Daniel A Coronado-Delgado1, Héctor M Garnica-Garza2. 1. Centro de Investigacion y de Estudios Avanzados Unidad Monterrey Apodaca, Nuevo León, Mexico. 2. CINVESTAV IPN Via del Conocimiento 201 Parque PIIT Ciudad de Mexico, Mexico.
Abstract
Using Monte Carlo simulation and a realistic patient model, it is shown that the volume of healthy tissue irradiated at therapeutic doses can be drastically reduced using a combination of standard megavoltage and kilovoltage X-ray beams with a contrast agent previously loaded into the tumor, without the need to reduce standard treatment margins. Four-dimensional computed tomography images of 2 patients with a centrally located and a peripherally located tumor were obtained from a public database and subsequently used to plan robotic stereotactic body radiotherapy treatments. Two modalities are assumed: conventional high-energy stereotactic body radiotherapy and a treatment with contrast agent loaded in the tumor and a kilovoltage X-ray beam replacing the megavoltage beam (contrast-enhanced radiotherapy). For each patient model, 2 planning target volumes were designed: one following the recommendations from either Radiation Therapy Oncology Group (RTOG) 0813 or RTOG 0915 task group depending on the patient model and another with a 2-mm uniform margin determined solely on beam penumbra considerations. The optimized treatments with RTOG margins were imparted to the moving phantom to model the dose distribution that would be obtained as a result of intrafraction motion. Treatment plans are then compared to the plan with the 2-mm uniform margin considered to be the ideal plan. It is shown that even for treatments in which only one-fifth of the total dose is imparted via the contrast-enhanced radiotherapy modality and with the use of standard treatment margins, the resultant absorbed dose distributions are such that the volume of healthy tissue irradiated to high doses is close to what is obtained under ideal conditions.
Using Monte Carlo simulation and a realistic patient model, it is shown that the volume of healthy tissue irradiated at therapeutic doses can be drastically reduced using a combination of standard megavoltage and kilovoltage X-ray beams with a contrast agent previously loaded into the tumor, without the need to reduce standard treatment margins. Four-dimensional computed tomography images of 2 patients with a centrally located and a peripherally located tumor were obtained from a public database and subsequently used to plan robotic stereotactic body radiotherapy treatments. Two modalities are assumed: conventional high-energy stereotactic body radiotherapy and a treatment with contrast agent loaded in the tumor and a kilovoltage X-ray beam replacing the megavoltage beam (contrast-enhanced radiotherapy). For each patient model, 2 planning target volumes were designed: one following the recommendations from either Radiation Therapy Oncology Group (RTOG) 0813 or RTOG 0915 task group depending on the patient model and another with a 2-mm uniform margin determined solely on beam penumbra considerations. The optimized treatments with RTOG margins were imparted to the moving phantom to model the dose distribution that would be obtained as a result of intrafraction motion. Treatment plans are then compared to the plan with the 2-mm uniform margin considered to be the ideal plan. It is shown that even for treatments in which only one-fifth of the total dose is imparted via the contrast-enhanced radiotherapy modality and with the use of standard treatment margins, the resultant absorbed dose distributions are such that the volume of healthy tissue irradiated to high doses is close to what is obtained under ideal conditions.
Tumor intrafraction motion represents perhaps the most serious impediment to fully
exploiting the recent innovations in technology and treatment strategies in the field of
radiotherapy. Traditionally, the most straightforward and perhaps most widely available
solution to this problem has been the use of generous treatment margins such that the gross
target volume (GTV) is fully encompassed by the radiation beam, regardless of its position
in the motion cycle.[1] This solution carries the obvious disadvantage that large volumes of healthy tissue,
the planning target volume (PTV), are irradiated to the same absorbed dose level as the GTV
itself. To address this shortcoming and try to reduce the treatment margins, several
technologies have been implemented which are capable of monitoring in real time the position
of the GTV. These technologies can be classified into 2 broad categories: static
synchronized delivery and dynamic radiologic tumor tracking. Among the former, respiratory
gating relies on the use of a breathing monitor placed on the patient surface and usually
optical in nature. The motion of the GTV is assumed to correlate with that of the breathing
monitor, and only when the latter is in a position previously determined as suitable to
carry out the treatment, the radiation beam is turned on. This effectively synchronizes the
treatment delivery to the respiratory cycle.[2] One problem with this approach is that the internal position of the target must be
inferred from surface motions. Another problem is the reduced efficiency, as the overall
treatment time is increased. Dynamic tumor tracking, on the other hand, aims to alleviate
these problems by continuously monitoring the target position during the irradiation, either
