Abstract
PURPOSE: Performance standards for quantitative 18F-FDG PET/CT studies are provided by the EANM Research Ltd. (EARL) to enable comparability of quantitative PET in multicentre studies. Yet, such specifications are not available for 68Ga. Therefore, our aim was to evaluate 68Ga-PET/CT quantification variability in a multicentre setting.Entities:
Keywords: 68Gallium PET/CT; Harmonisation; Image quality; Quantification
Year: 2019 PMID: 31705215 PMCID: PMC6841905 DOI: 10.1186/s40658-019-0253-z
Source DB: PubMed Journal: EJNMMI Phys ISSN: 2197-7364
Acquisition and reconstruction settings of clinical 68Ga PET/CT imaging for prostate cancer and neuroendocrine tumours. One hospital per row is presented
| Site | PET/CT system | Reconstruction settings | Prostate cancer | Neuroendocrine tumours | ||||
|---|---|---|---|---|---|---|---|---|
| Minutes per bed position | Injected activity | Minutes per bed position | Injected activity | |||||
| A | Philips Gemini TOF 64 | BLOB-OS-TF 4 mm 3i33ss | Pelvis: 4 | Body: 3 | 1.5 MBq/kg (range 50–250 MBq) | < 90 kg: 2.5 | > 90 kg: 3.5 | 2.6 MBq/kg (range 100–160 MBq) |
| B | Philips Gemini TF and XL | Astonish iterative reconstruction | 4 | 2.0 MBq/kg | 4 | 2.6 MBq/kg | ||
| C | Siemens mCT Flow | TrueX + TOF 2i21ss Gaussian 5mm | 1.5 mm/s CTM | 2.0 MBq/kg | 2.5 | 100 MBq | ||
| D | Philips Ingenuity TF | BLOB-OS-TF 4 mm 3i33ss 2 mm smooth B filter | NA | 4 | < 90 kg: 150 MBq > 90 kg: 200 MBq | |||
| E | Siemens mCT TrueV | OSEM3D, TOF + PSF 2i21ss Gaussian 5 mm | 4 | 1.5 MBq/kg (min 80 MBq) | NA | |||
| F | Philips Gemini TOF | BLOB-OS-TF 4 mm 3i33ss | Pelvis: 3 | Body: 2 | 100 MBq | 2.5 | 100 MBq | |
| G | Siemens mCT | TrueX + TOF 4i21ss Gaussian 5 mm | 3 | 1.5 MBq/kg | 3 | 1.5 MBq/kg | ||
| H | Siemens mCT40 and mCT128 | TrueX + TOF 3i21ss Gaussian 3 mm | < 70 kg: 1.5 MBq/kg: 3 1.13 MBq/ml: 4 0.9 MBq/ml: 5 | > 70 kg: 1.5 MBq/kg: 4 1.2 MBq/ml: 5 1 MBq/ml: 6 | 1.5 MBq/kg | < 70 kg: 1.5 MBq/kg: 3 1.13 MBq/ml: 4 0.9 MBq/ml: 5 | > 70 kg: 1.5 MBq/kg: 4 1.2 MBq/ml: 5 1 MBq/ml: 6 | 1.5 MBq/kg |
NA = not applicable, i = iteration, ss = subsets, TOF = time-of-flight, PSF = point-spread-function, CTM = continuous table motion
PET/CT reconstruction settings for phantom measurements
| No. | Manufacturer | PET/CT system | Reconstruction | Iterations | Subsets | Filter size | Matrix | Voxel size | Slice thickness |
|---|---|---|---|---|---|---|---|---|---|
| 1 | Siemens | Biograph mCT 40 (1) | PFS + TOF | 3 | 21 | 7.00 | 256 × 256 | 3.18 | 3 |
| 2 | Siemens | Biograph mCT 40 (2) | PFS + TOF | 3 | 21 | 7.00 | 256 × 256 | 3.18 | 3 |
| 3 | Siemens | mCT 123 X3R | Back projection | – | – | 5.00 | 200 × 200 | 4.07 | 5 |
| 4 | Siemens | Biograph mCT Flow 20 | PFS + TOF | 2 | 21 | 5.00 | 200 × 200 | 4.07 | 2.027 |
| 5 | GE | VCT | 3D IR† | NS | NS | NS | 128 × 128 | 5.47 | 3.27 |
| 6 | GE | Discovery D690 | VPFXS* | 4 | 8 | NS | 192 × 192 | 3.65 | 3.27 |
| 7 | Philips | Gemini TOF | BLOB-OS-TF | 3 | 31 | NS | 144 × 144 | 4 | 4 |
| 8 | Philips | Gemini TOF BigBore | BLOB-OS-TF | 3 | 31 | NS | 144 × 144 | 4 | 4 |
| 9 | Philips | Ingenuity | BLOB-OS-TF | 3 | 31 | NS | 169 × 169 | 4 | 4 |
| 10 | Philips | Vereos | BLOB-OS-TF | 3 | 15 | 3.00 | 144 × 144 | 4 | 4 |
| 11 | GE | Discovery 710 | VPFX§ | NS | NS | NS | 256 × 256 | 2.73 | 3.27 |
| 12 | Siemens | mCT 40 | PFS + TOF | 3 | 21 | 6.50 | 256 × 256 | 3.18 | 2 |
| 13 | Siemens | mCT 64 | PFS + TOF | 3 | 21 | 6.50 | 256 × 256 | 3.18 | 2 |
TOF or TF = time-of-flight, PSF = point-spread-function, NS = not specified
†3D OSEM
*3D OSEM with TOF and PSF
§3D OSEM with TOF
Fig. 1Accuracy of the measured activity by the PET/CT system and local dose calibrator, based on the average between the cylindrical and IQ phantom. Numbers correspond to Table 2
Fig. 2Noise across the cylindrical phantom filled with 68Ga, visualized as coefficient of variation (CoV)
Fig. 3RC for 18F with the current EARL standards and RC of 68Ga. Solid lines: maximum and minimum values according to EARL limits as applicable before 2019
Fig. 468Ga RC-curves corrected for the 18F/68Ga calibration mismatch according to local cross-calibration. Solid lines: maximum and minimum values according to EARL limits as applicable before 2019