Short-term repeatability of myocardial blood flow using 82Rb PET/CT: The effect of arterial input function position and motion correction.

Yuka Otaki1, Martin Lyngby Lassen1, Osamu Manabe1,2, Evann Eisenberg1, Heidi Gransar1, Frances Wang1, Yoon Jae Lee1, Evangelos Tzolos1,3, Daniel S Berman1, Piotr J Slomka4.   

Abstract

BACKGROUND: We tested the repeatability of myocardial blood flow (MBF) quantified using 82Rb with and without motion correction (MC) and with arterial input functions estimated from left ventricle (LV) and atrium (LA).
METHODS: Twenty-one patients referred for clinical 82Rb PET/CT underwent repeated rest scans in a single imaging session. Global MBF was quantified using three different assessments by two operators: (1) automatic processing without MC and LV arterial input function (AIF), (2) with MC and LV-AIF, and (3) with MC and LA-AIF. Inter-scan and inter-operator repeatability were tested using coefficient of variation (CV).
RESULTS: MC with LV-AIF did not change MBF (no MC: 1.01 ± 0.30 mL/min/g vs MC with LV-AIF: 1.01 ± 0.29, P = 0.70), whereas MC with LA-AIF showed significantly lower MBF assessments (0.95 ± 0.28 mL/min/g, P = 0.0006). We report significant improvement for test-retest reproducibility for global MBF following MC (CV; No MC: 16.0, MC (LV-AIF): 9.2, MC (LA-AIF): 8.8). Good inter-operator repeatability was observed for LV-AIF (CV = 4.7) and LA-AIF (CV = 5.6) for global MBF assessments.
CONCLUSIONS: MC significantly improved the test-retest repeatability between operators and between scans. MBF obtained after MC with LV-AIF were comparable, whereas MBFs after MC and LA-AIF were significantly reduced.
© 2019. American Society of Nuclear Cardiology.

Entities:  

Keywords:  Myocardial blood flow; arterial input function; positron emission tomography; rubidium

Mesh:

Substances:

Year:  2019        PMID: 31559536     DOI: 10.1007/s12350-019-01888-5

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  3 in total

1.  Automatic quantification of ejection fraction from gated myocardial perfusion SPECT.

Authors:  G Germano; H Kiat; P B Kavanagh; M Moriel; M Mazzanti; H T Su; K F Van Train; D S Berman
Journal:  J Nucl Med       Date:  1995-11       Impact factor: 10.057

2.  Automated dynamic motion correction using normalized gradient fields for 82rubidium PET myocardial blood flow quantification.

Authors:  Benjamin C Lee; Jonathan B Moody; Alexis Poitrasson-Rivière; Amanda C Melvin; Richard L Weinberg; James R Corbett; Venkatesh L Murthy; Edward P Ficaro
Journal:  J Nucl Cardiol       Date:  2018-11-07       Impact factor: 5.952

3.  Comparison of diagnostic accuracy of PET-derived myocardial blood flow parameters: A meta-analysis.

Authors:  Sang-Geon Cho; Soo Jin Lee; Myung Hwan Na; Yun Young Choi; Henry Hee-Seung Bom
Journal:  J Nucl Cardiol       Date:  2018-11-02       Impact factor: 5.952

  3 in total
  13 in total

1.  Myocardial blood flow reserve and absolute myocardial blood flow for the assessment of patients with coronary artery disease with or without microvascular dysfunction.

Authors:  Raffaele Giubbini; Elisabetta Cerudelli; Luca Camoni
Journal:  J Nucl Cardiol       Date:  2020-08-04       Impact factor: 5.952

2.  Clinical implications of compromised 82Rb PET data acquisition.

Authors:  Andrew Van Tosh; J Jane Cao; John R Votaw; C David Cooke; Christopher J Palestro; Kenneth J Nichols
Journal:  J Nucl Cardiol       Date:  2021-08-20       Impact factor: 3.872

3.  Myocardial flow reserve estimation with contemporary CZT-SPECT and 99mTc-tracers lacks precision for routine clinical application.

Authors:  Edward P Ficaro; Venkatesh L Murthy; Jennifer M Renaud; Alexis Poitrasson-Rivière; Tomoe Hagio; Jonathan B Moody; Liliana Arida-Moody
Journal:  J Nucl Cardiol       Date:  2021-08-23       Impact factor: 3.872

4.  Myocardial revascularization driven by functional testing and PET imaging.

Authors:  Pierre-Yves Marie; Alain Manrique
Journal:  J Nucl Cardiol       Date:  2019-12-16       Impact factor: 5.952

5.  A big step towards clinical implementation of myocardial blood flow quantification with CZT SPECT.

Authors:  Matthieu Pelletier-Galarneau; Terrence D Ruddy
Journal:  J Nucl Cardiol       Date:  2019-09-18       Impact factor: 5.952

6.  Optimizing accuracy and precision with motion correction of PET myocardial blood flow measurements.

Authors:  Alexis Poitrasson-Rivière; Venkatesh L Murthy
Journal:  J Nucl Cardiol       Date:  2019-10-23       Impact factor: 5.952

7.  More evidence for adequate test-retest repeatability of myocardial blood flow quantification with 82Rb PET/CT.

Authors:  Robert A deKemp; Emel Celiker Guler; Terrence D Ruddy
Journal:  J Nucl Cardiol       Date:  2020-06-25       Impact factor: 5.952

Review 8.  Quantitative clinical nuclear cardiology, part 2: Evolving/emerging applications.

Authors:  Piotr J Slomka; Jonathan B Moody; Robert J H Miller; Jennifer M Renaud; Edward P Ficaro; Ernest V Garcia
Journal:  J Nucl Cardiol       Date:  2020-10-16       Impact factor: 5.952

9.  Anomalous origin of the left coronary artery from the pulmonary artery: The role of CZT camera.

Authors:  Hong-Xin Chu; Xiu-Li Wang; Mahmoud Alsalenhi; Zhi-Gang Liu; Jian-Ming Li
Journal:  J Nucl Cardiol       Date:  2021-12-06       Impact factor: 5.952

10.  Moving forward with motion reduction, detection and correction in cardiac PET.

Authors:  Ernest V Garcia; Jonathon A Nye
Journal:  J Nucl Cardiol       Date:  2021-03-21       Impact factor: 3.872

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.