Standardization of I-124 by three liquid scintillation-based methods.

D E Bergeron1, J T Cessna2, R Fitzgerald2, L Pibida2, B E Zimmerman2.   

Abstract

A solution of 124I was standardized for activity by 4πβ(LS)-γ(NaI) live-timed anticoincidence (LTAC) counting, with confirmatory measurements by triple-to-double coincidence ratio (TDCR) and CIEMAT-NIST efficiency tracing (CNET) liquid scintillation counting. The LTAC-based standard was shown to be in agreement (within k = 1 uncertainties) with previous measurements at NIST and elsewhere. Calibration settings for radionuclide calibrators were determined and a discrepancy with literature values, partially due to a calibration methodology dependent upon an erroneous setting for 18F, was identified and explained. Published by Elsevier Ltd.

Entities:  

Keywords:  Anticoincidence; Dose calibrator; Fluorine-18; Iodine-124; Liquid scintillation counting; Monte Carlo; Positron emission tomography; Quantitative imaging

Year:  2019        PMID: 31465950      PMCID: PMC6839768          DOI: 10.1016/j.apradiso.2019.108849

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  20 in total

1.  Software studies for germanium detectors data analysis.

Authors:  L Pibida; E Hsieh; A Fuentes-Figueroa; M M Hammond; L Karam
Journal:  Appl Radiat Isot       Date:  2006-03-29       Impact factor: 1.513

2.  A new NIST primary standardization of 18F.

Authors:  R Fitzgerald; B E Zimmerman; D E Bergeron; J C Cessna; L Pibida; D S Moreira
Journal:  Appl Radiat Isot       Date:  2013-12-04       Impact factor: 1.513

3.  Monte Carlo based approach to the LS–NaI 4πβ–γ anticoincidence extrapolation and uncertainty.

Authors:  R Fitzgerald
Journal:  Appl Radiat Isot       Date:  2016-03       Impact factor: 1.513

4.  An automated ionization chamber for secondary radioactivity standards.

Authors:  R Fitzgerald
Journal:  Appl Radiat Isot       Date:  2010-01-19       Impact factor: 1.513

5.  Impact of Recent Change in the National Institute of Standards and Technology Standard for 18F on the Relative Response of 68Ge-Based Mock Syringe Dose Calibrator Standards.

Authors:  Brian E Zimmerman; Denis E Bergeron; Jeffrey T Cessna
Journal:  J Nucl Med       Date:  2015-07-16       Impact factor: 10.057

6.  Experimental determination of some nuclear decay data in the decays of (177)Lu, (186)Re and (124)I.

Authors:  Aurelian Luca; Maria Sahagia; Mihail-Razvan Ioan; Andrei Antohe; Beatris Luminita Savu
Journal:  Appl Radiat Isot       Date:  2015-12-02       Impact factor: 1.513

7.  A phantom study: Should 124 I-mIBG PET/CT replace 123 I-mIBG SPECT/CT?

Authors:  Casper Beijst; Bart de Keizer; Marnix G E H Lam; Geert O Janssens; Godelieve A M Tytgat; Hugo W A M de Jong
Journal:  Med Phys       Date:  2017-04-17       Impact factor: 4.071

8.  Standardisation and half-life of 89Zr.

Authors:  E García-Toraño; V Peyrés; M Roteta; M Mejuto; A Sánchez-Cabezudo; E Romero
Journal:  Appl Radiat Isot       Date:  2017-10-18       Impact factor: 1.513

Review 9.  Production of iodine-124 and its applications in nuclear medicine.

Authors:  Ana Maria S Braghirolli; William Waissmann; Juliana Batista da Silva; Gonçalo R dos Santos
Journal:  Appl Radiat Isot       Date:  2014-04-04       Impact factor: 1.513

10.  A recommendation for revised dose calibrator measurement procedures for 89Zr and 124I.

Authors:  Bradley J Beattie; Keith S Pentlow; Joseph O'Donoghue; John L Humm
Journal:  PLoS One       Date:  2014-09-09       Impact factor: 3.240