Literature DB >> 31605643

Novel acoustic coupling bath using magnetite nanoparticles for MR-guided transcranial focused ultrasound surgery.

Steven P Allen1, Tom Steeves2, Austin Fergusson3, Dave Moore4, Richey M Davis5, Eli Vlaisialjevich3,6, Craig H Meyer1,7.   

Abstract

PURPOSE: Acoustic coupling baths, nominally composed of degassed water, play important roles during transcranial focused ultrasound surgery. However, this large water bolus also degrades the quality of intraoperative magnetic resonance (MR) guidance imaging. In this study, we test the feasibility of using dilute, aqueous magnetite nanoparticle suspensions to suppress these image degradations while preserving acoustic compatibility. We examine the effects of these suspensions on metrics of image quality and acoustic compatibility for two types of transcranial focused ultrasound insonation regimes: low-duty cycle histotripsy procedures and high-duty cycle thermal ablation procedures.
METHODS: Magnetic resonance guidance imaging was used to monitor thermal ablations of in vitro gel targets using a coupling bath composed of various concentrations of aqueous, suspended, magnetite nanoparticles in a clinical transcranial transducer under stationary and flowing conditions. Thermal deposition was monitored using MR thermometry simultaneous to insonation. Then, using normal degassed water as a coupling bath, various concentrations of aqueous, suspended, magnetite nanoparticles were placed at the center of this same transducer and insonated using high-duty cycle pulsing parameters. Passive cavitation detectors recorded cavitation emissions, which were then used to estimate the relative number of cavitation events per insonation (cavitation duty cycle) and the cavitation dose estimates of each nanoparticle concentration. Finally, the nanoparticle mixtures were exposed to low-duty cycle, histotripsy pulses. Passive cavitation detectors monitored cavitation emissions, which were used to estimate cavitation threshold pressures.
RESULTS: The nanoparticles reduced the MR signal of the coupling bath by 90% in T2- and T2*-weighted images and also removed almost all imaging artifacts caused by coupling bath motion. The coupling baths caused <5% changes in peak temperature change achieved during sonication, as observed via MR thermometry. At low duty cycle insonations, the nanoparticles decreased the cavitation threshold pressure by about 15 ± 7% in a manner uncorrelated with nanoparticle concentration. At high duty cycle insonations, the 0.5 cavitation duty cycle acoustic power threshold varied linearly with nanoparticle concentration.
CONCLUSIONS: Dilute aqueous magnetite nanoparticle suspensions effectively reduced MR imaging artifacts caused by the acoustic coupling bath. They also attenuated acoustic power deposition by <5%. For low duty cycle insonation regimes, the nanoparticles decreased the cavitation threshold by 15 ± 7%. However, for high-duty cycle regimes, the nanoparticles decreased the threshold for cavitation in proportion to nanoparticle concentration.
© 2019 American Association of Physicists in Medicine.

Entities:  

Keywords:  HIFU; MRI; cavitation; focused ultrasound; iron oxide; magnetite

Mesh:

Substances:

Year:  2019        PMID: 31605643      PMCID: PMC6899167          DOI: 10.1002/mp.13863

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

1.  The effect of polymer coatings on proton transverse relaxivities of aqueous suspensions of magnetic nanoparticles.

Authors:  Matthew R J Carroll; Phillip P Huffstetler; William C Miles; Jonathon D Goff; Richey M Davis; Judy S Riffle; Michael J House; Robert C Woodward; Timothy G St Pierre
Journal:  Nanotechnology       Date:  2011-07-19       Impact factor: 3.874

Review 2.  Image-guided ultrasound phased arrays are a disruptive technology for non-invasive therapy.

Authors:  Kullervo Hynynen; Ryan M Jones
Journal:  Phys Med Biol       Date:  2016-08-05       Impact factor: 3.609

Review 3.  Neurosurgical Applications of High-Intensity Focused Ultrasound with Magnetic Resonance Thermometry.

