Literature DB >> 31727754

Outcome Study of the Pipeline Embolization Device with Shield Technology in Unruptured Aneurysms (PEDSU).

D Atasoy1, N Kandasamy2, J Hart2, J Lynch2, S-H Yang3,4, D Walsh5, C Tolias5, T C Booth6,7.   

Abstract

BACKGROUND AND
PURPOSE: The recently introduced Pipeline Flex Embolization Device with Shield Technology (Pipeline Shield) is the third generation of Pipeline flow-diverter devices. It has a new stent-surface modification, which reduces thrombogenicity. We aimed to evaluate clinical and radiographic (safety and efficacy) outcomes of the Pipeline Shield.
MATERIALS AND METHODS: The 30-day and 1-year mortality and morbidity rates and the 6- and 18-month radiographic aneurysm occlusion outcomes for procedures performed between March 2016 and January 2018 were analyzed. 3D-TOF-MRA was used for follow-up.
RESULTS: Forty-four attempted Pipeline Shield procedures were performed for 41 patients with 44 target aneurysms (total of 52 aneurysms treated). A total of 88.5% of devices were inserted in the anterior circulation, and 11.5%, in the posterior circulation; 49/52 (94.2%) aneurysms were saccular; and 1/52 (1.9%) was fusiform. One (1.9%) aneurysm was an iatrogenic pseudoaneurysm, and 1 (1.9%) was a dissecting aneurysm. Seventy-one percent (35/49) of the saccular aneurysms were wide-neck (neck, >4 mm), 34.6% (18/52) were large (≥10 mm), and 3.8% (2/52) were giant (≥25 mm). The mean aneurysm sac maximal diameter was 9.0 mm, and the mean neck width was 5.0 mm. The cumulative mortality and morbidity rates were 2.3% and 6.8% at 1 year, respectively. The adequate occlusion rate was 78.8% at 6 months and 90.3% at 18 months.
CONCLUSIONS: In this pragmatic and non-industry-sponsored study, the occlusion rates and safety outcomes were similar to those seen in previously published studies with flow-diverter devices and earlier generation Pipeline Embolization Devices.
© 2019 by American Journal of Neuroradiology.

Entities:  

Mesh:

Year:  2019        PMID: 31727754      PMCID: PMC6911729          DOI: 10.3174/ajnr.A6314

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  39 in total

1.  Effect of flow diversion with silk on aneurysm size: A single center experience.

Authors:  Lee-Anne Slater; Cathy Soufan; Michael Holt; Winston Chong
Journal:  Interv Neuroradiol       Date:  2015-02       Impact factor: 1.610

2.  Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial.

Authors:  Tibor Becske; David F Kallmes; Isil Saatci; Cameron G McDougall; István Szikora; Giuseppe Lanzino; Christopher J Moran; Henry H Woo; Demetrius K Lopes; Aaron L Berez; Daniel J Cher; Adnan H Siddiqui; Elad I Levy; Felipe C Albuquerque; David J Fiorella; Zsolt Berentei; Miklós Marosfoi; Saruhan H Cekirge; Peter K Nelson
Journal:  Radiology       Date:  2013-02-15       Impact factor: 11.105

Review 3.  Flow-diverter devices for intracranial aneurysms: systematic review and meta-analysis.

Authors:  Ignacio Arrese; Rosario Sarabia; Rebeca Pintado; Miguel Delgado-Rodriguez
Journal:  Neurosurgery       Date:  2013-08       Impact factor: 4.654

4.  Flow diversion in the treatment of aneurysms: a randomized care trial and registry.

Authors:  Jean Raymond; Jean-Christophe Gentric; Tim E Darsaut; Daniela Iancu; Miguel Chagnon; Alain Weill; Daniel Roy
Journal:  J Neurosurg       Date:  2016-11-04       Impact factor: 5.115

5.  Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device.

Authors:  I Szikora; Z Berentei; Z Kulcsar; M Marosfoi; Z S Vajda; W Lee; A Berez; P K Nelson
Journal:  AJNR Am J Neuroradiol       Date:  2010-02-11       Impact factor: 3.825

6.  Pipeline-assisted coiling versus pipeline in flow diversion treatment of intracranial aneurysms.

Authors:  A Sweid; E Atallah; N Herial; H Saad; N Mouchtouris; G Barros; M R Gooch; S Tjoumakaris; H Zarzour; D Hasan; N Chalouhi; R H Rosenwasser; P Jabbour
Journal:  J Clin Neurosci       Date:  2018-10-24       Impact factor: 1.961

7.  Experience using pipeline embolization device with Shield Technology in a patient lacking a full postoperative dual antiplatelet therapy regimen.

