| Literature DB >> 31491870 |
Ching-Yi Liu1,2, Hsin-Kuo Ko3,4, James B Fink5, Gwo-Hwa Wan6,7,8, Chung-Chi Huang9,10, Yu-Chun Chen11, Hui-Ling Lin12,13,14.
Abstract
Although aerosol delivery through mechanical ventilators has been used to administer various medications, little is known of administration with colistin. This in vitro evaluation aimed to evaluate size distribution of colistin delivery by different types of nebulizers and concentrations during mechanical ventilation. Colistin methanesulfonate (colistin) for injection was dissolved in 6 mL of distilled water to produce a low concentration (L; 156 mg) and a high concentration (H; 312 mg). A dose volume of 6 mL was placed in a vibrating mesh nebulizer (VMN) and a jet nebulizer (JN). The inhaled mass (mean ± SD) of the VMN-L (53.80 ± 14.79 mg) was greater than both the JN-L (19.82 ± 3.34 mg, P = 0.001) and JN-H (31.72 ± 4.48 mg, P = 0.017). The nebulization time of the VMN-L (42.35 ± 2.30 min) was two times longer than the JN-L (21.12 ± 0.8 min) or JN-H (21.65 ± 0.42 min; P < 0.001). The mass median aerodynamic distal to the endotracheal tube was within a similar range at 2.03 to 2.26 μm (P = 0.434), independent of neb or formulation concentration. In conclusion, the VMN-L yields greater inhaled mass than the JN with either concentration. Therefore, a standard nominal dose of colistin results in a higher delivered dose during mechanical ventilation with a VMN compared with a JN and may be considered the preferred device. If JN must be used, multiple doses of low concentration colistin may compensate for poor delivery performance.Entities:
Keywords: drug concentration; inhaled colistin; inhaled drug mass; jet nebulizer; mechanical ventilation; particle size distributions; vibrating mesh nebulizer
Year: 2019 PMID: 31491870 PMCID: PMC6781281 DOI: 10.3390/pharmaceutics11090459
Source DB: PubMed Journal: Pharmaceutics ISSN: 1999-4923 Impact factor: 6.321
Figure 1Experimental apparatus set-up. A ventilator was connected to a test lung via an endotracheal tube with an inline collecting filter or Andersen Cascade Impactor. SVN: small volume nebulizer, VMN: vibrating mesh nebulizer.
Nebulizer performance on the delivery of colistin.
| Variables | VMN-L | JN-L | JN-H | |
|---|---|---|---|---|
| Inhaled mass (mg) | 53.80 ± 14.79 | 19.82 ± 3.34 * | 31.72 ± 4.48 * | <0.001 |
| Inhaled mass (%) | 34.44 ± 9.47 | 12.69 ± 2.14 † | 10.15 ± 1.43 † | <0.001 |
| Nebulization time (min) | 42.35 ± 2.30 | 21.12 ± 0.86 ‡ | 21.65 ± 0.42 ‡ | <0.001 |
| Delivery efficiency (mg/min) | 1.27 ± 0.32 | 0.94 ± 0.17 | 1.46 ± 0.20 § | 0.023 |
| Delivery efficiency (%/min) | 0.81 ± 0.20 | 0.60 ± 0.1 | 0.47 ± 0.06 ‖ | 0.014 |
Note: Data are presented as means ± standard deviation; VMN: vibrating mesh nebulizer; JN: jet nebulizer. * P = 0.001, JN-L versus VMN-L; P = 0.017 JN-H versus VMN-L; † P < 0.001, JN-L and JN-H versus VMN-L; P < 0.001, JN-L and JN-H versus VMN-L; § P = 0.025, JN-L versus JN-H; ‖ P = 0.015, JN-H versus VMN-L.
Particle size distribution of nebulized colistin distal to the endotracheal tube.
| Variables | VMN-L | JN-L | JN-H | |
|---|---|---|---|---|
| MMAD (µm) | 2.03 ± 0.24 | 2.09 ± 0.17 | 2.26 ± 0.05 | 0.434 |
| GSD | 1.58 ± 0.14 | 1.59 ± 0.10 | 1.58 ± 0.01 | 0.994 |
| Fine particle % (1.1 to 4.7 µm) | 87.83 ± 3.08 | 90.69 ± 4.55 | 95.83 ± 0.36 | 0.128 |
Note: Data are presented as means ± standard deviation; VMN: vibrating mesh nebulizer; JN: jet nebulizer; GSD: geometric standard deviation; MMAD: median aerodynamic diameter.
Figure 2Particle size distribution at each stage of the Andersen Cascade Impactor.