D E Bergeron1, J T Cessna2, R Fitzgerald2, L Pibida2, B E Zimmerman2.
Abstract
A solution of 124I was standardized for activity by 4πβ(LS)-γ(NaI) live-timed anticoincidence (LTAC) counting, with confirmatory measurements by triple-to-double coincidence ratio (TDCR) and CIEMAT-NIST efficiency tracing (CNET) liquid scintillation counting. The LTAC-based standard was shown to be in agreement (within k = 1 uncertainties) with previous measurements at NIST and elsewhere. Calibration settings for radionuclide calibrators were determined and a discrepancy with literature values, partially due to a calibration methodology dependent upon an erroneous setting for 18F, was identified and explained. Published by Elsevier Ltd.Entities:
Keywords: Anticoincidence; Dose calibrator; Fluorine-18; Iodine-124; Liquid scintillation counting; Monte Carlo; Positron emission tomography; Quantitative imaging
Year: 2019 PMID: 31465950 PMCID: PMC6839768 DOI: 10.1016/j.apradiso.2019.108849