Validation of a large-scale audit technique for CT dose optimisation.

T J Wood1, A W Davis, C S Moore, A W Beavis, J R Saunderson.   

Abstract

The expansion and increasing availability of computed tomography (CT) imaging means that there is a greater need for the development of efficient optimisation strategies that are able to inform clinical practice, without placing a significant burden on limited departmental resources. One of the most fundamental aspects to any optimisation programme is the collection of patient dose information, which can be compared with appropriate diagnostic reference levels. This study has investigated the implementation of a large-scale audit technique, which utilises data that already exist in the radiology information system, to determine typical doses for a range of examinations on four CT scanners. This method has been validated against what is considered the 'gold standard' technique for patient dose audits, and it has been demonstrated that results equivalent to the 'standard-sized patient' can be inferred from this much larger data set. This is particularly valuable where CT optimisation is concerned as it is considered a 'high dose' technique, and hence close monitoring of patient dose is particularly important.

Entities:  

Mesh:

Year:  2011        PMID: 22090412     DOI: 10.1093/rpd/ncr434

Source DB:  PubMed          Journal:  Radiat Prot Dosimetry        ISSN: 0144-8420            Impact factor:   0.972


  4 in total

1.  Adult exposures from MDCT including multiphase studies: first Italian nationwide survey.

Authors:  Federica Palorini; Daniela Origgi; Claudio Granata; Domenica Matranga; Sergio Salerno
Journal:  Eur Radiol       Date:  2014-02       Impact factor: 5.315

2.  Correlation of the clinical and physical image quality in chest radiography for average adults with a computed radiography imaging system.

Authors:  C S Moore; T J Wood; A W Beavis; J R Saunderson
Journal:  Br J Radiol       Date:  2013-04-08       Impact factor: 3.039

3.  Investigating the use of an antiscatter grid in chest radiography for average adults with a computed radiography imaging system.

Authors:  C S Moore; T J Wood; G Avery; S Balcam; L Needler; A Smith; J R Saunderson; A W Beavis
Journal:  Br J Radiol       Date:  2015-01-09       Impact factor: 3.039

4.  Use of a computer simulator to investigate optimized tube voltage for chest imaging of average patients with a digital radiography (DR) imaging system.

Authors:  Craig Steven Moore; Tim Wood; Ged Avery; Steve Balcam; Liam Needler; Hiten Joshi; Najeeb Ahmed; John Saunderson; Andrew Beavis
Journal:  Br J Radiol       Date:  2019-10-07       Impact factor: 3.039

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.