Andres Kaalep1, Marc Huisman2, Terez Sera3,4, Danielle Vugts2, Ronald Boellaard5,6.
Abstract
PURPOSE: The aim of this study was to investigate the variability in quantitative performance and feasibility of quantitative harmonisation in 89Zr PET/CT imaging.Entities:
Keywords: 89Zr; EARL accreditation; Harmonisation; PET/CT; Performance; Quantification
Year: 2018 PMID: 30460577 PMCID: PMC6246752 DOI: 10.1186/s40658-018-0226-7
Source DB: PubMed Journal: EJNMMI Phys ISSN: 2197-7364
Fig. 1Transversal slice of 89Zr-filled NEMA NU2-2007 body phantom
Reconstruction settings
| System | System | Pixel spacing (mm) | Slice thickness (mm) | Reconstruction method | Post filter width (mm) | Resolution recovery |
|---|---|---|---|---|---|---|
| A | Discovery 690 | 3.65 | 3.27 | VPFXS | 6.4 | Yes |
| B | Discovery 690 | 2.73 | 3.27 | VPHD | 6.5 | No |
| C | Discovery 710 | 2.73 | 3.27 | VPFX | 6.4 | No |
| D | Discovery 710 | 2.73 | 3.27 | VPFXS | 9.0 | Yes |
| E | Biograph 40 mCT | 3.18 | 2.00 | PSF+TOF | 7.0 | Yes |
| F | Somatom Definition AS mCT | 3.18 | 5.00 | PSF+TOF | 8.0 | Yes |
| G | Ingenuity TF PET/CT | 4.00 | 4.00 | BLOB-OS-TF | 4.0 | No |
| H | Biograph 64 mCT | 4.07 | 5.00 | OSEM3D+TOF | 5.0 | No |
Fig. 2SUV accuracy of the PET scanners relative to the dose calibrator measurements plotted by system number. Systems 1–2 and 3–4 represent GE scanners Discovery 690 and Discovery 710, respectively; systems 5–6 are Siemens Biograph 40_mCT and Somatom Definition AS_mCT; system 7 is the Philips Ingenuity TF PET/CT; system 8 is the Siemens Biograph 64_mCT
Fig. 3RC curves derived from the 89Zr phantom experiments using SUVmean (a), SUVmax (b) and SUVpeak (c) quantitative metrics and corresponding RC curves derived from the EARL [18F]FDG phantom experiments using SUVmean (d), SUVmax (e) and SUVpeak (f) quantitative metrics. Current EARL specifications for [18F]FDG-PET/CT accreditation are presented as bold dashed lines. Systems 1–2 and 3–4 represent GE scanners Discovery 690 and Discovery 710, respectively; systems 5–6 are Siemens Biograph 40_mCT and Somatom Definition AS_mCT; system 7 is the Philips Ingenuity TF PET/CT; system 8 is the Siemens Biograph 64_mCT
Fig. 4Background SUV bias-corrected RC curves derived from the phantom experiments using SUVmean (a), SUVmax (b) and SUVpeak (c) quantitative metrics. Current EARL specifications for [18F]FDG-PET/CT accreditation are presented as bold dashed lines. Systems 1–2 and 3–4 represent GE scanners Discovery 690 and Discovery 710, respectively; systems 5–6 are Siemens Biograph 40_mCT and Somatom Definition AS_mCT; system 7 is the Philips Ingenuity TF PET/CT; system 8 is the Siemens Biograph 64_mCT
Fig. 5RC curves normalised to the largest sphere, derived from the phantom experiments using SUVmean (a), SUVmax (b) and SUVpeak (c) quantitative metrics. Systems 1–2 and 3–4 represent GE scanners Discovery 690 and Discovery 710, respectively; systems 5–6 are Siemens Biograph 40_mCT and Somatom Definition AS_mCT; system 7 is the Philips Ingenuity TF PET/CT; system 8 is the Siemens Biograph 64_mCT