using sets of radiological images,[3] 2 at least, or a combination of a single image and a target motion model.[4,5] By continuously monitoring the location of the GTV, adjustments to the position and
direction from which the radiation beam is aimed to the target are made in order to account
for the GTV motion. While these technologies represent state-of-the-art solutions to the
intrafraction motion problem, they introduce an extra layer of complexity to the already
challenging problem posed by the irradiation of an internal structure with what for
practical purposes is an invisible radiation beam. From this perspective, the simplicity of
the generous margin strategy remains unrivaled, and therefore it is worth exploring
techniques that might lead to its improvement with regard to the large amount of healthy
tissue irradiated to therapeutic doses without resorting to the minimization of the PTV
itself. In this context, we hypothesize that if the radiation absorption properties of the
GTV were somehow modified in order to create a larger dose difference with respect to the
surrounding tissue, the size of the PTV margin could be kept as large as the motion
amplitude would dictate, as the surrounding tissue would never reach the same dose level as
the GTV. This is precisely the rationale behind contrast-enhanced radiotherapy (CERT),[6,7] where a radiological contrast agent previously loaded into the tumor, where it tends
to accumulate, enhances the absorption of X-rays at the kilovoltage energy range, thus
creating a large dose gradient between the GTV and the surrounding tissues.In this work, we will indeed show that a stereotactic body radiotherapy (SBRT) lung
treatment with recommended PTV margins can be dramatically improved and that in fact can be
made to resemble what would be obtained under ideal conditions of no tumor movement and
minimum PTV margin by imparting a fraction of the total dose under the CERT treatment
scheme.
Materials and Methods
Patient Models
For this study, 2 patient models, both with non-small cell lung tumors, were obtained
from The Cancer Imaging Archive[8-10] in the form of 4-dimensional computed tomography (4DCT) images with 10 breathing
phases. The resolution of these 4DCT images is 1 × 1 mm with a slice thickness of 3 mm,
and the GTV, lungs, heart, and other major structures were already segmented at each phase
in the respiratory cycle. Patient model A presents a centrally located GTV in the right
lung having a volume of 30.3 cm3. Patient model B on the other hand has a
peripherally located GTV with a volume of 31.1 cm3. This second model in
particular was chosen as a portion of the ribs, highly efficient in absorbing kilovoltage
X-ray beams, is located near the GTV in such a way that the PTV overlaps with it. Coronal
views of both models are shown in Figure
1. As the skin and bone structures are also of interest for our purposes, a
separate software, 3D Slicer (www.slicer.org), was used to delineate the
skin, ribs, and sternum on the whole set of 4DCT images. Following the recommendations of
RTOG 0813 and RTOG 0915, the PTV margins for both the patients were contoured directly
around the GTV and having dimensions of 1 cm in the craniocaudal direction and 0.5 cm in
the lateral direction.[11,12] In order to simulate an ideal treatment, that is, a treatment in which there is no
intrafraction motion, and therefore the PTV margin can be as small as the beam penumbra
will allow, separate patient models were generated with PTV margins of 2 mm in all
directions using only the CT images obtained at the deep inspiration stage. While an
average image over the whole set of 4DCT images would perhaps be a more appropriate
representation, this resulted in blurred organs and structures.
Figure 1.
Coronal views of the 2 patient models used in this work. Several relevant structures
were already contoured, including the gross target volume (GTV), both lungs, heart,
and trachea.
Coronal views of the 2 patient models used in this work. Several relevant structures
were already contoured, including the gross target volume (GTV), both lungs, heart,
and trachea.
Treatment Planning
It is assumed that the treatment is delivered under conditions of robotic SBRT.[3] A total of 300 circular beams with a 2 cm diameter for patient model A and 1 cm
diameter for patient model B were used for each treatment, and each beam could be turned
on and off by the optimization algorithm, described below. These diameters were selected
after comparing treatments with different beam sizes. A separate software was used to
determine the irradiation points and orientation for each beam, tailored to each patient
model, as described previously.[13] Only ipsilateral irradiations were allowed in order to minimize the unnecessary
exposure of healthy tissue and the same irradiation points and directions were used for
both the high-energy and the CERT treatments. Figure 2 shows the irradiation setup for one of the
treatment cases examined in this work. For each patient model, the following plans were
calculated, in all cases using the deep inspiration CT images:
Figure 2.
Irradiation setup for the treatment plans modeled in this work. A semispherical shell
with 80-cm radius from the geometric center of the gross target volume (GTV) was used
to lay a grid of dwell positions. The right panel shows actual irradiation points and
directions. The geometric center of the tumor is located at 5.7, 15.3, and 17.0
cm.
