Spencer P Lake1, Nathaniel F N Stoikes2, Amit Badhwar3, Corey R Deeken4.
Abstract
BACKGROUND: Hybrid hernia meshes combine biological tissue-derived extracellular matrix with permanent or resorbable synthetic. The objective of this study was to evaluate hybrid meshes (Gore® Synecor, Zenapro™, Ovitex™ 1S Reinforced Bioscaffold Permanent, and Ovitex™ 1S Reinforced Bioscaffold Resorbable) compared to non-hybrid, bioresorbable synthetic mesh (Phasix™ Mesh) in a rabbit bacterial inoculation model.Entities:
Keywords: Bacterial colonization; Hernia mesh; Hybrid material; Infection; Methicillin-resistant Staphylococcus aureus (MRSA); Rabbit model
Year: 2019 PMID: 31467674 PMCID: PMC6710816 DOI: 10.1016/j.amsu.2019.08.004
Source DB: PubMed Journal: Ann Med Surg (Lond) ISSN: 2049-0801
Fig. 1Representative macroscopic photographs of each mesh type prior to implantation (first column), during implantation of mesh into subcutaneous pocket (second column), 7 days after implantation/inoculation with MRSA just prior to explantation of mesh (third column), and after explantation of mesh (fourth column). Note: Each mesh was a 3.8 cm diameter disk prior to implantation.
Fig. 2Semi-quantitative abscess analysis demonstrates significantly lower white material scores for Synecor and Phasix™ Mesh compared to Zenapro™, Ovitex™ 1S Permanent, Ovitex™ 1S Resorbable (*p < 0.05 vs. Phasix™ Mesh; #p < 0.05 vs. Synecor; all values are shown with lines representing median values by group.
Fig. 3(A) Percentage of pockets for each mesh type that contained recoverable bacteria and (B) measured quantity of MRSA extracted from pockets containing meshes (reported in colony forming units or CFU) demonstrate favorable bacterial clearance for Synecor and superior bacterial clearance for Phasix™ Mesh compared to other hybrid meshes tested (*p < 0.05 vs. Phasix™ Mesh; #p < 0.05 vs. Synecor; all values are shown with lines representing median values by group).
Evaluation of microbial colonization of different types of implanted meshes, reported as colony forming units (CFU) of MRSA. Note: the initial inoculation level was 106 CFU of MRSA. (n = 10 for all groups except Zenapro™ with n = 6).
| Phasix™ Mesh | Synecor | Zenapro™ | Ovitex™ 1S Permanent | Ovitex™ 1S Resorbable |
|---|---|---|---|---|
| 0 | 1.64E+07 | 9.38E+07 | 1.68E+07 | 2.53E+08 |
| 0 | 8.58E+04 | 1.49E+07 | 2.07E+07 | 1.68E+08 |
| 0 | 0 | 5.18E+07 | 9.33E+07 | 1.19E+08 |
| 0 | 0 | 2.88E+07 | 8.28E+07 | 2.69E+08 |
| 0 | 0 | 8.28E+07 | 3.50E+07 | 2.18E+08 |
| 0 | 0 | 9.50E+07 | 6.23E+07 | 1.35E+08 |
| 0 | 0 | -- | 2.24E+08 | 1.28E+08 |
| 0 | 0 | -- | 7.70E+07 | 1.56E+08 |
| 0 | 0 | -- | 8.05E+07 | 1.21E+07 |
| 0 | 0 | -- | 1.46E+08 | 8.00E+07 |