Isabel Rodríguez-Pérez1,2, Juan Gros-Otero3, Miguel A Teus4,5, Rafael Cañones5, Montserrat García-González4,3.
Abstract
BACKGROUND: To compare the visual and refractive outcomes after myopic LASEK using three different excimer lasers and standardized surgical and mitomycin C (MMC) application protocols.Entities:
Keywords: Excimer laser; LASEK; Mitomycin C; Myopia
Mesh:
Year: 2019 PMID: 31615462 PMCID: PMC6792330 DOI: 10.1186/s12886-019-1214-y
Source DB: PubMed Journal: BMC Ophthalmol ISSN: 1471-2415 Impact factor: 2.209
Preoperative data for the 394 eyes treated with Allegretto, Esiris and Technolas excimer lasers in a follow-up period of 3 months after myopic LASEK with the adjuvant use of mitomycin C
| Parameter | ALLEGRETTO | ESIRIS | TECHNOLAS | |
|---|---|---|---|---|
| Eyes | 122 | 135 | 137 | |
| Sphere (D) (≤ − 11 D) | −4.10 ± 2.16 | − 4.16 ± 2.20 | −4.23 ± 2.18 | 0.9 |
| Cylinder (D) (≤ −4.5 D) | −0.90 ± 0.96 | − 0.90 ± 0.86 | − 0.89 ± 0.81 | 0.9 |
| CDVA (LogMAR) | −0.04 ± 0.06 | − 0.04 ± 0.05 | − 0.03 ± 0.07 | 0.06 |
| CDVA (Decimal) | 1.12 ± 0.13 | 1.11 ± 0.12 | 1.08 ± 0.10 | 0.05 |
| CCT (μm) | 524 ± 26.00 | 514 ± 31.20 | 520 ± 28.00 | 0.03 |
| Keratometry K1 (D) | 43.12 ± 1.50 | 43.77 ± 1.50 | 43.65 ± 1.50 | 0.0005 |
| Keratometry K2 (D) | 44.10 ± 1.60 | 44.90 ± 1.60 | 44.60 ± 1.50 | 0.0001 |
| Age (years) | 28.60 ± 4.50 | 30.40 ± 5.50 | 31.80 ± 7.40 | 0.0002 |
CDVA corrected distance visual acuity, CCT central corneal thickness, D dioptres
aFactorial ANOVA test
UDVA evolution up to three months after LASEK with MMC of 394 myopic eyes treated with Allegretto, Esiris and Technolas excimer lasers
| Parameter | Follow-up visit | ALLEGRETTO | ESIRIS | TECHNOLAS | |
|---|---|---|---|---|---|
| Eyes | ( | ( | ( | ||
| UDVA (Decimal) | 1 day | 0.64 ± 0.33 | 0.63 ± 0.24 | 0.57 ± 0.21 | 0.2 |
| 1 week | 0.72 ± 0.21 | 0.80 ± 0.24 | 0.81 ± 0.61 | 0.4 | |
| 1 month | 0.94 ± 0.23 | 0.92 ± 0.17 | 0.88 ± 0.19 | 0.07 | |
| 3 months | 1.11 ± 0.19 | 1.01 ± 0.21 | 0.98 ± 0.20 | 0.0001 | |
| UDVA (LogMAR) | 1 day | 0.29 ± 0.38 | 0.23 ± 0.20 | 0.27 ± 0.17 | 0.5 |
| 1 week | 0.16 ± 0.15 | 0.12 ± 0.17 | 0.13 ± 0.17 | 0.4 | |
| 1 month | 0.04 ± 0.13 | 0.04 ± 0.10 | 0.06 ± 0.11 | 0.2 | |
| 3 months | −0.03 ± 0.10 | 0.09 ± 0.11 | 0.01 ± 0.10 | 0.0001 |
UDVA uncorrected distance visual acuity, SD standard deviation
aFactorial ANOVA test
Fig. 1Cumulative histogram of three-month uncorrected distance visual acuity after myopic LASEK with MMC of 394 eyes treated with the Allegretto, Esiris and Technolas excimer lasers. LASEK = Laser-assisted subepithelial keratomileusis; MMC = Mitomycin C; VA = Visual acuity. a- Allegretto, b- Esiris, c-Technolas
Three-months postop outcomes after LASEK with MMC of 394 myopic eyes treated with Allegretto, Esiris and Technolas excimer lasers
| Parameter | ALLEGRETTO | ESIRIS | TECHNOLAS | |
|---|---|---|---|---|
| Eyes | 122 | 135 | 137 | |
| Residual Sphere (D) | −0.01 ± 0.24 | 0.29 ± 0.65 | 0.27 ± 0.58 | 0.0001 |
| Residual Cylinder (D) | −0.06 ± 0.32 | −0.37 ± 0.53 | − 0.26 ± 0.53 | 0.0001 |
| CDVA (LogMAR) | −0.02 ± 0.17 | −0.04 ± 0.10 | − 0.01 ± 0.08 | 0.05 |
| CDVA (Decimal) | 1.13 ± 0.15 | 1.10 ± 0.13 | 1.04 ± 0.17 | 0.0001 |
| Efficacy index | 0.99 ± 0.16 | 0.90 ± 0.18 | 0.91 ± 0.18 | 0.0004 |
| Safety index | 1.00 ± 0.12 | 1.07 ± 0.94 | 0.90 ± 0.14 | 0.03 |
| Change in lines of CDVA | −0.001 ± 0.11 | −0.007 ± 0.10 | − 0.025 ± 0.11 | 0.06 |
CDVA corrected distance visual acuity, D dioptres, SD Standard deviation
aFactorial ANOVA test
Fig. 2Changes in lines of corrected distance visual acuity at 3 months after myopic LASEK with MMC of 394 eyes treated with Allegretto, Esiris and Technolas excimer lasers. LASEK = Laser-assisted subepithelial keratomileusis; MMC = Mitomycin C; CDVA = Corrected distance visual acuity. a- Allegretto, b- Esiris, c-Technolas
