Yifan Lu1,2, Jay C Wang1,2,3, Rebecca Zeng2, Raviv Katz2, Demetrios G Vavvas2,3, Joan W Miller2,3, John B Miller1,2,3.
Abstract
PURPOSE: Optical coherence tomography angiography (OCT-A) has emerged as a novel tool for the non-invasive imaging and evaluation of the retinal microvasculature. There is little existing literature that compares OCT-A microvasculature metrics across different OCT-A devices in chorioretinal diseases. Herein, we examined these metrics on three available OCT-A platforms. PATIENTS AND METHODS: All subjects were scanned on each of three OCT-A devices: Optovue Avanti Angiovue, Topcon DRI-OCT Triton Swept-Source OCT, and Zeiss Cirrus 5000-HD-OCT Angioplex. Two investigators independently measured foveal avascular zone (FAZ) area. Superficial capillary plexus (SCP) and deep capillary plexus (DCP) vessel densities (VD) were calculated from binarized images with ImageJ software. Image quality across devices was qualitatively compared. Interclass correlation coefficient (ICC), Bland-Altman analysis, repeated measures ANOVA, and post-hoc tests were performed for statistical analysis.Entities:
Keywords: Angioplex; Angiovue; OCTA; Optovue; Triton; Zeiss
Year: 2019 PMID: 31749603 PMCID: PMC6816077 DOI: 10.2147/OPTH.S215322
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Comparison Of 3 OCT-A Devices
| Device | Cirrus HD-OCT Angioplex | RTvue XR Avanti Angiovue | DRI OCT-1 Triton |
|---|---|---|---|
| Manufacturer | Carl Zeiss Meditec | Optovue | Topcon |
| OCTA Algorithm | OMAG | SSADA | OCTARA |
| Scanning Speed | 68,000 A-scans/sec | 70,000 A-scans/sec | 100,000 A-scans/sec |
| Repeated B-scan Count | 2 (6 x 6), 4 (3 x 3) | 2 | 4 |
| Motion correction | Real-time tracking | Real-time tracking and post-scan orthogonal registration | Real-time tracking |
| OCTA scanning protocols | 3 x 3, 6 x 6 mm centered on macula | 3 x 3, 6 x 6, 8 x 8 mm centered on macula | 3 x 3, 4.5 x 4.5, 6 x 6 mm centered on either macula or optic nerve head |
| FDA Approval | Yes | Yes | No |
Abbreviatons: OMAG, optical micro-angiography; SSADA, split-spectrum amplitude decorrelation angiography; OCTARA, OCTA ratio analysis.
Patient Demographic Information And Chorioretinal Disease Diagnosis
| Patient | Age & Gender | Eye | Diagnosis |
|---|---|---|---|
| 1 | 81, F | R | Dry AMD |
| L | Wet AMD | ||
| 2 | 53, M | R | Aneurismal Type 1 NV |
| L | CSR | ||
| 3 | 80, M | R | CSR vs AMD |
| 4 | 55, M | R | MacTel2 |
| L | MacTel2 | ||
| 5 | 67, M | R | AMD, quiescent CNVM |
| L | AMD, quiescent CNVM | ||
| 6 | 66, F | R | MacTel2 |
| L | MacTel2 | ||
| 7 | 68, F | R | Wet AMD |
| L | Wet AMD/CSR |
Abbreviations: AMD, age-related macular degeneration; Aneurismal Type 1 NV, aneurismal type 1 neovascularization; CSR, central serous retinopathy; MacTel2, macular telangiectasia type 2; CNVM, choroidal neovascular membranes.
Figure 1SCP images and segmentation using Optovue RTVue-XR, Topcon DRI-OCT Triton Swept-Source OCT, and Zeiss Cirrus 5000-HD-OCT Angioplex.
Figure 2SCP images after binarization using the Niblack automated local thresholding method.
SCP FAZ Measurements And ICC
| Machine | SCP FAZ | ICC |
|---|---|---|
| Triton | 0.161 ± 0.103 | |
| Angiovue | 0.161 ± 0.106 | 0.95 |
| Angioplex | 0.164 ± 0.100 |
Figure 3(A) Bland-Altman plot showed agreement between the two investigators for the SCP FAZ area measurements. (B–D) Bland-Altman plot showing the level of agreement between FAZ area measurements for devices. (B) Triton vs Angiovue. (C) Triton vs Angioplex. (D) Angiovue vs Angioplex.
SCP And DCP VD Measurements With Respect To Individual Devices From Binarized Images Using The Niblack Automated Local Thresholding Method
| Machine | SCP Mean VD | DCP Mean VD |
|---|---|---|
| Triton | 0.340 ± 0.014 | 0.334 ± 0.009 |
| Angiovue | 0.307 ± 0.013 | 0.332 ± 0.025 |
| Angioplex | 0.352 ± 0.014 | 0.364 ± 0.012 |
Post Hoc Comparisons Of SCP VD, DCP VD, And FAZ
| Machine | Mean Difference | SE | p value | ||
|---|---|---|---|---|---|
| Triton | Angiovue | SCP VD | 0.033 | 0.003 | < 0.001 |
| DCP VD | 0.002 | 0.007 | 1 | ||
| FAZ | 1.1e -4 | 0.005 | 1 | ||
| Triton | Angioplex | SCP VD | −0.012 | 0.004 | 0.041 |
| DCP VD | −0.030 | 0.004 | < 0.001 | ||
| FAZ | −0.003 | 0.006 | 1 | ||
| Angiovue | Angioplex | SCP VD | −0.045 | 0.005 | < 0.001 |
| DCP VD | −0.032 | 0.008 | 0.006 | ||
| FAZ | −0.004 | 0.006 | 1 | ||