Authors:  Kirill Orlov; Dmitry Kislitsin; Nikolay Strelnikov; Vadim Berestov; Anton Gorbatykh; Timur Shayakhmetov; Pavel Seleznev; Anton Tasenko
Journal:  Interv Neuroradiol       Date:  2018-01-29       Impact factor: 1.610

8.  Silent ischemic events after Pipeline embolization device: a prospective evaluation with MR diffusion-weighted imaging.

Authors:  Leonardo B C Brasiliense; Morgan A Stanley; Sanjeet S Grewal; Harry J Cloft; Eric Sauvageau; Giuseppe Lanzino; David Miller; David F Kallmes; Ricardo Hanel
Journal:  J Neurointerv Surg       Date:  2016-01-08       Impact factor: 5.836

9.  Aneurysm Study of Pipeline in an Observational Registry (ASPIRe).

Authors:  David F Kallmes; Waleed Brinjikji; Edoardo Boccardi; Elisa Ciceri; Orlando Diaz; Rabih Tawk; Henry Woo; Pascal Jabbour; Felipe Albuquerque; Rene Chapot; Alain Bonafe; Shervin R Dashti; Josser E Delgado Almandoz; Curtis Given; Michael E Kelly; DeWitte T Cross; Gary Duckwiler; Nasser Razack; Ciaran J Powers; Sebastian Fischer; Demetrius Lopes; Mark R Harrigan; Daniel Huddle; Raymond Turner; Osama O Zaidat; Luc Defreyne; Vitor Mendes Pereira; Saruhan Cekirge; David Fiorella; Ricardo A Hanel; Pedro Lylyk; Cameron McDougall; Adnan Siddiqui; Istvan Szikora; Elad Levy
Journal:  Interv Neurol       Date:  2016-05-26

10.  Treatment of intracranial aneurysms using the pipeline flex embolization device with shield technology: angiographic and safety outcomes at 1-year follow-up.

Authors:  Mario Martínez-Galdámez; Saleh M Lamin; Konstantinos G Lagios; Thomas Liebig; Elisa F Ciceri; Rene Chapot; Luc Stockx; Swarupsinh Chavda; Christoph Kabbasch; Giuseppe Faragò; Hannes Nordmeyer; Thierry Boulanger; Mariangela Piano; Edoardo P Boccardi
Journal:  J Neurointerv Surg       Date:  2018-09-27       Impact factor: 5.836

View more
  6 in total

1.  Hydrophilic Surface Coating of Flow Diverters: A Possible Way to Omit Dual Platelet Aggregation Inhibition.

Authors:  Ansgar Berlis
Journal:  Cardiovasc Intervent Radiol       Date:  2020-08       Impact factor: 2.740

2.  Postprocedural Thrombosis following Endovascular Treatment of Intracranial Aneurysm with Flow Diverters or Coiling: A Histologic Study.

Authors:  J A Larco; M Abbasi; Y Liu; D Dai; G Lanzino; L E Savastano; H Cloft; D F Kallmes; R Kadirvel; W Brinjikji
Journal:  AJNR Am J Neuroradiol       Date:  2021-12-16       Impact factor: 3.825

3.  The Aneurysm Occlusion Assistant, an AI platform for real time surgical guidance of intracranial aneurysms.

Authors:  Kyle A Williams; Alexander R Podgorsak; Mohammad Mahdi Shiraz Bhurwani; Ryan A Rava; Kelsey N Sommer; Ciprian N Ionita
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2021-02-15

4.  Outcome of Flow Diverters with Surface Modifications in Treatment of Cerebral Aneurysms: Systematic Review and Meta-analysis.

Authors:  Y-L Li; A Roalfe; E Y-L Chu; R Lee; A C O Tsang
Journal:  AJNR Am J Neuroradiol       Date:  2020-12-31       Impact factor: 3.825

Review 5.  Advances in endovascular aneurysm management: flow modulation techniques with braided mesh devices.

Authors:  Jessica K Campos; Barry Cheaney Ii; Brian V Lien; David A Zarrin; Chau D Vo; Geoffrey P Colby; Li-Mei Lin; Alexander L Coon
Journal:  Stroke Vasc Neurol       Date:  2020-03-25

6.  Periprocedural to 1-year safety and efficacy outcomes with the Pipeline Embolization Device with Shield technology for intracranial aneurysms: a prospective, post-market, multi-center study.

Authors:  Hal Rice; Mario Martínez Galdámez; Markus Holtmannspötter; Laurent Spelle; Konstantinos Lagios; Maria Ruggiero; Pedro Vega; Hemant Sonwalkar; René Chapot; Saleh Lamin
Journal:  J Neurointerv Surg       Date:  2020-06-01       Impact factor: 5.836

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.