Kananan Utitsarn1,2, Giordano Biasi1, Nauljun Stansook1,3, Ziyad A Alrowaili1,4, Marco Petasecca1, Martin Carolan5, Vladimir L Perevertaylo6, Wolfgang A Tomé7, Tomas Kron1,8,9, Michael L F Lerch1, Anatoly B Rosenfeld1.
Abstract
PURPOSE: We introduce a technique that employs a 2D detector in transmission mode (TM) to verify dose maps at a depth of dmax in Solid Water. TM measurements, when taken at a different surface-to-detector distance (SDD), allow for the area at dmax (in which the dose map is calculated) to be adjusted.Entities:
Keywords: 2D solid-state array detector; MP512; in vivo QA; transmission detector
Mesh:
Year: 2019 PMID: 31609090 PMCID: PMC6839376 DOI: 10.1002/acm2.12744
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1(Upper panel) A snapshot of the active area of the MP512 and protective PMMA slabs. (Lower panel) A schematic of the packaging of the MP512 system, with the MP512 active area proper, upper and lower PMMA protective slabs, and the PCB‐FR4 board on which the detector is wire bounded for signal readout.
Figure 2Use of the MP512 system in transmission mode (not to scale).
Figure 3Central sensitive volume of the MP512: ratio of dose mode measurements to transmission mode measurements as a function of surface‐to‐detector distance, for all considered square fields. Error bars did not exceed symbol size.
Central sensitive volume of the MP512: slope (M) and axis intercept (BA0) of the ratio of dose mode measurements to transmission mode measurements, for all considered square fields; absolute values.
| Parameter | Square field, side [cm] | ||||
|---|---|---|---|---|---|
| 2 | 3 | 5 | 8 | 10 | |
| M | 0.020 | 0.021 | 0.018 | 0.019 | 0.020 |
| BAo | 1.165 | 1.144 | 1.028 | 0.969 | 0.953 |
Figure 4Ratio of measurements in dose mode at dmax to measurements in transmission mode: BAo as a function of radiation field area A. Error bars did not exceed symbol size.
Static square field of 1 and 4 cm side: calculated dose [Gy] at dmax (using transmission mode measurements at 4 and 24 cm SDD) compared with treatment planning system (TPS) calculations and dose mode measurements performed with the MP512, with films and with a Farmer ionization chamber.
| Square field, side [cm] | MP512, calc. | MP512, meas. | TPS | EBT3 | Farmer chamber | |
|---|---|---|---|---|---|---|
| 4 cm SDD | 24 cm SDD | |||||
| 1 | 0.798 | 0.821 | 0.816 | 0.810 | 0.809 | 0.813 |
| 4 | 1.013 | 1.020 | 1.004 | 0.996 | 0.998 | 1.010 |
Intensity‐modulated radiotherapy fields: gamma evaluation for dose calculations at dmax (using transmission mode measurements at 4 and 24 cm SDD) and corresponding treatment planning system calculations and dose mode measurements performed with the MP512 and with films.
| SDD [cm] | Acceptance criteria | ||||||||
|---|---|---|---|---|---|---|---|---|---|
| 3%/3 mm | 2%/2 mm | 1%/1 mm | |||||||
| MP512, calc. vs. TPS | MP512, calc. vs. EBT3 | MP512, calc. vs. MP512, meas. | MP512, calc. vs. TPS | MP512, calc. vs. EBT3 | MP512, calc. vs. MP512, meas. | MP512, calc. vs. TPS | MP512, calc. vs. EBT3 | MP512, calc. vs. MP512, meas. | |
| 4 | 98.14% | 96.89% | 99.79% | 90.50% | 92.00% | 98.59% | 62.20% | 69.40% | 99.40% |
| 24 | 97.22% | 97.53% | 99.69% | 93.80% | 93.80% | 97.69% | 59.00% | 71.00% | 99.00% |