Irradiation setup for the treatment plans modeled in this work. A semispherical shell
with 80-cm radius from the geometric center of the gross target volume (GTV) was used
to lay a grid of dwell positions. The right panel shows actual irradiation points and
directions. The geometric center of the tumor is located at 5.7, 15.3, and 17.0
cm.A CyberKnife-like ideal treatment plan with a uniform PTV margin of 2 mm. This was
the minimum possible margin without degrading the absorbed dose distribution in the
GTV. This plan served as the reference plan.A high-energy treatment plan with the PTV margin as recommended by the RTOG 0813
protocol.A CERT plan with the margin recommended in RTOG 0813 and RTOG 0915.A set of plans combining the treatment plans 2 and 3 mentioned earlier with different
proportions of the total prescribed dose imparted by each modality, from 80% MV to 20%
CERT to 50% of the prescribed dose by each modalityEach of these plans was optimized according to the same set of prescription goals as
recommended in either RTOG 0813 or RTOG 0915 and listed in Table 1. The simultaneous projection feasibility
algorithm of Cimmino[14] was used to carry out the optimization. The application of this algorithm in
radiotherapy is described in Censor et al,[15] and our implementation has been discussed before.[13]
Table 1.
Prescription Goals Used in the Optimization of the Treatment Plans.a
a DLOW and DUP refer to the minimum and maximum
dose limits imposed to each structure.
Prescription Goals Used in the Optimization of the Treatment Plans.aAbbreviations: GTV, gross target volume; PTV, planning target volume.a DLOW and DUP refer to the minimum and maximum
dose limits imposed to each structure.
Absorbed Dose Calculations
Conversion from Hounsfield units to material composition
In order to define the different materials present in the patient model, a calibration
curve that converts CT numbers into elemental weight data using 12 different materials
as the basis was used.[16] Computer scripts were developed to read the CT images, convert the voxel data
into its respective material type, and write the corresponding files in the format
needed by the Monte Carlo code, described below. The material representing the contrast
agent embedded in the tumor tissue was treated separately, as it is not actually present
in the patient models used: It is assumed that the contrast agent is based on gold
nanoparticles (GNP)[6] and that its concentration in the GTV is 10 mg-Au/g, a concentration deemed feasible.[17] The tumor is assumed to consist of soft tissue with the weight fraction of each
material adjusted to incorporate gold at the concentration stated earlier. The contrast
agent was assumed to be present both in the high-energy and the CERT treatment, although
in the former it resulted in no discernible effect.
Monte Carlo simulations
The Monte Carlo code PENELOPE[18] and its accompanying set of subroutines from the PenEasy implementation[19] were used to carry out all the absorbed dose calculations reported in this work.
Separate calculations were run for each of the 300 beams participating in the 2 modeled
treatments. In order to determine the impact of the intrafraction motion, the absorbed
dose imparted by each beam was calculated at every phase of the respiratory cycle;
therefore, for each beam energy and treatment, a total of 3000 dose matrices were
obtained. Following the recommendations of Task Group 268,[20] the relevant description of the software and transport parameters is shown in
Table 2. The CERT
treatment was imparted using a 220-kVp X-ray beam produced by a tungsten target and
filtered by 2 mm of copper, while the megavoltage beam model was taken from the
literature, and it is based on the Monte Carlo modeling of a CyberKnife treatment machine.[21] Our full implementation of both X-ray sources into the PENELOPE software has been
detailed before.[13] X-rays of 220 kVp were used as we have previously shown that a X-ray beam with
220-kVp spectrum is an optimal compromise between penetration at depth and sizable
fluence in the energy interval that maximized the absorption of the incident beam by the GNPs.[22]
Table 2.
Relevant Monte Carlo Information Per TG-268.
Parameter
Value
Monte Carlo code
PENELOPE-2006
Cross-sections
Built-in analytical models and tabulated data
Transport parameters
Eab = 10 keV (photons and electrons); C1 = C2 = 0.1;
Wcc = Wcr = 100 eV
Statistical uncertainty
2% on average for those voxels receiving at least 50% of the maximum
dose
Relevant Monte Carlo Information Per TG-268.
Incorporation of the Effect of Intrafraction Tumor Motion Into the Treatment
Delivery
In order to determine the effect that the intrafraction motion has on the resultant
absorbed dose distributions for the calculated treatment plans, the optimized plans were
imparted to the moving phantom at each stage in the respiratory cycle. This is done by
calculating the dose distribution from each of the 300 beams, weighted according to the
optimization results and adding the dose matrices. Note that, as the beams would be
sequentially delivered, we would have to know in advance the particular sequence in which
the patient is irradiated. To avoid this problem and make our conclusions as general as
possible, each of the 10 moving phantoms was irradiated with the 300 beams, and it was
then assumed that each stage in the breathing cycle received an equal proportion,
one-tenth, of the total prescribed dose. The treatments thus obtained will be referred to
as the imparted treatments.
