Shu Tu1, Kang Li1, Xiaohu Ding1, Chengguo Zuo1, Dongpeng Hu1, Jian Ge1.
Abstract
Purpose: The purpose of this study is to evaluate the consistency of and deviation in intraocular pressure (IOP) measurements obtained by the TonoVet rebound tonometer and the Tonopen applanation tonometer in a primate model.Entities:
Year: 2019 PMID: 31523117 PMCID: PMC6707614
Source DB: PubMed Journal: Mol Vis ISSN: 1090-0535 Impact factor: 2.367
Figure 1Consistency analysis in monkey eyes with normal intraocular pressure (IOP). A: Regression analysis between TonoVet rebound and Tonopen applanation tonometer readings (n= 90 readings) demonstrated that the tonometer readings obtained using these two devices were strongly and linearly correlated; B: Consistency between TonoVet and Tonopen readings was evaluated using Bland-Altman analysis, with the means plotted against the deviations between TonoVet and Tonopen readings.
Evaluation of the consistency between the Tonopen and the TonoVet measurement in control healthy monkeys.
| Tonometers | IOP ≤21 mmHg (n=90 readings) | |
|---|---|---|
| Tonopen | TonoVet | |
| Mean ± SD | 12.92±2.31 | 12.99±3.04 |
| Range | 6-20 | 8-21 |
| Δ Mean ± SD | 0.06±2.08 | |
| Δ Range | -5-7 | |
| 95% CI | -4.13 – 4.00 | |
| Δ < ±1 mmHg* | 56.67% | |
| Δ < ±3 mmHg† | 91.12% | |
| P | 0.761 | |
Δ Mean ± SD: mean and standard deviation of difference (TonoVet minus Tonopen) * IOP difference smaller than ± 1 mmHg † IOP difference smaller than ± 3 mmHg
Figure 2A monkey model of chronic ocular hypertension was established. A: Fundus: The enlarged optic cup. B: Optical coherence tomography (OCT): The damaged retinal nerve fiber layer. C: OCT: The enlarged optic cup.
Figure 3Consistency analysis in COHT monkey models. A: Regression analysis between TonoVet rebound and Tonopen applanation tonometer readings (n= 72 readings); B: Deviations in IOP values (calculated as TonoVet minus Tonopen) are shown plotted against the average IOP ((TonoVet plus Tonopen)/2). The solid line represents the regression line, and the dotted lines represent 95% confidence interval limits. Duplicates were distributed in the graph.
Figure 4Consistency between TonoVet and Tonopen measurements was evaluated using the Bland–Altman analysis in animals with different IOP levels in the COHT monkey group. The averages are shown plotted against the deviations. Higher intraocular pressure (IOP) values were associated with worse consistency for both tonometers. The solid line represents the average deviation, and the dotted lines indicate the 95% limits of the confidence intervals.
Evaluation of the consistency between the Tonopen and the TonoVet measurement in COHT monkey models.
| Tonometers | IOP 7–16 mmHg
(n=13 readings) | IOP 17–22 mmHg
(n=22 readings) | IOP 23–60 mmHg
(n=37 readings) | |||
|---|---|---|---|---|---|---|
| Tonopen | TonoVet | Tonopen | TonoVet | Tonopen | TonoVet | |
| Mean ± SD | 14.15 | 18.62 | 19.23 | 24.18 | 29.3 | 43.05 |
| | ± 1.14 | ± 1.98 | ± 1.31 | ± 3.62 | ± 6.12 | ± 14.29 |
| Range | 12-16 | 16 - 21 | 17 - 22 | 19 - 34 | 23 - 47 | 22 - 71 |
| Δ Mean ± SD | 4.46±1.66 | | 4.95±3.57 | | 13.76±9.19 | |
| Δ Range | 1-7 | 0 - 15 | -32 | |||
| 95% CI | 3.46 - 5.47 | 3.37 - 6.54 | 10.69 - 16.82 | |||
| Δ > ±5 mmHg* | 23.08% | 36.36% | 75.68% | |||
| P | 0 | 0 | 0.003 | |||
Δ Mean ± SD: mean and standard deviation of difference (TonoVet minus Tonopen) * IOP difference bigger than ± 5 mmHg