Optimization of the difference-of-Gaussian channel sets for the channelized Hotelling observer.

Christiana Balta1,2, Ramona W Bouwman1, Mireille J M Broeders1,3, Nico Karssemeijer2, Wouter J H Veldkamp4, Ioannis Sechopoulos1,2, Ruben E van Engen1.   

Abstract

The channelized-Hotelling observer (CHO) was investigated as a surrogate of human observers in task-based image quality assessment. The CHO with difference-of-Gaussian (DoG) channels has shown potential for the prediction of human detection performance in digital mammography (DM) images. However, the DoG channels employ parameters that describe the shape of each channel. The selection of these parameters influences the performance of the DoG CHO and needs further investigation. The detection performance of the DoG CHO was calculated and correlated with the detection performance of three humans who evaluated DM images in 2-alternative forced-choice experiments. A set of DM images of an anthropomorphic breast phantom with and without calcification-like signals was acquired at four different dose levels. For each dose level, 200 square regions-of-interest (ROIs) with and without signal were extracted. Signal detectability was assessed on ROI basis using the CHO with various DoG channel parameters and it was compared to that of the human observers. It was found that varying these DoG parameter values affects the correlation ( r 2 ) of the CHO with human observers for the detection task investigated. In conclusion, it appears that the the optimal DoG channel sets that maximize the prediction ability of the CHO might be dependent on the type of background and signal of ROIs investigated.
© 2019 Society of Photo-Optical Instrumentation Engineers (SPIE).

Entities:  

Keywords:  2-alternative forced choice; anthropomorphic breast phantom; channelized-Hotelling observer; difference-of-Gaussian; mammography; model observers

Year:  2019        PMID: 31572746      PMCID: PMC6763759          DOI: 10.1117/1.JMI.6.3.035501

Source DB:  PubMed          Journal:  J Med Imaging (Bellingham)        ISSN: 2329-4302


  18 in total

1.  Additional factors for the estimation of mean glandular breast dose using the UK mammography dosimetry protocol.

Authors:  D R Dance; C L Skinner; K C Young; J R Beckett; C J Kotre
Journal:  Phys Med Biol       Date:  2000-11       Impact factor: 3.609

2.  Monte Carlo calculation of conversion factors for the estimation of mean glandular breast dose.

Authors:  D R Dance
Journal:  Phys Med Biol       Date:  1990-09       Impact factor: 3.609

3.  Optimization of digital breast tomosynthesis (DBT) acquisition parameters for human observers: effect of reconstruction algorithms.

Authors:  Rongping Zeng; Aldo Badano; Kyle J Myers
Journal:  Phys Med Biol       Date:  2017-02-02       Impact factor: 3.609

4.  Evaluating the sensitivity of the optimization of acquisition geometry to the choice of reconstruction algorithm in digital breast tomosynthesis through a simulation study.

Authors:  Rongping Zeng; Subok Park; Predrag Bakic; Kyle J Myers
Journal:  Phys Med Biol       Date:  2015-01-16       Impact factor: 3.609

5.  Addition of a channel mechanism to the ideal-observer model.

Authors:  K J Myers; H H Barrett
Journal:  J Opt Soc Am A       Date:  1987-12       Impact factor: 2.129

6.  Toward image quality assessment in mammography using model observers: Detection of a calcification-like object.

Authors:  Ramona W Bouwman; Alistair Mackenzie; Ruben E van Engen; Mireille J M Broeders; Kenneth C Young; David R Dance; Gerard J den Heeten; Wouter J H Veldkamp
Journal:  Med Phys       Date:  2017-10-13       Impact factor: 4.071

7.  A model observer study using acquired mammographic images of an anthropomorphic breast phantom.

Authors:  Christiana Balta; Ramona W Bouwman; Ioannis Sechopoulos; Mireille J M Broeders; Nico Karssemeijer; Ruben E van Engen; Wouter J H Veldkamp
Journal:  Med Phys       Date:  2017-12-21       Impact factor: 4.071

8.  Human- and model-observer performance in ramp-spectrum noise: effects of regularization and object variability.

Authors:  C K Abbey; H H Barrett
Journal:  J Opt Soc Am A Opt Image Sci Vis       Date:  2001-03       Impact factor: 2.129

9.  Correlation between a 2D channelized Hotelling observer and human observers in a low-contrast detection task with multislice reading in CT.

Authors:  Lifeng Yu; Baiyu Chen; James M Kofler; Christopher P Favazza; Shuai Leng; Matthew A Kupinski; Cynthia H McCollough
Journal:  Med Phys       Date:  2017-07-13       Impact factor: 4.071

10.  Incorporating human contrast sensitivity in model observers for detection tasks.

Authors:  Subok Park; Aldo Badano; Brandon D Gallas; Kyle J Myers
Journal:  IEEE Trans Med Imaging       Date:  2009-03       Impact factor: 10.048

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.