Dose Accumulation Through the Motion Cycle
The open source medical image informatics software 3D Slicer was used to perform B-spline
deformable image registration throughout the whole respiratory cycle (www.slicer.org).
Separate computer scripts were implemented to read the vector field output files yielded
by 3D Slicer and use them to correlate voxel indices among the CT images at each stage in
the respiratory cycle. This correlation in turn was used to add the dose matrices as
yielded by PENELOPE and our optimization software for each of the treatments modeled in
this work.
Results
Patient Model A
Isodose curves on the coronal plane are shown in Figure 3 for each treatment modeled in this work,
namely, ideal 6 MV and imparted, that is, accounting for intrafraction motion, 6-MV, and
CERT treatments. Note that for the ideal high-energy treatment, the 30-Gy isodose curve
almost spills into the contralateral lung, something that does not occur for the
CERT-imparted treatment. In general, the isodose curves for the CERT treatment, in spite
of patient motion, are more tightly wrapped around the GTV.
Figure 3.
Isodose curves on the coronal plane for the 3 treatments modeled in this work: upper
left panel—ideal 6 MV treatment; right panel—imparted 6 MV; and bottom panel—imparted
contrast-enhanced radiotherapy (CERT).
Isodose curves on the coronal plane for the 3 treatments modeled in this work: upper
left panel—ideal 6 MV treatment; right panel—imparted 6 MV; and bottom panel—imparted
contrast-enhanced radiotherapy (CERT).Figure 4 shows the cumulative
dose–volume histogram (cDVH) for the target volumes resulting from the imparted
high-energy treatment plan, in the left panel, and for the CERT plan on the right
panel.
Figure 4.
Cumulative dose-volume histogram (cDVHs) for the imparted treatment plans in each of
the modalities modeled in this work. Target coverage in each treatment adheres to the
recommendations set forth in RTOG 0813.
Cumulative dose-volume histogram (cDVHs) for the imparted treatment plans in each of
the modalities modeled in this work. Target coverage in each treatment adheres to the
recommendations set forth in RTOG 0813.The cDVHs for the ideal megavoltage treatment are shown as a reference. As can be seen
from this figure and regardless of patient motion, 95% of the PTV and 100% of the GTV
receives the prescribed dose of 50 Gy, which is basically the purpose of the treatment
margin recommended in RTOG 0813. While the dose distribution in the PTV is similar among
the ideal and imparted treatment, it must be kept in mind that the ideal margin is smaller
than the margin recommended by RTOG 0813. The ideal treatment of course also offers an
appropriate target coverage but with the obvious advantage of a total PTV about half that
recommended by the RTOG 0813 protocol. Table 3 shows the lung volume, including both
lungs, irradiated to at least 2 particular maximum dose levels as stated in RTOG 0813. In
all cases and regardless of the treatment modality, the irradiated critical volumes are
well below the 1000 cm3 limit stated in the said protocol.
Table 3.
Total Lung Volume (cm3) Irradiated to at Most 12.5 Gy and 13.5
Gy.a
a Both lungs are included. RTOG 0813 recommended limit at each dose level
is 1000 cm3.
Total Lung Volume (cm3) Irradiated to at Most 12.5 Gy and 13.5
Gy.aAbbreviation: CERT, contrast-enhanced radiotherapy.a Both lungs are included. RTOG 0813 recommended limit at each dose level
is 1000 cm3.Table 4 shows the maximum
absorbed doses imparted to at least 1 cm3 of several irradiated organs at risk
(OAR) by each treatment modality along with the recommended limits for the 50-Gy dose arm
of the protocol, which again regardless of the treatment modality are never exceeded. All
3 treatment plans are therefore acceptable from a clinical perspective.
Table 4.
Maximum Absorbed Dose (Gy) Imparted to at Least 1 cm3 of Selected
Irradiated Structures.a
a This does not consider the 2-cm margin around the PTV as specified by
RTOG 0813.
