| Literature DB >> 31754418 |
Mijoo Kim1, Jaewon Kim2, Hang-Nga Mai3, Tae-Yub Kwon4, Yong-Do Choi3,5, Cheong-Hee Lee1, Du-Hyeong Lee1,3.
Abstract
PURPOSE: The present study was designed to examine the clinical fit of fixed dental prosthesis fabricated by the milling-sintering method using a presintered cobalt-chromium alloy.Entities:
Keywords: Alloy; Clinical; Dental; Fixed dental prosthesis; Marginal discrepancy; Presintered
Year: 2019 PMID: 31754418 PMCID: PMC6856311 DOI: 10.4047/jap.2019.11.5.280
Source DB: PubMed Journal: J Adv Prosthodont ISSN: 2005-7806 Impact factor: 1.904
Fig. 1Workflow of this study.
Fig. 2Representative designs of custom abutment (A) and prosthetic coping (B) of implant-supported metalceramic prosthesis.
Fig. 3Contraction phenomenon of presintered cobaltchromium soft alloy during the sintering process.
Materials used for fabricating metal-ceramic prostheses
| Material | Brand | Manufacturer | Composition (wt%) |
|---|---|---|---|
| Core | Soft Metal | LHK, Chilgok, Korea | Co 63.4, Cr 29, Mo 5.8, Si 0.8, other elements < 1 |
| 4-all | Ivoclar Vivadent, Schaan, Liechtenstein | Ni 61.4, Cr 25.7, Mo 11.0, Si 1.5, Mn < 1.0, Al < 1.0, C < 1.0 | |
| Porcelain | VITA VMK Master | Vita Zahnfabrik, Bad Säckingen, Germany |
Fig. 4Custom abutment and prosthesis placement. (A) Custom abutment, (B) Metal-ceramic prosthesis made by milling-sintering method using a presintered alloy, (C) Metal-ceramic prosthesis made by conventional casting method using a casting alloy.
Fig. 5Representative images for marginal discrepancy analyses using the triple-scan technique. (A) Image merging of scan images, (B) Cross-sectional image showing absolute marginal discrepancy.
Fig. 6Absolute marginal discrepancy of the prosthesis fabricated using milling-sintering and casting methods in each patient.
| (bar): Difference in marginal discrepancy between the prostheses.
Absolute marginal discrepancy of metal-ceramic prostheses fabricated by different fabrication methods (µm)
| Buccal | Lingual | Mesial | Distal | Mean | |
|---|---|---|---|---|---|
| Milling-Sintering | 94.2 ± 34.7 | 74.7 ± 31.4 | 115.1 ± 35.3 | 78.6 ± 24.3 | 90.6 ± 21.3 |
| Casting | 74.0 ± 49.6 | 95.8 ± 24.0 | 106.8 ± 35.3 | 112.1 ± 31.7 | 97.2 ± 25.9 |
| .782 |
P value is calculated by Mann-Whitney U test (α = .05).