Authors:  Rivka R Colen; Iman Sahnoune; Jeffrey S Weinberg
Journal:  Neurosurg Clin N Am       Date:  2017-10       Impact factor: 2.509

Review 4.  Transcranial MRI-Guided Focused Ultrasound: A Review of the Technologic and Neurologic Applications.

Authors:  Pejman Ghanouni; Kim Butts Pauly; W Jeff Elias; Jaimie Henderson; Jason Sheehan; Stephen Monteith; Max Wintermark
Journal:  AJR Am J Roentgenol       Date:  2015-07       Impact factor: 3.959

Review 5.  Designed synthesis of uniformly sized iron oxide nanoparticles for efficient magnetic resonance imaging contrast agents.

Authors:  Nohyun Lee; Taeghwan Hyeon
Journal:  Chem Soc Rev       Date:  2011-12-02       Impact factor: 54.564

6.  MRI-guided gas bubble enhanced ultrasound heating in in vivo rabbit thigh.

Authors:  S D Sokka; R King; K Hynynen
Journal:  Phys Med Biol       Date:  2003-01-21       Impact factor: 3.609

7.  Effects of f-number on the histotripsy intrinsic threshold and cavitation bubble cloud behavior.

Authors:  Eli Vlaisavljevich; Tyler Gerhardson; Tim Hall; Zhen Xu
Journal:  Phys Med Biol       Date:  2016-12-20       Impact factor: 3.609

8.  Magnetic Nanoclusters with Hydrophilic Spacing for Dual Drug Delivery and Sensitive Magnetic Resonance Imaging.

Authors:  Nipon Pothayee; Sharavanan Balasubramaniam; Nikorn Pothayee; Neeta Jain; Nan Hu; Yinnian Lin; Richey M Davis; Nammalwar Sriranganathan; Alan P Koretsky; J S Riffle
Journal:  J Mater Chem B       Date:  2013       Impact factor: 6.331

9.  Inertial cavitation dose and hemolysis produced in vitro with or without Optison.

Authors:  Wen-Shiang Chen; Andrew A Brayman; Thomas J Matula; Lawrence A Crum
Journal:  Ultrasound Med Biol       Date:  2003-05       Impact factor: 2.998

10.  Unilateral magnetic resonance guided focused ultrasound thalamotomy for essential tremor: practices and clinicoradiological outcomes.

Authors:  Won Seok Chang; Hyun Ho Jung; Eun Jung Kweon; Eyal Zadicario; Itay Rachmilevitch; Jin Woo Chang
Journal:  J Neurol Neurosurg Psychiatry       Date:  2014-05-29       Impact factor: 10.154

View more
  4 in total

1.  Technical Note: Effect of transducer position and ground plane configuration on image quality in MR-guided focused ultrasound therapies.

Authors:  Allison Payne; Emilee Minalga; Robb Merrill; Dennis L Parker; J Rock Hadley
Journal:  Med Phys       Date:  2020-04-06       Impact factor: 4.071

2.  A preclinical study of diffusion-weighted MRI contrast as an early indicator of thermal ablation.

Authors:  Steven P Allen; Francesco Prada; Zhiyuan Xu; Jeremy Gatesman; Xue Feng; Helen Sporkin; Yekaterina Gilbo; Sydney DeCleene; Kim Butts Pauly; Craig H Meyer
Journal:  Magn Reson Med       Date:  2020-11-11       Impact factor: 3.737

3.  The Potential of Adjusting Water Bolus Liquid Properties for Economic and Precise MR Thermometry Guided Radiofrequency Hyperthermia.

Authors:  Kemal Sumser; Gennaro G Bellizzi; Gerard C van Rhoon; Margarethus M Paulides
Journal:  Sensors (Basel)       Date:  2020-05-22       Impact factor: 3.576

Review 4.  Magnetic mediators for ultrasound theranostics.

Authors:  Arkadiusz Józefczak; Katarzyna Kaczmarek; Rafał Bielas
Journal:  Theranostics       Date:  2021-11-02       Impact factor: 11.556

  4 in total