Maximum Absorbed Dose (Gy) Imparted to at Least 1 cm3 of Selected
Irradiated Structures.aAbbreviation: CERT, contrast-enhanced radiotherapy; PTV, planning target
volume.a This does not consider the 2-cm margin around the PTV as specified by
RTOG 0813.Figure 5 shows the effect of
combining the high-energy and CERT-imparted treatments with various weights. The GTV
coverage, shown in the right panel, is not affected as expected, since the RTOG 0813
treatment margins are large enough to account for tumor motion. Note however that the cDVH
curve for the PTV, shown in the right panel of Figure 5, steadily shifts toward lower dose values,
as the percentage of the total dose imparted via the CERT treatment is increased. As an
example, imparting 40% of the total dose via the CERT modality reduces by more than half
the volume of lung tissue receiving at least 55 Gy.
Figure 5.
Cumulative dose-volume histogram (cDVHs) for the gross target volume (GTV), right
panel, and PTV for different combination of high-energy and contrast-enhanced
radiotherapy (CERT) treatments. Each pair of percentage figures in the labels refer to
the weights of the megavoltage and the CERT treatments in this order.
Cumulative dose-volume histogram (cDVHs) for the gross target volume (GTV), right
panel, and PTV for different combination of high-energy and contrast-enhanced
radiotherapy (CERT) treatments. Each pair of percentage figures in the labels refer to
the weights of the megavoltage and the CERT treatments in this order.Figure 6 shows the total volume
receiving a dose of at least 45 Gy, 50 Gy, and 55 Gy as a function of the percentage of
the total dose delivered by each modality. At the higher dose level even imparting
one-fifth of the total dose with the CERT modality results in a treatment volume
irradiated at 55 Gy that is on par with what is obtained under ideal conditions of no
tumor movement and minimum PTV margin, although the RTOG PTV volume is larger than the
ideal margin. At lower dose levels, the reduction in the total volume irradiated is not as
drastic as observed at higher doses, but nevertheless a favorable change is clearly
discernible.
Figure 6.
Total planning target volume irradiated to at least 55 Gy, 50 Gy, and 45 Gy. In these
and subsequent graphs, the set of 2 numbers separated by a dash indicate the
percentage of absorbed dose imparted by the high-energy and contrast-enhanced
radiotherapy (CERT) modalities in this order.
Total planning target volume irradiated to at least 55 Gy, 50 Gy, and 45 Gy. In these
and subsequent graphs, the set of 2 numbers separated by a dash indicate the
percentage of absorbed dose imparted by the high-energy and contrast-enhanced
radiotherapy (CERT) modalities in this order.The same trend is observed for other structures, such as the lungs and heart, as shown in
Figure 7. For the lungs in
particular, reductions in the total volume receiving at least 15 Gy ranges from 10% to 25%
for the combinations of megavoltage and CERT treatments examined in this work. Although
the change in the heart volume irradiated at a given dose level, 10 Gy for the plot in the
right panel of Figure 7, is less
pronounced, a downward trend is clearly discernible. It must be kept in mind that all this
is happening without affecting the coverage of the GTV as previously shown in Figure 5 and with the use of standard
PTV margins which ensure that the GTV is fully covered regardless of its position in the
motion cycle. Perhaps most important, no assumptions regarding the knowledge of the
particular position of the GTV in the motion cycle are made. By combining a megavoltage
and a CERT treatment, even with a modest fraction of the total dose imparted by the latter
modality, we are able to obtain a treatment plan that closely resembles what would be
obtained under ideal conditions of no tumor motion and beam penumbra-determined PTV
margin.
Figure 7.
Left panel: total volume of lung receiving at least 15 Gy, including both lungs;
right panel: volume of heart irradiated to at least 10 Gy. The 2 numbers separated by
a dash on the x-axis indicate the percentage of absorbed dose imparted by the
high-energy and contrast-enhanced radiotherapy (CERT) modalities, respectively.
Left panel: total volume of lung receiving at least 15 Gy, including both lungs;
right panel: volume of heart irradiated to at least 10 Gy. The 2 numbers separated by
a dash on the x-axis indicate the percentage of absorbed dose imparted by the
high-energy and contrast-enhanced radiotherapy (CERT) modalities, respectively.On the other hand, as shown in Figure
8, the volume of skin and ribs irradiated to a given level of absorbed dose
increases as the fraction of the total dose delivered by CERT increases. For the skin in
particular, splitting the total dose at 80% to 20% ratio between the high-energy and CERT
modalities, respectively, almost doubles the skin volume irradiated at 5 Gy. It should be
pointed out however that, as shown in Table 3, even if 100% of the prescribed dose was imparted via the CERT modality,
the maximum skin dose would not exceed the RTOG 0813 stated limit of 36 Gy. The same
applies to the ribs, whose maximum dose is well below the 40 Gy limit stated in the RTOG
protocol. These are simply the negative consequences of the depth dose characteristics and
interaction physics of kilovoltage X-ray beams that are, however, greatly reduced when
combining CERT and high-energy irradiations in a single treatment strategy.
