| Literature DB >> 32461263 |
Xinwen Yao1,2,3, Mengyuan Ke3, Yijie Ho4, Emily Lin3, Damon W K Wong1,2,3, Bingyao Tan1,2,3, Leopold Schmetterer2,3,5,6,7,8,9, Jacqueline Chua10,8.
Abstract
BACKGROUND/IMS: To compare the retinal vessel diameter measurements obtained from the swept-source optical coherence tomography angiography (OCTA; Plex Elite 9000, Carl Zeiss Meditec, USA) and adaptive optics ophthalmoscope (AOO; RTX1, Imagine Eyes, France).Entities:
Keywords: angiogenesis; imaging; retina
Year: 2020 PMID: 32461263 PMCID: PMC7907556 DOI: 10.1136/bjophthalmol-2020-316111
Source DB: PubMed Journal: Br J Ophthalmol ISSN: 0007-1161 Impact factor: 4.638
Figure 1Example of the manually registered OCTA 9×9 mm2, 3×3 mm2 and AOO scans on the same eye. (A) The OCTA 9×9 mm2 scan centred at the macular area, marked with the red box representing the corresponding 3×3 mm2 scan displayed in (B) and the light blue box representing the corresponding AOO scan displayed in (C). (D–F) The ROIs for performing vessel diameter measurement indicated by the yellow boxes taken from the (D) OCTA 9×9 mm2, (E) 3×3 mm2 and (F) AOO scans. For the OCTA scans, binarisation masks (green) were imposed on the original images in (D) and (E). Vessel diameter measurements were made perpendicular to the vessel direction, as shown by the red lines. AOO, adaptive optics ophthalmoscope; OCTA, optical coherence tomography angiography; ROIs, regions of interest.
Clinical and ocular characteristics of participants (n=15)
| Characteristics | Mean (SD), n (%) |
| Clinical characteristics | |
| Age (yrs) | 30.87 (6.19) |
| Sex, male | 5 to 33.33 |
| Ethnicity, Chinese | 12 to 80.00 |
| Ocular characteristics | |
| Imaged eye | 27 |
| Right | 14 to 51.85 |
| Left | 13 to 48.15 |
| Spherical equivalent | −3.82 (2.79) |
.
Vessel diameter measurements in two OCTA protocols and AOO
| Unit: µm | AOO | SS-OCTA 3×3 mm2 (OCTA3) | SS-OCTA 9×9 mm2 (OCTA9) |
| Mean±SD (COV) | Mean±SD (COV) | Mean±SD (COV) | |
| Total (n=240) | 50.5±15.6 (30.9) | 55.2±16.3 (29.5) | 54.7±14.3 (26.1) |
| Arteries (n=147) | 51.8±13.9 (26.8) | 56.7±14.5 (25.6) | 56.1±13.2 (23.5) |
| Veins (n=93) | 48.4±18.0 (37.2) | 52.8±18.6 (35.2) | 52.6±15.9 (30.2) |
| D≥45 µm (n=133) | 60.8±13.5 (22.1) | 65.4±14.4 (22.0) | 63.5±13 (20.5) |
| D<45 µm (n=107) | 37.7±5.1 (13.7) | 42.4±6.7 (15.8) | 43.7±5.7 (13.1) |
AOO, adaptive optics ophthalmoscope; COV, coefficient of variance; OCTA, optical coherence tomography angiography; SS-OCTA, swept-source OCTA.
Comparison of mean difference in vessel diameter (µm) measured from two OCTA protocols and AOO
| Unit: µm | Total | Arteries | Veins | D≥45 µm | D<45 µm |
| Mean difference |
|
|
|
|
|
|
|
|
|
|
| |
| ICC (95% CI) | 0.94 (0.92 to 0.95) | 0.91 (0.88 to 0.94) | 0.96 (0.94 to 0.97) | 0.90 (0.86 to 0.92) | 0.68 (0.57 to 0.77) |
| Mean difference |
|
|
|
|
|
|
|
|
|
|
| |
| ICC (95% CI) | 0.93 (0.92 to 0.95) | 0.92 (0.89 to 0.94) | 0.95 (0.92 to 0.96) | 0.90 (0.87 to 0.93) | 0.70 (0.59 to 0.79) |
| Mean difference | −0.5 (−1.3 to 0.4) | −0.6 (−1.8 to 0.5) | −0.2 (−1.6 to 1.1) | − |
|
| 0.298 | 0.301 | 0.729 |
|
| |
| ICC (95% CI) | 0.90 (0.87 to 0.92) | 0.87 (0.82 to 0.90) | 0.93 (0.89 to 0.95) | 0.84 (0.78 to 0.88) | 0.66 (0.54 to 0.76) |
p < 0.05 for values in bold
AOO, adaptive optics ophthalmoscope; ICC, intraclass correlation coefficient; OCTA, optical coherence tomography angiography.
Figure 2Bland-Altman plots of vessel diameter measurements from OCTA 9×9 mm2, 3×3 mm2 and AOO scans. Red dots: arterioles. Blue dots: venules. AOO, adaptive optics ophthalmoscope; OCTA, optical coherence tomography angiography.