Retrospective analysis of reduced energy switching and room switching times on throughput efficiency of a multi-room proton therapy center.

Dennis Mah1, Chin Cheng Chen1, A Omer Nawaz1, Greg Galbreath2, Reuven Shmulenson1, Nancy Lee3, Brian Chon1.   

Abstract

OBJECTIVE: To quantify how a control software upgrade changed beam delivery times and impacted efficiency and capacity of a multiroom proton therapy center.
METHODS: A four-room center treating approximately 90 patients/day, treating for approximately 7 years with optimized operations, underwent a software upgrade which reduced room and energy switching times from approximately 30 to 20 s and approximately 4 s to ~0.5 s, respectively. The center uses radio-frequency identification data to track patient treatments and has software which links this to beam delivery data extracted from the treatment log server. Two 4-month periods, with comparable patient volume, representing periods before and after the software change, were retrospectively analyzed.
RESULTS: A total of 16,168 and 17,102 fields were analyzed. For bilateral head and neck and prostate patients, the beam waiting time was reduced by nearly a factor of 3 and the beam delivery times were reduced by nearly a factor of 2.5. Room switching times were reduced more modestly. Gantry capacity has increased from approximately 30 patients to 40-45 patients in a 16-h daily operation.
CONCLUSIONS: Many proton centers are striving for increased efficiencies. We demonstrated that reductions in energy and room switching time can significantly increase center capacity. Greater potential for further gains would come from improvements in setup and imaging efficiency. ADVANCES IN KNOWLEDGE: This paper provides detailed measured data on the effect on treatment times resulting from reducing energy and room switching times under controlled conditions. It helps validate the models of previous investigations to establish treatment capacity of a proton therapy center.

Entities:  

Mesh:

Year:  2019        PMID: 31746631      PMCID: PMC7066962          DOI: 10.1259/bjr.20190820

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  14 in total

1.  Clinical characterization of a proton beam continuous uniform scanning system with dose layer stacking.

Authors:  J B Farr; A E Mascia; W C Hsi; C E Allgower; F Jesseph; A N Schreuder; M Wolanski; D F Nichiporov; V Anferov
Journal:  Med Phys       Date:  2008-11       Impact factor: 4.071

2.  A comprehensive and efficient daily quality assurance for PBS proton therapy.

Authors:  O Actis; D Meer; S König; D C Weber; A Mayor
Journal:  Phys Med Biol       Date:  2017-02-06       Impact factor: 3.609

3.  Technical Note: An efficient daily QA procedure for proton pencil beam scanning.

Authors:  James E Younkin; Jiajian Shen; Martin Bues; Daniel G Robertson; Daniel W Mundy; Edward Clouser; Wei Liu; Xiaoning Ding; Joshua B Stoker
Journal:  Med Phys       Date:  2018-02-19       Impact factor: 4.071

4.  Spot scanning system for proton radiotherapy.

Authors:  T Kanai; K Kawachi; Y Kumamoto; H Ogawa; T Yamada; H Matsuzawa; T Inada
Journal:  Med Phys       Date:  1980 Jul-Aug       Impact factor: 4.071

5.  Use of treatment log files in spot scanning proton therapy as part of patient-specific quality assurance.

Authors:  Heng Li; Narayan Sahoo; Falk Poenisch; Kazumichi Suzuki; Yupeng Li; Xiaoqiang Li; Xiaodong Zhang; Andrew K Lee; Michael T Gillin; X Ronald Zhu
Journal:  Med Phys       Date:  2013-02       Impact factor: 4.071

6.  Proton facility economics: the importance of "simple" treatments.

Authors:  Peter A S Johnstone; John Kerstiens; Helsper Richard
Journal:  J Am Coll Radiol       Date:  2012-08       Impact factor: 5.532

7.  Quantitative analysis of treatment process time and throughput capacity for spot scanning proton therapy.

Authors:  Kazumichi Suzuki; Matthew B Palmer; Narayan Sahoo; Xiaodong Zhang; Falk Poenisch; Dennis S Mackin; Amy Y Liu; Richard Wu; X Ronald Zhu; Steven J Frank; Michael T Gillin; Andrew K Lee
Journal:  Med Phys       Date:  2016-07       Impact factor: 4.071

8.  Experiences at the Paul Scherrer Institute with a remote patient positioning procedure for high-throughput proton radiation therapy.

Authors:  Alessandra Bolsi; Antony J Lomax; Eros Pedroni; Gudrun Goitein; Eugen Hug
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008-08-01       Impact factor: 7.038

9.  Highly efficient and sensitive patient-specific quality assurance for spot-scanned proton therapy.

Authors:  J E Johnson; C Beltran; H Wan Chan Tseung; D W Mundy; J J Kruse; T J Whitaker; M G Herman; K M Furutani
Journal:  PLoS One       Date:  2019-02-14       Impact factor: 3.240

10.  A novel daily QA system for proton therapy.

Authors:  Xiaoning Ding; Yuanshui Zheng; Omar Zeidan; Anthony Mascia; Wen Hsi; Yixiu Kang; Eric Ramirez; Niek Schreuder; Ben Harris
Journal:  J Appl Clin Med Phys       Date:  2013-03-04       Impact factor: 2.102

View more
  1 in total

1.  Proton therapy special feature: introductory editorial.

Authors:  Kathryn D Held; Antony J Lomax; Esther G C Troost
Journal:  Br J Radiol       Date:  2020-03       Impact factor: 3.039

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.