Figure 8.
Volume of skin and ribs irradiated at doses of 5 Gy and 10 Gy, respectively, as a
function of the total dose fraction imparted with high-energy and contrast-enhanced
radiotherapy (CERT).
Volume of skin and ribs irradiated at doses of 5 Gy and 10 Gy, respectively, as a
function of the total dose fraction imparted with high-energy and contrast-enhanced
radiotherapy (CERT).
Patient Model B
Figure 9 shows isodose curves on
the coronal plane for the 3 treatments modeled, namely, 6 MV ideal, 6 MV imparted, and
CERT imparted. Again, in all 3 treatments, 100% of the GTV receives the prescribed dose of
48 Gy, so the margin as recommended by RTOG 0915 fully compensates for intrafraction
motion.
Figure 9.
Isodose curves on the coronal plane for the 3 treatments modeled in this work: upper
left panel—ideal 6 MV treatment; right panel—imparted 6 MV; bottom panel—imparted
contrast-enhanced radiotherapy (CERT).
Isodose curves on the coronal plane for the 3 treatments modeled in this work: upper
left panel—ideal 6 MV treatment; right panel—imparted 6 MV; bottom panel—imparted
contrast-enhanced radiotherapy (CERT).Figure 10 shows the effect that
such a motion has on the target coverage for the high-energy and CERT treatments, using
the ideal 6-MV plan as the reference. As mentioned before, GTV coverage is excellent
independent of the treatment modality and is not affected by the patient motion. For the
high-energy treatment, left panel of Figure 10, and unlike in the patient case previously discussed, a larger
degradation is seen in the absorbed dose distribution of the PTV when comparing the
delivered against the ideal treatment. In spite of this, 97% of the PTV is receiving the
prescribed dose of 48 Gy.
Figure 10.
Cumulative dose-volume histogram (cDVHs) for the imparted treatment plans in each of
the modalities modeled in this work. Target coverage in each treatment adheres to the
recommendations set forth in RTOG 0915.
Cumulative dose-volume histogram (cDVHs) for the imparted treatment plans in each of
the modalities modeled in this work. Target coverage in each treatment adheres to the
recommendations set forth in RTOG 0915.As mentioned before, a portion of the ribs is encompassed by the PTV, and while for the
megavoltage treatment its presence is not clearly discernible from the resultant cDVH, for
the CERT-imparted treatment, right panel in Figure 10, the presence of bone with its high
kilovoltage X-ray absorption efficiency results in a distorted cDVH: At least 5% of the
PTV volume is now receiving absorbed doses exceeding 40 Gy. From the isodose curves show
in Figure 9, however, it is clear
that these high doses are being imparted to bone and not to the healthy lung tissue.
Moreover, the maximum PTV dose is approximately the same as in the ideal high-energy
treatment but with an average PTV dose of 22.5 Gy as opposed to 50.6 Gy for the 6-MV
treatment.Table 5 shows the maximum
absorbed doses imparted to at least 1 cm3 of different OARs, using the
recommendations set forth in RTOG 0915. All treatments modeled are within these
recommendations. For the lungs, heart, and esophagus, the CERT treatment closely
approaches the ideal, static patient and minimum PTV margin, 6-MV treatment. Note again
that CERT imparts doses to the skin and bone, not considering the portion inside the PTV,
ranging from 30% to 100% above the dose delivered by the ideal high-energy treatment but
still within the safety limits recommended by RTOG 0915.
Table 5.
Maximum Absorbed dose (Gy) Imparted to at Least 1 cm3 of Selected
Irradiated Structures.
a Per RTOG 0915, the 2-cm margin around the PTV is not considered.
Maximum Absorbed dose (Gy) Imparted to at Least 1 cm3 of Selected
Irradiated Structures.Abbreviation: CERT, contrast-enhanced radiotherapy; PTV, planning target
volume.a Per RTOG 0915, the 2-cm margin around the PTV is not considered.Figure 11 shows the effect on
both the GTV and the PTV of combining the high-energy and CERT-imparted treatments with
several weights. For the GTV, as the CERT treatment weight is increased, the average dose
also increases, while for the PTV the opposite holds true. From the right panel of Figure 11, it is evident that bone
behaves essentially as a contrast agent loaded into the PTV.
Figure 11.
Cumulative dose-volume histogram (cDVHs) for the gross target volume (GTV), right
panel, and planning target volume (PTV) for different combination of high-energy and
contrast-enhanced radiotherapy (CERT) treatments. Each pair of percentage figures in
the labels refer to the weights of the high-energy and the CERT treatments in this
order.
