| Literature DB >> 31525218 |
Sheila M Keating1,2, Wes Rountree3, Eduard Grebe1,2, Andrea L Pappas3, Mars Stone1,2, Dylan Hampton1, Christopher A Todd3, Marek S Poniewierski3, Ana Sanchez3, Cassandra G Porth3, Thomas N Denny3, Michael P Busch1,2.
Abstract
Laboratory assays for identifying recent HIV-1 infections are widely used for estimating incidence in cross-sectional population-level surveys in global HIV-1surveillance. Adequate assay and laboratory performance are required to ensure accurate incidence estimates. The NIAID-supported External Quality Assurance Program Oversight Laboratory (EQAPOL) established a proficiency testing program for the most widely-used incidence assay, the HIV-1 Limiting Antigen Avidity EIA (LAg), with US Centers for Disease Control and Prevention (CDC)-approved kits manufactured by Sedia Biosciences Corporation and Maxim Biomedical. The objective of this program is to monitor the performance of participating laboratories. Four rounds of blinded external proficiency (EP) panels were distributed to up to twenty testing sites (7 North American, 5 African, 4 Asian, 2 South American and 2 European). These panels consisted of ten plasma samples: three blinded well-characterized HIV-1-seropositive samples that were included as replicates and an HIV-negative control. The seropositive samples spanned the dynamic range of the assay and are categorized as either recent or long-term infection. Participating sites performed the assay according to manufacturers' instructions and completed an online survey to gather information on kit manufacturer, lot of kit used, laboratory procedures and the experience of technicians. On average, fifteen sites participated in each round of testing, with an average of four sites testing with only the Maxim assay, seven testing with only the Sedia assay and five sites utilizing both assays. Overall, the Sedia and Maxim assays yielded similar infection status categorization across the laboratories; however, for most of the nine HIV+ samples tested, there were significant differences in the optical density readouts, ODn (N = 8) and OD (N = 7), between LAg kit manufacturers (p < 0.05 based on mixed effects models. The EQAPOL LAg program is important for monitoring laboratory performance as well as detecting variations between manufacturers of HIV-1incidence assays.Entities:
Mesh:
Substances:
Year: 2019 PMID: 31525218 PMCID: PMC6746377 DOI: 10.1371/journal.pone.0222290
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Proficiency criteria and average scores.
| Grading Criteria and Points Allocation | Average Scores | |||||||
|---|---|---|---|---|---|---|---|---|
| EP2 | EP3 | EP4 | ||||||
| Points in EP3 | Points in EP4 | Sedia | Maxim (N = 8) | Sedia (N = 13) | Maxim (N = 8) | Sedia (N = 12) | Maxim (N = 9) | |
| 9.1 | 7.5 | 10 | 10 | 10 | 10 | |||
| 8.2 | 10 | 10 | 10 | 24 | 23.6 | |||
| 50.5 | 54.4 | 60 | 60 | 45 | 46.7 | |||
| 15.1 | 13.3 | 14.9 | 14 | 12.3 | 12.7 | |||
Fig 1Comparison of Maxim and Sedia OD and ODn values.
Panel A shows OD readings and Panel B shows ODn readings. The blue line is the linear regression line and the red line shows the slope if the two kits had produced equivalent results.
Between-site and within-site variances for kit calibrators, EPs 1–4 (mixed effects model estimates).
| Model-Based Variance Mean (95% CI) | |||
|---|---|---|---|
| Variance Type | Maxim Kit | Sedia Kit | Result |
| Between Site | 0.00473 (0.00205, 0.02014) | 0.00247 (0.00118, 0.00806) | Variances Equal |
Comparison of between-site and within-site variances of OD and ODn values for EPs 1–4 (mixed effects models estimates).
| Model-Based Variance Mean (95% CI) | Intraclass Correlation | |||||
|---|---|---|---|---|---|---|
| Kit Type | Variance Type | OD Values | ODn Value | Result | OD ICC | ODn ICC |
| . | . | |||||
| 0.949 | 0.967 | |||||
| . | . | |||||
| 0.962 | 0.959 | |||||
| . | . | |||||
| 0.959 | 0.963 | |||||
Variances adjusted for site, kit type, and sample.
Comparison of average OD values by kit type for EPs 1–4 (mixed effects model estimates).
| Model-Based Mean (95% CI) | |||||
|---|---|---|---|---|---|
| ID | Sedia Mean | Maxim Mean | Ratio | Sedia—Maxim: Mean (95% CI) | p-value |
| LA_0001 | 0.0570 | 0.0247 | 2.3 | 0.0323 (-0.0003, 0.0649) | 0.0518 |
| LA_0009 | 0.6608 | 0.6470 | 1.0 | 0.0138 (-0.0188, 0.0464) | 0.4067 |
Comparison of average ODn values by kit type for EPs 1–4 (mixed effects model estimates).
| Model-Based Mean (95% CI) | |||||
|---|---|---|---|---|---|
| ID | Sedia Mean | Maxim Mean | Ratio | Sedia—Maxim: Mean (95% CI) | p-value |
| LA_0009 | 0.9561 | 0.9630 | 1.0 | -0.0069 (-0.0515, 0.0376) | 0.7604 |
Fig 2Mean ODn values for each sample by site and kit type.
Maxim kits are shown in green and Sedia kits are in purple. Eight out of nine samples had significantly higher Sedia means than Maxim means.
Data summary of laboratories by kit type for EPs 1–4.
| Site | Maxim EPs | Sedia EPs | Maxim Calibrators | Sedia Calibrators | Total Calibrators | Maxim | Sedia | Total |
|---|---|---|---|---|---|---|---|---|
| 059 | 4 | 4 | 24 | 24 | 48 | 95 | 89 | 184 |
| 060 | 4 | 4 | 24 | 24 | 48 | 99 | 89 | 188 |
| 061 | 1 | 4 | 6 | 24 | 30 | 21 | 89 | 110 |
| 065 | 4 | 24 | 24 | 89 | 89 | |||
| 066 | 3 | 18 | 18 | 77 | 77 | |||
| 067 | 2 | 12 | 12 | 47 | 47 | |||
| 069 | 1 | 6 | 6 | 21 | 21 | |||
| 071 | 2 | 12 | 12 | 47 | 47 | |||
| 072 | 4 | 24 | 24 | 83 | 83 | |||
| 073 | 1 | 6 | 6 | 21 | 21 | |||
| 074 | 1 | 6 | 6 | 28 | 28 | |||
| 075 | 4 | 24 | 24 | 87 | 87 | |||
| 076 | 1 | 6 | 6 | 21 | 21 | |||
| 077 | 3 | 18 | 18 | 74 | 74 | |||
| 078 | 4 | 24 | 24 | 95 | 95 | |||
| 086 | 2 | 12 | 12 | 53 | 53 | |||
| 090 | 2 | 12 | 12 | 47 | 47 | |||
| 106 | 4 | 4 | 24 | 24 | 48 | 95 | 89 | 184 |
| 107 | 4 | 4 | 24 | 24 | 48 | 95 | 91 | 186 |
| 216 | 4 | 24 | 24 | 89 | 89 | |||