| Literature DB >> 12843746 |
Bhupat D Rawal1, Azucena Degula, Ludmila Lebedeva, Robert S Janssen, Frederick M Hecht, Haynes W Sheppard, Michael P Busch.
Abstract
The sensitive/less-sensitive (S/LS) enzyme immunoassay (EIA) testing strategy for discriminating "early" from "longstanding" HIV infection has been widely applied for detecting recent seroconverters and estimating HIV incidence rates. The originally developed assay (3A11-LS EIA; Abbott Laboratories, Abbott Park, IL) involved performance of LS EIAs using a bead-based assay that required specialized equipment and reagents of limited availability. In contrast, 96-microwell-based EIAs are more universally applied for HIV serodiagnosis throughout the world. The authors report development and preliminary validation of an LS protocol using an EIA in a 96-well format: the Vironostika HIV-1 MicroElisa System (Vironostika-LS EIA; Bio Merieux, Raleigh, NC). The results with samples from recent HIV-1 seroconverters, persons with longstanding HIV-1 asymptomatic infection, patients on highly active antiretroviral therapy, and AIDS patients show a high degree of correlation between the Vironostika-LS EIA and 3A11-LS EIA. The authors also demonstrate that the Abbott 3A11-LS EIA and Vironostika-LS EIA performed comparably on HIV-1-positive samples from persons infected with non-B HIV-1 subtypes. These results support the potential use of the Vironostika-LS EIA for detection of recent HIV-1 infections for incidence projections and for other epidemiologic, clinical, and molecular surveillance applications.Entities:
Mesh:
Substances:
Year: 2003 PMID: 12843746 DOI: 10.1097/00126334-200307010-00009
Source DB: PubMed Journal: J Acquir Immune Defic Syndr ISSN: 1525-4135 Impact factor: 3.731