Cumulative dose-volume histogram (cDVHs) for the gross target volume (GTV), right
panel, and planning target volume (PTV) for different combination of high-energy and
contrast-enhanced radiotherapy (CERT) treatments. Each pair of percentage figures in
the labels refer to the weights of the high-energy and the CERT treatments in this
order.Yet, even if only 20% of the prescribed dose is imparted via the CERT modality, the total
PTV receiving doses at or above 50 Gy is reduced from 95% down to 40%, despite the
inherent patient motion, the large PTV margins and the presence of bone in the margins
used in both treatments. Furthermore, as shown on the left panel of Figure 12, at doses near the prescription level,
there is an important reduction in the planning volume irradiated even when only one-fifth
of the total dose is imparted via the CERT modality. However, the reduction in the volume
receiving intermediate doses, central panel in Figure 12, is more modest. This is the dose level,
as shown un Figure 10, where
absorption by bone prevents the cDVH for the PTV from falling to zero as was the case in
the centrally located tumor. Note however that the average PTV dose, right panel of Figure 11, decreases by at least 10%
when the total dose is split between the 6-MV and CERT modalities at a ratio of 80% to
20%, with even larger reductions for other ratios.
Figure 12.
Total planning target volume (PTV) irradiated to at least 55 Gy and 45 Gy and average
PTV dose as a function of the percentage of the total dose delivered by each
modality
Total planning target volume (PTV) irradiated to at least 55 Gy and 45 Gy and average
PTV dose as a function of the percentage of the total dose delivered by each
modalityFigure 13 shows both the lung,
outside the 2-cm margin defined by RTOG 0915, and the esophagus volume receiving at least
15 Gy and 10 Gy, respectively. For the lung tissue in particular, the reduction in volume
is not as pronounced as for the centrally located tumor because, as mentioned before, the
PTV encompasses not only the lung but a sizable portion of other tissues. The esophagus
does exhibit a reduction of almost 30% in the volume receiving at least 10 Gy when the 6
MV and CERT modalities are combined with different ratios.
Figure 13.
Left panel: total volume of lung receiving at least 15 Gy (both lungs); right panel:
volume of the esophagus irradiated to at least 10 Gy.
Left panel: total volume of lung receiving at least 15 Gy (both lungs); right panel:
volume of the esophagus irradiated to at least 10 Gy.As noted in the centrally located tumor, the absorbed dose imparted to both the skin and
the ribs, shown in Figure 14,
increases as the percentage of total dose imparted via the CERT modality increases, which
as discussed before is a consequence of the physical characteristics of the absorption
process undergone by kilovoltage X-ray beams. It is important to emphasize that according
to Table 5, the tolerances
established by RTOG 0915 are never exceeded.
Figure 14.
Volumes of skin and ribs irradiated to 5 Gy and 10 Gy, respectively.
Volumes of skin and ribs irradiated to 5 Gy and 10 Gy, respectively.
Discussion
There already exists clinical experience related to the irradiation method proposed in this
work: a phase I study to evaluate the use of a CT scanner for the treatment of metastatic
brain tumors upon which an iodinated contrast agent had been previously loaded was carried
out without adverse reactions observed, with treatments imparted with a combination of
megavoltage and CERT modalities.[23] Remarkably, some of the irradiated tumors vanished after 4 treatments. While in the
aforementioned trial the PTV margin did not play a role as the whole brain was being
irradiated, in this work, we have shown an additional advantage of combining megavoltage and
CERT therapies: Its robustness against intrafraction motion as the only restriction placed
on the PTV margin would be that it needs to be large enough to fully encompass the motion
amplitude of the GTV to be irradiated and thus resulting in absorbed dose distributions that
closely follow what would be obtained under ideal conditions.While it has been shown that the presence of bone inside the relatively high-dose PTV
region does not adversely affect GTV coverage in CERT treatments, see Figure 10, the presence of air pockets or low-density
tissue and the ensuing disruption of charged particle equilibrium must also be considered.