Fig. 3Three-month predictability (SE ± 0.50 D and ± 1.00 D) after LASEK with MMC in the Allegretto, Esiris and Technolas groups for myopia correction. LASEK = Laser-assisted subepithelial keratomileusis; MMC = Mitomycin C; SE = Spherical Equivalent. a- Allegretto, b- Esiris, c-Technolas
Fig. 4Attempted versus achieved spherical equivalent refraction scatterplots at 3 months after LASEK with MMC for myopia correction in the Allegretto, Esiris and Technolas groups. The linear regression equation and coefficient of determination (r2) are displayed. LASEK = Laser-assisted subepithelial keratomileusis; MMC = Mitomycin C. a- Allegretto, b- Esiris, c-Technolas
Comparative publications among different excimer platforms for Anterior Surface Ablation/LASIK performance
| Study design | Treatment | MMC 0.02% | UDVA decimal | UDVA LogMAR | Efficacy Index | Residual Sphere | Residual Cylinder | CDVA decimal | CDVA LogMAR | Safety Index | Change in CDVA lines preop/postop | SE ±0.5 D | SE ±1 D | Correlation preopSE/ Corrected refraction | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| ANTERIOR SURFACE ABLATION | |||||||||||||||
| Our Study | Retrospective study. 394 eyes. ≤ 40 y.o. Full range myopia 3-months follow-upU Allegretto 400 Hz vs Esiris vs Technolas 217c. | Standard (no customized) | MMC: 15 s each 50 μm | *Allegretto better than others | *Allegretto better than others | *Allegretto better than others | *Allegretto better than others | *Allegretto better than others | *Allegretto better than others | No differences were found | *Technolas worse than others | No differences were found | *Allegretto better than others | No differences were found | Strong correlation in the 3 groups |
[ JCRS. 2011;37:1858–64. | Prospective study 131 eyes. ≤ 40 y.o. Low to moderate myopia. 3-month follow-up Allegretto 400 Hz vs Technolas 217z. | WFG vs Planoscan | MMC: 10–25 s > 70 μm | Allegretto better than Technolas | No differences were found | No differences were found | Allegretto better than Technolas | ||||||||
[ JRS. 2015;31:683–90. | Prospective study. 151 eyes. ≤ 45 y.o. Moderate myopia. 3-month follow-upU Allegretto 400 Hz vs Technolas 217z | WFO vs Tissue-saving | MMC: 10–25 s > 70 μm | Allegretto better than Technolas | No differences were found | No differences were found | 20/20 (Snellen) in both groups | ||||||||
| LASIK | |||||||||||||||
[ Clin Ophthalmol. 2012; 6:1159–68. | Retrospective study. 442 eyes. ≤ 52 y.o. Full range myopia 3-month follow-upU Allegretto 400 Hz vs Technolas 217z | WFO vs Planoscan Flap creation: Visumax 500 kHz | Allegretto better than Technolas | No differences were found | Allegretto better than Technolas | *Allegretto better than Technolas | No differences were found | No differences were found | No differences were found | No differences were found | |||||
[ J Refract Surg. 2008;24:885–90. | Prospective study 44 eyes. Full range myopia and hyperopia Allegretto eye-Q vs Esiris | Q-value vs aspheric standard Flap creation: Moria M2 | No differences in myopia subgroup were found | No differences in total group were found | No differences in myopia subgroup were found | No differences in myopia subgroup were found | No differences in total group were found | No differences in total group were found | No differences in total group were found | No differences in total group were found | |||||
UDVA uncorrected distance visual acuity; CDVA corrected distance visual acuity, D dioptres, MMC mitomycin C, SE spherical equivalent, y.o.: years old; *statistically significant differences (P < 0.05)