| Literature DB >> 31739653 |
Min Young Kwak1, Woo Ri Choi1, Jun Woo Park1, Eun Jeong Hwang1, Yeo Ra Ha1, Jong Woo Chung1, Woo Seok Kang1.
Abstract
OBJECTIVES: To investigate the correlation of objective audiometry with user satisfaction as measured with the questionnaire scores.Entities:
Keywords: Audiometry; Hearing Aids
Year: 2019 PMID: 31739653 PMCID: PMC7248607 DOI: 10.21053/ceo.2019.00871
Source DB: PubMed Journal: Clin Exp Otorhinolaryngol ISSN: 1976-8710 Impact factor: 3.372
Changes in hearing test results over 1-, 3-, and 6-month periods
| Variable | Unaided (1st visit) | Aided 1 month | Aided 3 months | Aided 6 months | |
|---|---|---|---|---|---|
| K-HHIE | 31.2±24.3 | 29.7±25.2 | 32.3±25.4 | 39.4±30.8 | 0.046* |
| PTA (dB) | 60.2±11.7 | 37.8±10.7 | 36.8±9.3 | 38.9±8.6 | <0.001* |
| SDS (%) | 46.4±31.0 | 78.8±15.3 | 81.3±12.7 | 82.4±12.0 | <0.001* |
| HINT | 0.4±3.7 | –1.0±2.3 | –1.3±2.1 | –1.2±2.2 | 0.003* |
| CA-f | 16.8±8.8 | 18.6±9.4 | 19.9±8.5 | 19.6±8.6 | 0.069 |
| CA-d | 21.4±6.7 | 21.9±6.8 | 23.9±6.7 | 23.9±6.3 | 0.024* |
| CA-Di | 8.5±8.1 | 14.7±10.7 | 15.2±10.6 | 10.9±12.2 | 0.028* |
Values are presented as mean±standard deviation.
K-HHIE, Korean version of the Hearing Handicap Inventory for the Elderly; PTA, pure tone audiometry; SDS, speech discrimination score; HINT, hearing in noise test; CA-f, central auditory frequency pattern test; CA-d, central auditory duration pattern test; CA-Di, central auditory dichotic test.
Repeated measures analysis of variance (*P<0.05).
Patients’ ages and audiometry results before hearing aid fitting
| Variable | A-HHIE (n=6) | B-HHIE (n=14) | A-IOI (n=12) | B-IOI (n=8) | ||||
|---|---|---|---|---|---|---|---|---|
| Age (yr) | 60.0±13.4 | 62.9±8.4 | –0.597 | 0.558 | 63.7±9.1 | 59.6±11.1 | 0.892 | 0.384 |
| PTA (dB) | 65.8±12.6 | 57.7±10.6 | 1.470 | 0.159 | 60.4±8.9 | 59.9±15.4 | 0.100 | 0.921 |
| SDS (%) | 34.7±30.6 | 51.4±30.9 | –1.115 | 0.279 | 48.7±28.6 | 43.0±36.0 | 0.392 | 0.700 |
| HINT | –0.38±4.89 | 0.69±3.17 | –0.591 | 0.562 | 1.02±4.07 | –0.60±2.95 | 0.963 | 0.348 |
| CA-f | 17.8±9.1 | 16.2±9.0 | 0.351 | 0.730 | 17.9±9.4 | 15.0±8.2 | 0.714 | 0.484 |
| CA-d | 18.5±10.7 | 22.6±4.0 | –1.291 | 0.213 | 22.6±6.7 | 19.6±6.6 | 0.967 | 0.346 |
| CA-Di | 6.0±6.0 | 9.6±8.8 | –0.902 | 0.379 | 8.0±7.8 | 9.3±9.0 | –0.331 | 0.744 |
Values are presented as mean±standard deviation.
HHIE, Hearing Handicap Inventory for the Elderly; IOI, International Outcome Inventory; PTA, pure tone audiometry; SDS, speech discrimination score; HINT, hearing in noise test; CA-f, central auditory frequency pattern test; CA-d, central auditory duration pattern test; CA-Di, central auditory dichotic test.
Independent Student t-test (*P<0.05).
Fig. 1.Objective audiometric test results before and 1, 3, and 6 months after wearing the hearing aid (A-HHIE, six patients; B-HHIE, 14 patients). HHIE, Hearing Handicap Inventory for the Elderly; PTA, pure tone audiometry; SDS, speech discrimination score; HINT, hearing in noise test; SNR, signal-to-noise ratio; CA-f, central auditory frequency pattern test; CA-d, central auditory duration pattern test; CA-Di, central auditory dichotic test. *Repeated measures analysis of variance.
Fig. 2.Objective audiometric test results before and 1, 3, and 6 months after wearing the hearing aid (A-IOI, 12 patients; B-IOI, eight patients). IOI, International Outcome Inventory; PTA, pure tone audiometry; SDS, speech discrimination score; HINT, hearing in noise test; SNR, signal-to-noise ratio; CA-f, central auditory frequency pattern test; CA-d, central auditory duration pattern test; CA-Di, central auditory dichotic test. *Repeated measures analysis of variance.