Literature DB >> 27791259

Are all dialyzers compatible with the convective volumes suggested for postdilution online hemodiafiltration?

Jacky Potier1, Guillaume Queffeulou1, Julien Bouet1.   

Abstract

INTRODUCTION: The recent analysis of 4 randomized controlled trials has confirmed the lower mortality risk for postdilution online hemodiafiltration (OL-HDF) compared to hemodialysis, and above all for patients with the highest delivered body surface area standardized convective volume (CV/BSA >23 L/1.73 m²/session). Since the impact of the dialyzers used in these trials has never been studied, we retrospectively analyzed clinical tests carried out with 19 commonly used dialyzers. The aim was to provide information on their performances and behavior to aid in an objective choice for therapies associated with OL-HDF.
METHODS: "Efficiency" was evaluated by measuring the reduction ratio of beta-2 microglobulin (RRß2M) and myoglobin (RRmyo) for a CV/BSA between 0 and 30 L, extrapolating them at CV/BSA = 23 L. "Safety" was defined by the safe CV (CVsafe), corresponding to the CV/BSA above which albumin loss is >5 g/session.
RESULTS: With CV/BSA = 23 L, all the dialyzers ensure an optimal ß2M extraction (RRß2M: 76%-84.5%). For myoglobin, efficiency disparities are bigger (RRmyo: 40%-85%). Above all, 4/19 dialyzers lose more than 5 g albumin and should not be used under these conditions.
CONCLUSIONS: It is recommended to prescribe dialyzers that are above all safe. Moreover, if one considers that the removal of middle molecule solutes prevails over the CV necessary for their transmembrane transport, some dialyzers that are more efficient with CV/BSA <23 L than others with CV/BSA >= 23 L might then be prescribed when the conditions do not permit the suggested CV/BSA goal to be achieved.

Entities:  

Mesh:

Substances:

Year:  2016        PMID: 27791259     DOI: 10.5301/ijao.5000525

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  4 in total

1.  Medium cut-off membranes - closer to the natural kidney removal function.

Authors:  Carina Zweigart; Adriana Boschetti-de-Fierro; Michael Hulko; Lars-Göran Nilsson; Werner Beck; Markus Storr; Bernd Krause
Journal:  Int J Artif Organs       Date:  2017-05-26       Impact factor: 1.595

2.  Early Light Chains Removal and Albumin Levels with a Double Filter-Based Extracorporeal Treatment for Acute Myeloma Kidney.

Authors:  Gabriele Donati; Fulvia Zappulo; Elisa Maietti; Anna Scrivo; Lorenzo Gasperoni; Elena Zamagni; Paola Tacchetti; Lucia Pantani; Olga Baraldi; Giorgia Comai; Maria Cappuccilli; Michele Cavo; Gaetano La Manna
Journal:  Toxins (Basel)       Date:  2022-06-05       Impact factor: 5.075

3.  Randomised trial on clinical performances and biocompatibility of four high-flux hemodialyzers in two mode treatments: hemodialysis vs post dilution hemodiafiltration.

Authors:  Marion Morena; Caroline Creput; Mouloud Bouzernidj; Annie Rodriguez; Lotfi Chalabi; Bruno Seigneuric; Céline Lauret; Anne-Sophie Bargnoux; Anne-Marie Dupuy; Jean-Paul Cristol
Journal:  Sci Rep       Date:  2019-12-04       Impact factor: 4.379

4.  Albumin losses during hemodiafiltration: all dialyzers are not created equal - a case report.

Authors:  Charles Cuvelier; Michel Tintillier; Gabriela Migali; Charlotte Van Ende; Jean-Michel Pochet
Journal:  BMC Nephrol       Date:  2019-10-28       Impact factor: 2.388