference and solid lines represent 95% limits of agreement. All values are expressed in g/session.
Figure 4Dialysate loss of albumin in post-dilution HDF with the 4 tested dialyzers. Continuous partial sampling of spent dialysate with a collection pump inserted into the dialysate outlet line was carried out in all patients from center 1 during the midweek session of each four treatment phases. At the end of the session, a sample was collected after stirring and albumin determined. (*p < 0.05 vs Polypure S+; $p < 0.05 vs Leoceed).
Characteristics of the membranes in the tested dialyzers.
| Membrane | Effective surface area | Inner diameter | Wall thickness | UF coefficient | |
|---|---|---|---|---|---|
| Leoceed 21HX | polysulfone | 2.1 | 200 | 35 | 94 |
| Polypure 22S+ | polysulfone | 2.2 | 200 | 40 | 74 |
| Rexsys 27H | polyethersulfone | 2.7 | 200 | 30 | 124 |
| VIE-21A | vitamin E-coated polysulfone | 2.1 | 185 | 45 | 89 |
Figure 5Study design and sequences of randomization per center. In this prospective multicenter randomized comparative cross-over trial in two parallel arms, HD vs on line HDF, all the patients were sequentially dialyzed with four different dialyzers (one/week): Leoceed 21HX, Polypure 22S+, Rexsys 27H and VIE-21A.