We argue that for kilovoltage X-ray beams, for which the secondary electron field has ranges
in the order of micrometers, the presence of low-density tissue or air pockets, particularly
in the periphery of the tumor, should not be as detrimental for the overall treatment plan
as it is for the megavoltage beams.From a radiobiological perspective, we have recently shown that kilovoltage and megavoltage
robotic SBRT for lung result in differences of less than 5% in the tumor control probability
(TCP) and normal tissue complication probability (NTCP),[13] although it is likely that the presence of the contrast agent embedded in the tumor,
not present in the referenced work, would change its biological response to kilovoltage
X-ray beams. This change, in principle, would be beneficial in terms of the TCP as the high
linear energy transfer of the electron cascade released from the high-atomic number atoms in
the contrast agent would result in an increased lethality.[24] As the radiobiological parameters needed to account for the presence of the contrast
agent in the GTV are not known, it is not possible to estimate the TCP for the treatment
plans modeled in this work. We have shown, however, that when combining high-energy and CERT
treatments, the OARs are receiving even lower maximum doses than when no contrast agent was
used as part of the treatment. Based on these findings, we believe that it is reasonable to
expect an even lower NTCP for CERT, perhaps on par with what it is usually obtained for
megavoltage treatments.One critical aspect of the proposed approach, and of CERT in general, is the specificity
with which the GNPs would predominantly accumulate in the tumor and not the healthy tissue
surrounding it. It is clear that conventional iodine-based contrast agent do accumulate
preferentially in the tumor tissue[23] which is important as the higher the specificity, the greater the difference between
the absorbed dose in the tumor and that of the surrounding tissue. While in this type of
contrast agent there is not much room for improving the specificity, nanoparticle-based
agents, through the optimization of the nanoparticle size, do offer this possibility. While
experimental data with mice seem to indicate that 2 nm in diameter GNP offers excellent specificity,[6] in humans, the average pore size of healthy and tumor tissue vasculature, 60 nm and
240 nm, respectively, suggests that GNP with diameter in the order of 100 nm would be better
suited for the purpose of CERT.[25]Several logistic aspects of the proposed method related to the treatment-planning process
would have to be addressed. In particular:Quantification of the contrast agent concentration in the tumor: This is important as
the absorption of kilovoltage X-rays, and therefore the total dose imparted, depends
on the amount of contrast agent present at a given point in the patient. This in
principle should not represent a serious impediment to the proposed method, as the
measurement of the contrast agent concentration can be carried out using the same set
of CT images used for treatment-planning purposes, provided that a suitable
calibration curve has been previously obtained[26] and, of course, that the contrast agent was administered prior to the CT
scanning process.Variation in the contrast agent concentration with time: This is by far the most
challenging problem associated with the proposed method and in general to the CERT
modality. For the particular case of GNP-based contrast agent, it has been reported
that, in mice, the concentration of GNP in the tumor reaches a plateau at about 5
hours postinjection and stays fairly constant for periods of up to 24 hours,[27] with clearance from other tissues and organs at a much faster rate. It is not
clear from this study how the concentration changes 24 hours after the administration.
Therefore, for a single fraction treatment, as in robotic SBRT, variation in contrast
agent concentration with time should not be a problem, as the treatment session
usually lasts between 30 and 90 minutes. For multifraction SBRT, further research
would be needed regarding the dynamics of the GNP in the tumor for time periods of at
least 1 week postadministration, in order to determine whether additional
administration of the GNP would be neededFrom a technological perspective, it has been assumed in this work that a robotic therapy
machine capable of dual mega- and kilovoltage X-ray beam generation is used to deliver the
treatment. While this technology is not currently available in the clinic, its main
components are already present in the treatment room. There are 2 possibilities to obtain
the kilovoltage X-ray beam needed to carry out the irradiation as modeled in this work:Target material and geometry optimization such that the megavoltage incident electron
beam yields a sizable component of kilovoltage X-rays that could then be used for the
purposes of CERT irradiation. This approach has been pursued in the context of portal imaging,[28,29] with the obvious disadvantage that it would be impossible to independently
control the kilovoltage beam quality and output rate.On linac with a gantry-mounted X-ray tube intended for tumor tracking, repurposing
such an X-ray tube in order to obtain the desired spectrum and beam output may be an
option. Note that this is not exactly what was modeled in this work, as we assumed
that both the kilovoltage and the megavoltage beams are aimed from the same position
and with the same direction. However, in light of the results presented in this work,
it is reasonable to conclude that using the already available X-ray tube is a viable
alternative to the proposed methodWe believe a third option that could be implemented in the linacs used in the CyberKnife
systems, as they do not have a flattening filter and bending magnet, would be to
independently accelerate, via DC voltage, a separate electron beam and use the same target,
with a slight modification, to generate both the high- and low-energy beams. This would
allow to independently manipulate the dose contribution from CERT as deemed fit and also to
fine-tune the kilovoltage X-ray spectra to the particular characteristics of the target to
be irradiated. This approach is currently under theoretical analysis at our laboratory,
particularly with regard to the cooling system necessary to dissipate the extra heat
generated in the target material due to the additional electron beam incident on it and also
with regard to the characteristics of the resultant kilovoltage X-ray beam.